Skip to content

Halyard PSSC004-03_0001Recalls & reported problems

Avid Medical, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10809160348895), received between Feb 21, 2025 and Feb 24, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Contamination” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 20, 2024Class IIOpen, Classified

    Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

    Z-0422-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Contamination2

Patient problems

  • Unspecified Infection2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 24, 2025Contamination; Contamination

    INCIDENT ONE: PATIENT WAS REPORTED TO HAVE A POST-SURGICAL INFECTION. THERE WAS MEDICAL INTERVENTION (ANTIBIOTICS) PROVIDED FOR THIS PATIENT WHICH REQUIRED A PROLONG HOSPITAL STAY DUE TO COMPLICATIONS FROM INFECTION. AN ADDITIONAL SURGICAL PROCEDURE WAS REQUIRED. ADDITIONAL INFORMATION WAS REQUESTED ON FEBRUARY 6, 2025, FEBRUARY 11, 2025, AND FEBRUARY 17, 2025, AT THIS TIME THE SITE IS AWAITING FURTHER INFORMATION ABOUT THE INFECTION, TYPE OF PROCEDURE, AND TECHNIQUE PERFORMED BY THE HOSPITAL.

    Patient problems: Unspecified Infection

    MAUDE report 21453830 ↗
  • InjuryFeb 21, 2025Contamination; Contamination

    INCIDENT TWO: PATIENT WAS REPORTED TO HAVE A POST-SURGICAL INFECTION. THERE WAS MEDICAL INTERVENTION (ANTIBIOTICS) PROVIDED FOR THIS PATIENT. ADDITIONAL INFORMATION WAS REQUESTED ON FEBRUARY 6, 2025, FEBRUARY 11, 2025, AND FEBRUARY 17, 2025, THE SITE IS CURRENTLY AWAITING FURTHER DETAILS INCLUDING DETAILS OF THE TYPE OF INFECTION, TYPE OF PROCEDURE, AND TECHNIQUE UTILIZED BY THE CUSTOMER.

    Patient problems: Unspecified Infection

    MAUDE report 21436894 ↗

Other Cardiac catheterization kit devices

All Cardiac catheterization kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page