Clearview® 92002Recalls & reported problems
Alere San Diego, Inc. · Clearview™ Strep A Exact II Dipstick · Class 1
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10815845020024), received between Jun 1, 2018 and Jun 1, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “False Negative Result” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jul 30, 2025Open, Classified
Due to a classification error in the inventory management system, specific medical devices were shipped to customers that did not have the requisite license to purchase such devices.
Z-2337-2025 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- False Negative Result1
Patient problems
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 1, 2018False Negative Result; False Negative Result
ON (B)(6) 2018, THE PATIENT PRESENTED AT THE URGENT CARE FACILITY WITH A SORE THROAT. UPON THE INITIAL EXAMINATION, THE CUSTOMER NOTED THAT THE PATIENT'S THROAT ONLY APPEARED A LITTLE RED AND IT WAS NOT IMMEDIATELY EVIDENT THE PATIENT HAD STREP A. A TEST WAS PERFORMED WITH A CLEARVIEW STREP A EXACT II DIPSTICK TEST USING A DUAL SWAB TECHNIQUE AND A NEGATIVE RESULT WAS OBTAINED AT THE 5 MINUTE READ TIME. THE SECOND THROAT SWAB WAS SENT TO (B)(6) LABORATORY FOR CULTURE CONFIRMATION. THE PATIENT WAS DIAGNOSED WITH PHARYNGITIS AND SENT HOME. TWO HOURS AFTER TEST WAS PERFORMED ON THE CLEARVIEW STREP A EXACT II DIPSTICK TEST, THE OPERATOR NOTICED A FAINT POSITIVE LINE HAD DEVELOPED IN THE TEST REGION AND FORWARDED THIS INFORMATION TO THE CUSTOMER. ON (B)(6) 2018, THE CUSTOMER CONTACTED THE PATIENT TO INFORM THE PATIENT OF THE POSITIVE RESULT AND PRESCRIBE AMOXICILLIN. AT THIS TIME, THE THROAT CULTURE RESULTS HAD NOT BEEN RECEIVED. ON (B)(6) 2018, THE CUSTOMER OBTAINED A POSITIVE RESULT FROM THE CONFIRMATORY THROAT CULTURE. THE REPORT INDICATED THAT THERE WERE A FEW COLONIES OF STREP A ON THE CULTURE. THE CULTURE WAS PERFORMED ON STREP A SPECIFIC BACITRACIN DISK. THE CUSTOMER PROCEEDED TO
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 7563182 ↗
Other Antigens, All Groups, Streptococcus Spp. devices
- BD Veritor™ System for Rapid Detection of Group A Strep CLIA-Waived Kit 256040Becton, Dickinson And Company · ANTIGENS, ALL GROUPS, STREPTOCOCCUS SPP.258 reports
- Acceava® 92001Alere San Diego, Inc. · Antigens, All Groups, Streptococcus Spp.2 recalls2 reports
- QuickVue 20108Quidel Corporation · ANTIGENS, ALL GROUPS, STREPTOCOCCUS SPP.2 recalls0 reports
- Sofia 20274Quidel Corporation · ANTIGENS, ALL GROUPS, STREPTOCOCCUS SPP.1 recall2 reports
- McKesson 5003Mckesson Medical-Surgical Inc. · Antigens, All Groups, Streptococcus Spp.1 recall0 reports
- McKesson 4999Mckesson Medical-Surgical Inc. · Antigens, All Groups, Streptococcus Spp.1 recall0 reports
- QuickVue In-Line 00345Quidel Corporation · ANTIGENS, ALL GROUPS, STREPTOCOCCUS SPP.1 recall0 reports
- QuickVue 20125Quidel Corporation · ANTIGENS, ALL GROUPS, STREPTOCOCCUS SPP.1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.