LAP-BAND AP Adjustable Gastric Banding System B-2265Recalls & reported problems
Apollo Endosurgery, Inc. · LAP-BAND AP Large with Access Port II · Class 3 · Implant
All models of LAP-BAND AP Adjustable Gastric Banding System (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10811955020237), received between Mar 3, 2015 and Dec 14, 2018 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak7
- Appropriate Device Problem Term/Code Not Available1
- Leak/Splash1
Patient problems
- Failure of Implant5
- Patient Problem/Medical Problem3
- Pain1
- Sepsis1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 14, 2018Fluid/Blood Leak; Fluid/Blood Leak
REPORTED AS: A PATIENT WITH THE LAP-BAND SYSTEM WAS REPORTED TO HAVE A "LEAKY" LAP-BAND. THE BAND WAS REMOVED.
Patient problems: Failure of Implant
MAUDE report 8166592 ↗MalfunctionOct 26, 2017Fluid/Blood Leak; Fluid/Blood Leak
REPORTED AS: A PATIENT WITH THE LAP-BAND SYSTEM WAS REPORTED TO HAVE A LEAK. "[PATIENT] HAD NOT BEEN ABLE TO HOLD VOLUME FOR SEVERAL MONTHS. [PATIENT] HAD GAINED WEIGHT. RADIOGRAPHIC STUDIES WITH CONTRAST ARE NEGATIVE FOR BAND OR TUBING LEAK." DEVICE WAS REMOVED.
Patient problems: Failure of Implant
MAUDE report 6980611 ↗MalfunctionMay 24, 2017Fluid/Blood Leak; Fluid/Blood Leak
REPORTED AS: A PATIENT WITH THE LAP-BAND SYSTEM WAS REPORTED TO HAVE A LEAK. "SIX YEARS AGO [PATIENT] NOTED LOSS OF SATIETY. EVALUATIONS CONFIRMED TUBING BREAK WHICH WAS REPAIRED IN 2011. VOLUME CHECK OF [PATIENT'S] SYSTEM WAS 0 CC AND REPEATED INJECTIONS MEET NO RESISTANCE AND FAIL TO RETURN ANY VOLUME INDICATING A TUBING BREAK. CONTRAST RADIOLOGY SHOWS A COMPLETE TUBING BREAK ABOUT 6 INCHES FROM THE BAND." PORT AND TUBING REPLACEMENT.
Patient problems: Failure of Implant
MAUDE report 6589187 ↗MalfunctionSep 22, 2016Fluid/Blood Leak; Fluid/Blood Leak
REPORTED AS A PATIENT WITH THE LAP-BAND SYSTEM WAS REPORTED TO HAVE A "PORT LEAK." THE DEVICE WAS EXPLANTED.
Patient problems: Failure of Implant
MAUDE report 5970218 ↗MalfunctionAug 16, 2016Fluid/Blood Leak; Fluid/Blood Leak
REPORTED AS: A PATIENT WITH THE LAP-BAND SYSTEM WAS REPORTED TO HAVE A "LEAKING PORT." THE DEVICE WAS EXPLANTED.
Patient problems: Patient Problem/Medical Problem
MAUDE report 5878658 ↗MalfunctionApr 29, 2016Fluid/Blood Leak; Fluid/Blood Leak
REPORTED AS: THE PATIENT'S LAP-BAND SYSTEM "WASN'T GETTING RESTRICTION, CAME FOR VOLUME CHECK [THE PATIENT] WAS SUPPOSED TO HAVE 8CC BUT ONLY HAD 4CC." THE PORT WAS REMOVED AND REPLACED.
Patient problems: Patient Problem/Medical Problem
MAUDE report 5619536 ↗InjuryApr 7, 2016Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
REPORTED AS: A PATIENT WITH THE LAP-BAND SYSTEM WAS REPORTED TO HAVE "HAD ABDOMINAL PAIN, PRESENTED TO HOSPITAL AND FOUND TO HAVE FREE FLUID AND SEPSIS, BAND EXPLANTED. PATIENT PRESENTED TO HOSPITAL AND WAS TAKEN TO THE OPERATING ROOM. PATIENT WANTS THE BAND REPLACED."
Patient problems: Pain; Sepsis
MAUDE report 5555132 ↗MalfunctionOct 21, 2015Leak/Splash
REPORTED AS: PATIENT WAS HAVING ADJUSTMENTS PERFORMED, STATED THE LAP-BAND SYSTEM WAS WORKING WELL UNTIL 6 MONTHS AGO WHEN THEY LOST ALL RESTRICTION. PATIENT HAD DIAGNOSTIC TESTING AND THE PHYSICIAN DETERMINED THERE WAS A PORT LEAK AND THE PORT WAS REPLACED.
Patient problems: Patient Problem/Medical Problem
MAUDE report 5163747 ↗MalfunctionMar 3, 2015Fluid/Blood Leak
HEALTHCARE PROFESSIONAL REPORTED EXPLANT AND REPLACEMENT OF ORIGINAL LAP-BAND DEVICE DUE TO LEAK AND EXPLANT AND REPLACEMENT OF SECOND PORT.
Patient problems: Failure of Implant
MAUDE report 4580092 ↗
Other Implant, Intragastric For Morbid Obesity devices
- ORBERA Intragastric Balloon System B-4800Apollo Endosurgery, Inc. · Implant, Intragastric For Morbid Obesity6 deaths327 reports
- ReShape™ Integrated Dual Balloon System RSM101Reshape Medical, Inc. · Implant, Intragastric For Morbid Obesity3 deaths98 reports
- LAP-BAND AP Adjustable Gastric Banding System B-2240Apollo Endosurgery, Inc. · Implant, Intragastric For Morbid Obesity82 reports
- OBALON Touch™ 4300-0003Medtimo, Inc · Implant, Intragastric For Morbid Obesity1 recall0 reports
- LAP-BAND AP Adjustable Gastric Banding System B-2245Apollo Endosurgery, Inc. · Implant, Intragastric For Morbid Obesity27 reports
- The LAP-BAND Adjustable Gastric Banding System B-2220Apollo Endosurgery, Inc. · Implant, Intragastric For Morbid Obesity25 reports
- OBALON® Balloon Kit HPMC 7600-0001Medtimo, Inc · Implant, Intragastric For Morbid Obesity20 reports
- LAP-BAND AP Adjustable Gastric Banding System B-2260Apollo Endosurgery, Inc. · Implant, Intragastric For Morbid Obesity18 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page