Skip to content

6ml Syringe to SCD Tube Accessory Set RF-T15Recalls & reported problems

Origen Biomedical, Inc. · 6ml syringe on a female luer with 15cm sealed SCD tubing · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10816203020205), received between Dec 8, 2023 and Jan 3, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Discolored” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 19, 2023Class IIOpen, Classified

    Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.

    Z-1068-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Discolored2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 3, 2024Material Discolored; Material Discolored

    CUSTOMER REPORTED THAT A 6 ML SYRINGE (PART #: RF-T15) WAS RECEIVED DISCOLORED (YELLOW). THIS PRODUCT WAS A REPLACEMENT FOR PREVIOUS COMPLAINED PRODUCT THAT HAD THE SAME ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18436256 ↗
  • MalfunctionDec 8, 2023Material Discolored; Material Discolored

    CUSTOMER REPORTED THAT A 6ML SYRINGE (PART #: RF-T15) WAS RECEIVED DISCOLORED (YELLOW) AND WITH LOOSE PLUNGERS. THE ISSUE WAS DISCOVERED AT INCOMING INSPECTION AND DID NOT INVOLVE A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18291823 ↗

Other Tube, Tissue Culture devices

All Tube, Tissue Culture devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page