CORFLO Enteral Feeding Tube 42-9366Recalls & reported problems
Corpak Medsystems, Inc. · Feeding tube with ENFit connectors non weighted with stylet · Class 2
All models of CORFLO Enteral Feeding Tube (14)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10815149020669), received between Apr 21, 2017 and Aug 14, 2019 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Crack” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Crack3
- Break2
- Obstruction of Flow1
Patient problems
- No Known Impact Or Consequence To Patient4
- Foreign Body In Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 14, 2019Break; Obstruction of Flow; Break; Obstruction of Flow
IT WAS REPORTED THAT "THE TUBE WAS PLACED IN A PEDIATRIC PATIENT ON (B)(6) [2019]. ON (B)(6)[2019], THE TUBE WAS REMOVED AND THE DOCTOR REALIZED THAT ONLY A PART OF THE TUBE WAS REMOVED. AFTER CHECK IT BY RX [RADIOGRAPHY] THEY SAW THAT A PART OF THE TUBE WAS STILL ON JEJUNUM (13 CM APPROX.). A SURGICAL INTERVENTION WAS NEEDED TO REMOVE THE BROKEN PART. BETWEEN (B)(6) [2019] AND (B)(6) [2019], THE TUBE WAS OBTURED [OBSTRUCTED] AND THEY IRRIGATED THE TUBE WITH 5CC OF WATER, BUT THEY DID NOT PERCEIVE THE BROKEN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8895672 ↗MalfunctionOct 1, 2018Break; Break
IT WAS REPORTED THAT UPON REMOVAL OF THE NASO-JEJUNAL (NJ) TUBE FROM A PATIENT IN PEDIATRIC INTENSIVE CARE, IT WAS NOTICED THAT THE TUBE HAD BROKEN, AND PART OF THE TUBE REMAINED INSIDE THE PATIENT. A PROCEDURE WAS PERFORMED TO REMOVE THE BROKEN PIECE OF TUBE, AND TO RE-PASS A NEW NJ TUBE. NO FURTHER INFORMATION PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 7923858 ↗MalfunctionApr 21, 2017Crack; Crack
HALYARD RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENTS, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE THIRD OF THREE REPORTS. REFER TO 3009124963-2017-00014 FOR THE FIRST PATIENT. REFER TO 3009124963-2017-00015 FOR THE SECOND PATIENT. IT WAS REPORTED THAT A PATIENT EXPERIENCED AN NJ TUBE THAT DETERIORATED. IT WAS PICKED UP ON X-RAY THAT THE NJ TUBE SPLIT AT THE TOP OF THE TUBE NEAR THE PATIENT'S TRACHEA. THE PATIENT WAS REQUIRED TO GO TO THE OPERATING ROOM TO HAVE THE TUBE REMOVED. NO FURTHER INFORMATION WAS RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6512917 ↗MalfunctionApr 21, 2017Crack; Crack
HALYARD RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENTS, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE SECOND OF THREE REPORTS. REFER TO 3009124963-2017-00014 FOR THE FIRST PATIENT. REFER TO 3009124963-2017-00016 FOR THE THIRD PATIENT. IT WAS REPORTED THAT A PATIENT EXPERIENCED A NJ TUBE THAT DETERIORATED. IT WAS PICKED UP ON X-RAY THAT THE NJ TUBE SPLIT AT THE TOP OF THE TUBE NEAR THE PATIENT'S TRACHEA. THE PATIENT WAS REQUIRED TO GO TO THE OPERATING ROOM TO HAVE THE TUBE REMOVED. NO FURTHER INFORMATION WAS RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6512844 ↗MalfunctionApr 21, 2017Crack; Crack
HALYARD RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENTS, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE FIRST OF THREE REPORTS. REFER TO 3009124963-2017-00015 FOR THE SECOND PATIENT. REFER TO 3009124963-2017-00016 FOR THE THIRD PATIENT. IT WAS REPORTED THAT A PATIENT EXPERIENCED AN NJ TUBE THAT DETERIORATED. IT WAS PICKED UP ON X-RAY THAT THE NJ TUBE SPLIT AT THE TOP OF THE TUBE NEAR THE PATIENT'S TRACHEA. THE PATIENT WAS REQUIRED TO GO TO THE OPERATING ROOM TO HAVE THE TUBE REMOVED. NO FURTHER INFORMATION WAS RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6512719 ↗
Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-05090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page