Skip to content

Universa G23417Recalls & reported problems

Cook Incorporated · Universa Firm Ureteral Stent and Positioner With Hydrophilic Coating and Monofilament Tether · Class 2 · Implant

All models of Universa (82)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002234175), received between Jun 25, 2019 and Jun 25, 2019 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component3

Patient problems

  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 25, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS ORIGINALLY REPORTED DURING KIDNEY/URETER STENT PLACEMENT USING A UNIVERSA FIRM URETERAL STENT SET, THE "STRING LOOSENS AT THE SLIGHTEST TOUCH." THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS AS A RESULT OF THIS ALLEGED PRODUCT MALFUNCTION. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES AS A RESULT OF THIS OCCURRENCE. NO UNINTENDED SECTION OF THE DEVICE REMAINED INSIDE THE PATIENT'S BODY. PHYSICAL EVALUATION OF THE RETURNED COMPLAINT DEVICE REVEALED THAT THE TETHER ON THE DEVICE IS BROKEN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8732598 ↗
  • MalfunctionJun 25, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    ADDITIONAL INFORMATION WAS RECEIVED ON 25JUN2019: IT WAS REPORTED, THE TETHERS ARE BREAKING AT THE KNOTS BY THE SLIGHTEST TOUCH. THE STENTS WERE TESTED BY THE ACCOUNT AND FOUND THAT THE TETHER BREAKS EASILY AND THESE TETHERS WERE NOT USED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8732178 ↗
  • MalfunctionJun 25, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    ADDITIONAL INFORMATION WAS RECEIVED ON 25JUN2019: IT WAS REPORTED, THE TETHERS ARE BREAKING AT THE KNOTS BY THE SLIGHTEST TOUCH. THE STENTS WERE TESTED BY THE ACCOUNT AND FOUND THAT THE TETHER BREAKS EASILY AND THESE TETHERS WERE NOT USED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8731626 ↗

Other STENT, URETERAL devices

All STENT, URETERAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page