X-Tack Endoscopic HeliX Tacking System CNH-C01-213-LRecalls & reported problems

Apollo Endosurgery, Inc. · Suture Cinch - Long · Class 2

All models of X-Tack Endoscopic HeliX Tacking System (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10811955020756), received between Jan 14, 2022 and Aug 20, 2026 — classified as 2 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure7
  • Failure to Cut2
  • Insufficient Device Problem Information2
  • No Apparent Adverse Event2
  • Break1
  • Material Integrity Problem1
  • Therapeutic or Diagnostic Output Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
  • Insufficient Information3

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 20, 2026Activation Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50351119 ↗
  • InjuryJun 9, 2026Break; Failure to Cut; Activation Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SUTURE CINCH - LONG WAS USED DURING AN ENDOSCOPIC MUCOSAL RESECTION (EMR) PROCEDURE ON (B)(6) 2026. DURING THE PROCEDURE, THE CINCH DID NOT CUT THE SUTURE OR DEPLOY, AND THE HANDLE BROKE. THE CINCH WAS MANUALLY DEPLOYED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25444836 ↗
  • MalfunctionFeb 19, 2026Activation Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SUTURE CINCH WAS USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE ON (B)(6) 2026. DURING THE PROCEDURE AND AFTER THE SUTURE WAS CUT, THE CINCH FAILED TO DEPLOY. THE PROCEDURE WAS COMPLETED WITH A NEW SUTURE CINCH. THERE WAS NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24385524 ↗
  • MalfunctionJul 18, 2025Activation Failure; Activation Failure

    IT WAS RETURNED TO BOSTON SCIENTIFIC THAT TWO SUTURING CINCH WAS USED IN A FISTULA CLOSURE PROCEDURE ON (B)(6) 2025. DURING THE PROCEDURE, THE PHYSICIAN FIRST USED ARGON TO CREATE A LESION SIMILAR TO AN ULCER AND THEN USED THE X-TACK DEVICE TO CLOSE THE LESION. THE X-TACK WORKED FINE, BUT WHEN THE PHYSICIAN USED THE SUTURING CINCH TO CUT THE SUTURE, THE CINCH DID NOT KNOT THE SUTURE, AND THE SUTURE CAME LOOSE FROM THE X-TACK. ADDITIONAL INFORMATION RECEIVED JUNE 23, 2025 CONFIRMED THE CINCH DID NOT PERFORM THE MECHANISM IT NORMALLY DOES TO PREVENT THE SUTURE FROM BECOMING LOOSE. THIS OCCURRED WITH TWO SUTURING CINCHES. THIS RECORD REPRESENTS THE FIRST OF TWO SUTURING CINCHES. THE PROCEDURE WAS COMPLETED WITH ANOTHER SUTURING CINCH. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22543698 ↗
  • MalfunctionJul 18, 2025Activation Failure; Activation Failure

    IT WAS RETURNED TO BOSTON SCIENTIFIC THAT TWO SUTURING CINCH WAS USED IN A FISTULA CLOSURE PROCEDURE ON (B)(6) 2025. DURING THE PROCEDURE, THE PHYSICIAN FIRST USED ARGON TO CREATE A LESION SIMILAR TO AN ULCER AND THEN USED THE X-TACK DEVICE TO CLOSE THE LESION. THE X-TACK WORKED FINE, BUT WHEN THE PHYSICIAN USED THE SUTURING CINCH TO CUT THE SUTURE, THE CINCH DID NOT KNOT THE SUTURE, AND THE SUTURE CAME LOOSE FROM THE X-TACK. ADDITIONAL INFORMATION RECEIVED JUNE 23, 2025, CONFIRMED THE CINCH DID NOT PERFORM THE MECHANISM IT NORMALLY DOES TO PREVENT THE SUTURE FROM BECOMING LOOSE. THIS OCCURRED WITH TWO SUTURING CINCHES. THIS RECORD REPRESENTS THE FIRST OF TWO SUTURING CINCHES. THE PROCEDURE WAS COMPLETED WITH ANOTHER SUTURING CINCH. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22543636 ↗
  • MalfunctionJul 8, 2025Failure to Cut; Activation Failure; Failure to Cut; Activation Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC THAT A SUTURING CINCH WAS USED IN THE ESOPHAGUS DURING A FISTULA CLOSURE PROCEDURE ON (B)(6) 2025. DURING THE PROCEDURE, THE CINCH DID NOT RELEASE CORRECTLY AS IT DID NOT CUT THE THREAD. THE PROCEDURE WAS COMPLETED WITH ANOTHER SUTURING CINCH. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22436303 ↗
  • MalfunctionApr 8, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC THAT A SUTURING CINCH WAS USED IN AN ESD PROCEDURE ON (B)(6) 2025. WHEN DEPLOYING A SUTURING CINCH, THERE WAS A RESISTANCE, AND IT WAS IMPOSSIBLE TO USE THAT CINCH BECAUSE THE PHYSICIAN FELT RESISTANCE AND FINALLY BROKE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT. PROCEDURE COMPLETED WITH AN ALTERNATIVE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21796664 ↗
  • MalfunctionNov 3, 2023Material Integrity Problem; Material Integrity Problem

    HEALTHCARE PROFESSIONAL REPORTED DURING LOADING OF THE SUTURE AND WHEN THE GOLD TAB WAS PULLED TO THREAD THE SUTURE, THE CINCH ALSO MOVED INTO THE CATHETER. DEVICE NOT ADMINISTERED TO THE PATIENT. THE CASE CONTINUED WITH A NEW CINCH. PATIENT WAS LATER ADMITTED TO ICU DUE TO A NON-DISCLOSED REASON HOWEVER UNRELATED TO THE DEVICE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information

    MAUDE report 18068179 ↗
  • MalfunctionAug 25, 2023No Apparent Adverse Event; Insufficient Device Problem Information; No Apparent Adverse Event; Insufficient Device Problem Information

    CINCH CUT THE SUTURE BUT DID NOT DISENGAGE FROM THE DEVICE. METAL WAS PULLED TO DISLODGE RESULTING IN REMOVAL OF THE SUTURES. PROCEDURE SUCCESSFULLY COMPLETED WITH COMPETITIVE DEVICE. PATIENT IS DOING WELL.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information

    MAUDE report 17622948 ↗
  • MalfunctionMay 12, 2023No Apparent Adverse Event; Insufficient Device Problem Information; No Apparent Adverse Event; Insufficient Device Problem Information

    HEALTHCARE PROVIDER CINCH SUTURE BROKE DURING THE CINCHING PROCESS. PROCEDURE SUCCESSFULLY COMPLETED WITH COMPETITIVE DEVICE. PATIENT IS DOING WELL.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information

    MAUDE report 16918417 ↗
  • MalfunctionJan 14, 2022Therapeutic or Diagnostic Output Failure

    APOLLO SUTURE CINCH- LONG FAILED UPON OPENING THE DEVICE - REP DISCOVERED UPON OPENING, IT WAS NOT USED ON THE PATIENT. ANOTHER OF THE SAME DEVICE WAS USED INSTEAD TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13257715 ↗

Other Hemostatic Metal Clip For The Gi Tract devices

All Hemostatic Metal Clip For The Gi Tract devices →