CORTRAK 2 EAS Electromagnetic Transmitting Stylet and CORFLO Feeding Tube 40-9551TRAK2Recalls & reported problems
Corpak Medsystems, Inc. · Feeding tube non weighted with stylet for use with CORTRAK 2 Enteral Access System · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10815149020911), received between Mar 6, 2019 and Jun 5, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Remove2
- Break1
Patient problems
- Low Oxygen Saturation2
- Foreign Body In Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 5, 2025Break; Break
IT WAS REPORTED, ¿WE HAD AN INCIDENT TODAY WHERE ONE OF OUR PATIENTS WITH A CORTRAK TUBE HAD HALF OF THE TUBE COME OUT OF HIS NOSE (THE BRIDLE WAS UNCLIPPED FOR REPOSITIONING) WHILE THE OTHER HALF REMAINED IN HIS STOMACH.¿ PER ADDITIONAL INFORMATION RECEIVED ON 23MAY2025, THE TUBE WAS IN PLACE FOR THREE WEEKS (PLACED ON 22APR2025). THE TUBE BREAK WAS FIRST IDENTIFIED ON 13MAY2025 VIA X-RAY. THERE WAS NO INJURY TO THE PATIENT. PATIENT REQUIRED ESOPHAGOGASTRODUODENOSCOPY TO RETRIEVE PORTION OF TUBE (APPROX. 100CM). THERE WAS A CLOG NOTED IN THE TUBE AND CLEARED ON 5MAY2025. NO PERMANENT HARM TO PATIENT. HE IS STILL ADMITTED BUT NOT FOR REASONS RELATED TO THIS INCIDENT.
Patient problems: Foreign Body In Patient
MAUDE report 22150572 ↗MalfunctionMar 6, 2019Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT THE CUSTOMER HAD DIFFICULTY REMOVING THE GUIDEWIRE FROM THE TUBE UPON INITIAL REPLACEMENT. THIS LED TO THE PATIENT DESATURATING. PER ADDITIONAL INFORMATION RECEIVED 27 FEB 2019, THE ISSUE WITH THE GUIDEWIRE PROLONGED THE PROCEDURE, AND THIS, ALONG WITH THE PATIENT'S UNDERLYING CONDITION, LED TO DESATURATION. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Low Oxygen Saturation
MAUDE report 8397604 ↗MalfunctionMar 6, 2019Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT THE CUSTOMER HAD DIFFICULTY REMOVING THE GUIDEWIRE FROM THE TUBE UPON INITIAL REPLACEMENT. THIS LED TO THE PATIENT DESATURATING. PER ADDITIONAL INFORMATION RECEIVED 27 FEB 2019, THE ISSUE WITH THE GUIDEWIRE PROLONGED THE PROCEDURE, AND THIS, ALONG WITH THE PATIENT'S UNDERLYING CONDITION, LED TO DESATURATION. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Low Oxygen Saturation
MAUDE report 8394276 ↗
Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-05090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page