Zenith G32535Recalls & reported problems
William A. Cook Australia Pty. Ltd. · Zenith Fenestrated, AAA Endovascular Graft, Proximal Body · Class 3 · Implant
All models of Zenith (32)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002325354), received between Jun 21, 2023 and Jul 11, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Component or Accessory Incompatibility” (1 report).
111 other recalls cover devices cleared under the same FDA submission (P020018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Component or Accessory Incompatibility1
Patient problems
- Thrombosis/Thrombus2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THIS (B)(4): LEFT RENAL STENT OCCLUDED, INFARCTED KIDNEY, PATIENT ON DIALYSIS. MENTIONED THAT THE STENT USED FOR RENAL ARTERY WAS 5MM STENT (NON-COOK). THE PHYSICIAN MENTIONED THAT THE ONLY ZFEN RENAL OCCLUSIONS HE HAS SEEN HAVE OCCURRED WHEN 5MM STENT WAS USED IN THE RENAL. SEE (B)(4) FOR: 2 OTHER PATIENTS RECEIVING 5MM STENT HAVE OCCLUDED. (TOTALLING 3 OUT OF 4 PATIENTS WHO RECEIVED A 5MM STENT).
Patient problems: Thrombosis/Thrombus
MAUDE report 17298355 ↗InjuryJun 21, 2023Component or Accessory Incompatibility; Adverse Event Without Identified Device or Use Problem; Component or Accessory Incompatibility; Adverse Event Without Identified Device or Use Problem
THIS (B)(4): LEFT RENAL STENT OCCLUDED, INFARCTED KIDNEY, PATIENT ON DIALYSIS. MENTIONED THAT THE STENT USED FOR RENAL ARTERY WAS 5MM STENT (NON-COOK). THE PHYSICIAN MENTIONED THAT THE ONLY ZFEN RENAL OCCLUSIONS HE HAS SEEN HAVE OCCURRED WHEN 5MM STENT WAS USED IN THE RENAL. SEE (B)(4) FOR: 2 OTHER PATIENTS RECEIVING 5MM STENT HAVE OCCLUDED. (TOTALLING 3 OUT OF 4 PATIENTS WHO RECEIVED A 5MM STENT).
Patient problems: Thrombosis/Thrombus
MAUDE report 17168304 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020018. One submission often covers many models, so check the product and lot details in each record.
- Z-3146-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-18-73-ZT, REF G48466, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3153-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-56-ZT, REF G48477, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3152-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-73-ZT, REF G48476, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3151-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-22-73-ZT, REF G48474, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3130-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-24-82-ZT, REF G48407, 18 Fr ID/ 7.1 mm OD, Sterile
- Z-3145-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-18-56-ZT, REF G48465, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3132-2011 · Apr 15, 2011 · TerminatedCook Medical, Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-26-96-ZT, REF G48413, 18 Fr ID/ 7.1 mm OD, Sterile.
- Z-3138-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-36-95-ZT, REF G48432, 22 Fr ID/ 8.5 mm OD, Sterile.
- Z-3147-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-56-ZT, REF G48469, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3148-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-73-ZT, REF G48470, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3156-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Main Body Extension, REF ESBE-22-29, REF G32179, 18 Fr. ID/ 6.90 OD, Sterile.
- Z-3141-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-14-56-ZT, REF G48455, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3134-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-28-96-ZT, REF G48418, 20 Fr ID/ 7.7 mm OD, Sterile
- Z-3139-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-10-71-ZT, REF G48444, 14 Fr ID/ 5.3 mm OD, Sterile.
- Z-3144-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-16-90-ZT, REF G48463, 16Fr ID/ 6.0 mm OD, Sterile.
…and 96 more.
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page