Skip to content

Cantata G23805Recalls & reported problems

Cook Incorporated · Cantata, 2.8 Duo Microcatheter · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10827002238050), received between Jun 15, 2017 and Nov 11, 2020 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Device Contamination with Chemical or Other Material1
  • Material Perforation1
  • Material Separation1

Patient problems

  • No Consequences Or Impact To Patient2
  • Device Embedded In Tissue or Plaque1
  • No Code Available1
  • No Patient Involvement1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 11, 2020Material Separation; Material Separation

    IT WAS REPORTED A MALE PATIENT REQUIRED PLACEMENT OF A CANTATA DUO 2.8 SUPERSELECTIVE MICROCATHETER FOR A SPLENIC EMBOLIZATION PROCEDURE. AFTER THE COILS WERE PLACED, THE OPERATOR NOTED THAT THE CATHETER BROKE INSIDE OF THE PATIENT, ABOUT 10CM FROM THE DISTAL END. THE OPERATOR RETRIEVED THE SEPARATED PORTION WITH A SNARE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO OTHER ADVERSE EFFECTS WERE REPORTED FOR THIS INCIDENT.

    Patient problems: Device Embedded In Tissue or Plaque; No Code Available

    MAUDE report 10825513 ↗
  • MalfunctionSep 11, 2018Material Perforation; Material Perforation

    IT WAS REPORTED A CANTATA DUO 2.8 SUPERSELECTIVE MICROCATHETER WAS PLACED FOR A UTERINE FIBROID EMBOLIZATION IN A PATIENT WITH A TORTUOUS ANATOMY. DURING THE PROCEDURE, THE PHYSICIAN REMOVED THE CATHETER FOR ASSESSMENT. WHILE FLUSHING WITH SALINE AND CONTRAST SOLUTION WITH A 1 ML SYRINGE, IT WAS OBSERVED THAT A HOLE HAD FORMED IN THE PROXIMAL THIRD OF THE CATHETER. THE PROCEDURE CONTINUED WITH ANOTHER LIKE CATHETER WHICH ADDED TO THE PROCEDURE TIME, BUT DID NOT RESULT IN ANY ADVERSE EFFECTS TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7865119 ↗
  • MalfunctionAug 14, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    FOREIGN MATERIAL WAS OBSERVED FREE-FLOATING WITHIN THE DEVICE PACKAGE. THE FOREIGN MATTER WAS OBSERVED DURING AN EVALUATION OF UNUSED SAMPLES RETURNED FOR A SEPARATE COMPLAINT INVESTIGATION. THERE WAS NO PATIENT CONTACT, ACCORDINGLY, NO ADVERSE PATIENT CONSEQUENCE. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

    Patient problems: No Patient Involvement

    MAUDE report 6792669 ↗
  • MalfunctionJun 15, 2017Detachment of Device or Device Component; Detachment of Device or Device Component

    CUSTOMER REPORTED THAT A PATIENT UNDERWENT SUCCESSFUL PLACEMENT OF A CANTATA DUO 2.8 SUPERSELECTIVE MICROCATHETER FOR AN UNSPECIFIED INDICATION. THE RUBBER PART OF THE CATHETER SEPARATED FROM THE HUB AT THE TIME OF MEDICINAL INJECTION. THE CIRCUMSTANCES AND HANDLING CONDITIONS AT THE TIME OF FLUID INJECTION ARE NOT KNOWN. THE PHYSICIAN RECONNECTED THE RUBBER PART TO THE HUB AND PROCEEDED WITH UNEVENTFUL FLUID INJECTION. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. THE DEVICE IS NOT AVAILABLE FOR EVALUATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6643273 ↗

FDA clearance and classification

The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Cantata G23805

Has the Cantata G23805 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10827002238050), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Cantata G23805?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jun 15, 2017 and Nov 11, 2020. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (1), “Device Contamination with Chemical or Other Material” (1) and “Material Perforation” (1); the most common patient problem is “Device Embedded In Tissue or Plaque” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Cantata G23805?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Cantata G23805, and is it still sold?

The FDA's device database lists Cook Incorporated as the company responsible for it (brand Cantata, model G23805). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, continuous flush devices

All Catheter, continuous flush devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page