The LAP-BAND Adjustable Gastric Banding System B-2210Recalls & reported problems
Apollo Endosurgery, Inc. · LAP-BAND SYSTEM, 9.75, STERILE, US · Class 3 · Implant
All models of The LAP-BAND Adjustable Gastric Banding System (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10811955020145), received between May 3, 2013 and Nov 21, 2017 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak5
- Leak/Splash1
Patient problems
- Failure of Implant4
- Patient Problem/Medical Problem1
- Weight Changes1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2017Fluid/Blood Leak; Fluid/Blood Leak
REPORTED AS: A PATIENT WITH THE LAP-BAND SYSTEM WAS REPORTED TO HAVE BEEN "EVALUATED FOR A LAP BAND TUBING LEAK." [PATIENT] "FELT THERE WAS NO RESTRICTION AND THERE WAS MINIMAL FLUID IN [THEIR] SYSTEM. WHEN SALINE WAS INJECTED IT DID NOT RETURN FULL VOLUME. X-RAY SHOWED PROXIMAL TUBING LEAK, WHICH REQUIRED LAPAROSCOPY TO REPAIR. LEAK WAS TOO CLOSE TO BAND TO REPAIR AT LAPAROSCOPY." DEVICE WAS REMOVED.
Patient problems: Failure of Implant
MAUDE report 7049844 ↗MalfunctionApr 24, 2017Fluid/Blood Leak; Fluid/Blood Leak
REPORTED AS: A PATIENT WITH THE LAP-BAND SYSTEM WAS REPORTED TO HAVE "DEFECTIVE BAND, LOCATION OF WHERE THE PORT AND THE TUBING MEETS HAS A HOLE IN IT". PHYSICIAN CONFIRMED, DEVICE REMAINS IMPLANTED AT THIS TIME, BUT WILL BE REMOVED AND REPLACED.
Patient problems: Failure of Implant
MAUDE report 6516383 ↗MalfunctionFeb 10, 2017Fluid/Blood Leak; Fluid/Blood Leak
REPORTED AS: A PATIENT WITH THE LAP-BAND SYSTEM WAS REPORTED TO HAVE "OVER THE PAST SIX OR SO MONTHS [PATIENT] HAD NOTICED NO RESTRICTION, AND HAD GAINED LOTS OF WEIGHT. [PATIENT] WAS FOUND TO NOT HAVE ANY VOLUME IN [PATIENT'S] LAP BAND IN THE CLINIC. A DIAGNOSTIC STUDY WAS DONE REVEALING A LARGE LEAK IN THE TUBING. IT WAS DIFFICULT TO TELL WERE THE LEAK WAS LOCATED WHETHER IT WAS DISTAL NEAR THE PORT OR PROXIMAL NEAR THE LAP BAND. A DIAGNOSTIC LAPAROSCOPY WAS DONE TO GET AT THE SECTION, WITH HOPES OF REPAIRING THE TUBING BY EXTERIORIZING THE LEAK AND BRING IN INTO THE SURFACE AND SLICING A NEW PIECE OF TUBING, HOWEVER THERE WAS A POSSIBILITY THAT THIS WILL BE TOO CLOSE TO THE LAP BAND ITSELF. THAT REQUIRED ULTIMATELY REPLACING THE ENTIRE SYSTEM." THE DEVICE WAS EXPLANTED AND REPLACED.
Patient problems: Weight Changes
MAUDE report 6323455 ↗MalfunctionOct 27, 2015Leak/Splash
REPORTED AS: PATIENT COMPLAINED OF NO RESTRICTION OF THEIR LAP-BAND SYSTEM DURING A VISIT. THE PHYSICIAN PERFORMED AN ULTRASOUND OF THE ABDOMEN WITH FLUOROSCOPY, AND A CRACK IN THE TUBING WAS NOTED. THE LAP-BAND PORT WAS REMOVED AND REPLACED.
Patient problems: Patient Problem/Medical Problem
MAUDE report 5181508 ↗MalfunctionApr 13, 2015Fluid/Blood Leak
HEALTHCARE PROFESSIONAL REPORTED PT "COMPLAINED OF SUDDEN LACK OF RESTRICTION" AND DURING AN ADJUSTMENT THE PHYSICIAN WAS UNABLE TO WITHDRAW AS MUCH SALINE AS INDICATED SHOULD HAVE BEEN IN THE BAND. THE FLUOROSCOPY WAS "INCONCLUSIVE FOR LEAK" AND THE "PT WANTED TO HAVE THE PORT REPLACED ANYWAY". THE PORT WAS EXPLANTED AND REPLACED.
Patient problems: Failure of Implant
MAUDE report 4693029 ↗MalfunctionMay 3, 2013Fluid/Blood Leak
HEALTHCARE PROFESSIONAL REPORTED A LAP-BAND "BROKEN TUBING" AND "LEAK/BREAK IN TUBING." THE EVENT WAS FIRST NOTED WHEN THE PATIENT REPORTED "SUDDEN LOSS OF SATIETY." THE "BROKEN TUBING PIECES" WERE EXPLANTED.
Patient problems: Failure of Implant
MAUDE report 3101810 ↗
Other Implant, Intragastric For Morbid Obesity devices
- ORBERA Intragastric Balloon System B-4800Apollo Endosurgery, Inc. · Implant, Intragastric For Morbid Obesity6 deaths327 reports
- ReShape™ Integrated Dual Balloon System RSM101Reshape Medical, Inc. · Implant, Intragastric For Morbid Obesity3 deaths98 reports
- LAP-BAND AP Adjustable Gastric Banding System B-2240Apollo Endosurgery, Inc. · Implant, Intragastric For Morbid Obesity82 reports
- OBALON Touch™ 4300-0003Medtimo, Inc · Implant, Intragastric For Morbid Obesity1 recall0 reports
- LAP-BAND AP Adjustable Gastric Banding System B-2245Apollo Endosurgery, Inc. · Implant, Intragastric For Morbid Obesity27 reports
- The LAP-BAND Adjustable Gastric Banding System B-2220Apollo Endosurgery, Inc. · Implant, Intragastric For Morbid Obesity25 reports
- OBALON® Balloon Kit HPMC 7600-0001Medtimo, Inc · Implant, Intragastric For Morbid Obesity20 reports
- LAP-BAND AP Adjustable Gastric Banding System B-2260Apollo Endosurgery, Inc. · Implant, Intragastric For Morbid Obesity18 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page