Cantata G23805Recalls & reported problems
Cook Incorporated · Cantata, 2.8 Duo Microcatheter · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10827002238050), received between Jun 15, 2017 and Nov 11, 2020 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Device Contamination with Chemical or Other Material1
- Material Perforation1
- Material Separation1
Patient problems
- No Consequences Or Impact To Patient2
- Device Embedded In Tissue or Plaque1
- No Code Available1
- No Patient Involvement1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 11, 2020Material Separation; Material Separation
IT WAS REPORTED A MALE PATIENT REQUIRED PLACEMENT OF A CANTATA DUO 2.8 SUPERSELECTIVE MICROCATHETER FOR A SPLENIC EMBOLIZATION PROCEDURE. AFTER THE COILS WERE PLACED, THE OPERATOR NOTED THAT THE CATHETER BROKE INSIDE OF THE PATIENT, ABOUT 10CM FROM THE DISTAL END. THE OPERATOR RETRIEVED THE SEPARATED PORTION WITH A SNARE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO OTHER ADVERSE EFFECTS WERE REPORTED FOR THIS INCIDENT.
Patient problems: Device Embedded In Tissue or Plaque; No Code Available
MAUDE report 10825513 ↗MalfunctionSep 11, 2018Material Perforation; Material Perforation
IT WAS REPORTED A CANTATA DUO 2.8 SUPERSELECTIVE MICROCATHETER WAS PLACED FOR A UTERINE FIBROID EMBOLIZATION IN A PATIENT WITH A TORTUOUS ANATOMY. DURING THE PROCEDURE, THE PHYSICIAN REMOVED THE CATHETER FOR ASSESSMENT. WHILE FLUSHING WITH SALINE AND CONTRAST SOLUTION WITH A 1 ML SYRINGE, IT WAS OBSERVED THAT A HOLE HAD FORMED IN THE PROXIMAL THIRD OF THE CATHETER. THE PROCEDURE CONTINUED WITH ANOTHER LIKE CATHETER WHICH ADDED TO THE PROCEDURE TIME, BUT DID NOT RESULT IN ANY ADVERSE EFFECTS TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7865119 ↗MalfunctionAug 14, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
FOREIGN MATERIAL WAS OBSERVED FREE-FLOATING WITHIN THE DEVICE PACKAGE. THE FOREIGN MATTER WAS OBSERVED DURING AN EVALUATION OF UNUSED SAMPLES RETURNED FOR A SEPARATE COMPLAINT INVESTIGATION. THERE WAS NO PATIENT CONTACT, ACCORDINGLY, NO ADVERSE PATIENT CONSEQUENCE. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
Patient problems: No Patient Involvement
MAUDE report 6792669 ↗MalfunctionJun 15, 2017Detachment of Device or Device Component; Detachment of Device or Device Component
CUSTOMER REPORTED THAT A PATIENT UNDERWENT SUCCESSFUL PLACEMENT OF A CANTATA DUO 2.8 SUPERSELECTIVE MICROCATHETER FOR AN UNSPECIFIED INDICATION. THE RUBBER PART OF THE CATHETER SEPARATED FROM THE HUB AT THE TIME OF MEDICINAL INJECTION. THE CIRCUMSTANCES AND HANDLING CONDITIONS AT THE TIME OF FLUID INJECTION ARE NOT KNOWN. THE PHYSICIAN RECONNECTED THE RUBBER PART TO THE HUB AND PROCEEDED WITH UNEVENTFUL FLUID INJECTION. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. THE DEVICE IS NOT AVAILABLE FOR EVALUATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6643273 ↗
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.