CORTRAK 2 EAS Electromagnetic Transmitting Stylet and CORFLO Feeding Tube 40-9551TRAK2Recalls & reported problems

Corpak Medsystems, Inc. · Feeding tube non weighted with stylet for use with CORTRAK 2 Enteral Access System · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10815149020911), received between Mar 6, 2019 and Jun 5, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove2
  • Break1

Patient problems

  • Low Oxygen Saturation2
  • Foreign Body In Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 5, 2025Break; Break

    IT WAS REPORTED, ¿WE HAD AN INCIDENT TODAY WHERE ONE OF OUR PATIENTS WITH A CORTRAK TUBE HAD HALF OF THE TUBE COME OUT OF HIS NOSE (THE BRIDLE WAS UNCLIPPED FOR REPOSITIONING) WHILE THE OTHER HALF REMAINED IN HIS STOMACH.¿ PER ADDITIONAL INFORMATION RECEIVED ON 23MAY2025, THE TUBE WAS IN PLACE FOR THREE WEEKS (PLACED ON 22APR2025). THE TUBE BREAK WAS FIRST IDENTIFIED ON 13MAY2025 VIA X-RAY. THERE WAS NO INJURY TO THE PATIENT. PATIENT REQUIRED ESOPHAGOGASTRODUODENOSCOPY TO RETRIEVE PORTION OF TUBE (APPROX. 100CM). THERE WAS A CLOG NOTED IN THE TUBE AND CLEARED ON 5MAY2025. NO PERMANENT HARM TO PATIENT. HE IS STILL ADMITTED BUT NOT FOR REASONS RELATED TO THIS INCIDENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 22150572 ↗
  • MalfunctionMar 6, 2019Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT THE CUSTOMER HAD DIFFICULTY REMOVING THE GUIDEWIRE FROM THE TUBE UPON INITIAL REPLACEMENT. THIS LED TO THE PATIENT DESATURATING. PER ADDITIONAL INFORMATION RECEIVED 27 FEB 2019, THE ISSUE WITH THE GUIDEWIRE PROLONGED THE PROCEDURE, AND THIS, ALONG WITH THE PATIENT'S UNDERLYING CONDITION, LED TO DESATURATION. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Low Oxygen Saturation

    MAUDE report 8397604 ↗
  • MalfunctionMar 6, 2019Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT THE CUSTOMER HAD DIFFICULTY REMOVING THE GUIDEWIRE FROM THE TUBE UPON INITIAL REPLACEMENT. THIS LED TO THE PATIENT DESATURATING. PER ADDITIONAL INFORMATION RECEIVED 27 FEB 2019, THE ISSUE WITH THE GUIDEWIRE PROLONGED THE PROCEDURE, AND THIS, ALONG WITH THE PATIENT'S UNDERLYING CONDITION, LED TO DESATURATION. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Low Oxygen Saturation

    MAUDE report 8394276 ↗