Halyard EAMC1000-05_0015Recalls & reported problems
Avid Medical, Inc. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Apr 30, 2026Class IOpen, Classified
Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels.
Z-2796-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Mar 13, 2026Class IOpen, Classified
Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Z-1970-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Other Cardiac catheterization kit devices
- Medline Industries, Inc. DYNJ44700JMedline Industries, Inc. · Cardiac catheterization kit3 recalls1 report
- Medline Industries, Inc. ECVC2555Centurion Medical Products Corporation · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. DYNJ83089Medline Industries, Inc. · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. DYNJ0578916XMedline Industries, Inc. · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. MNS9935Medline Industries, Inc. · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. ECVC5540ACenturion Medical Products Corporation · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. ECVC2745Centurion Medical Products Corporation · Cardiac catheterization kit3 recalls0 reports
- Medline Industries, Inc. ECVC6505Centurion Medical Products Corporation · Cardiac catheterization kit3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page