Skip to content

Zenith G32536Recalls & reported problems

William A. Cook Australia Pty. Ltd. · Zenith Fenestrated, AAA Endovascular Graft, Proximal Body · Class 3 · Implant

All models of Zenith (32)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002325361), received between Aug 6, 2017 and Mar 15, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Coagulation in Device or Device Ingredient” (1 report).

111 other recalls cover devices cleared under the same FDA submission (P020018); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Coagulation in Device or Device Ingredient1
  • Material Twisted/Bent1
  • Occlusion Within Device1
  • Patient Device Interaction Problem1

Patient problems

  • Occlusion2
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1
  • Urinary Retention1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 15, 2026Material Twisted/Bent; Patient Device Interaction Problem

    AN 81-YEAR-OLD FEMALE PATIENT UNDERWENT A ZFEN FOR AAA PROCEDURE IN WHICH THE ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT PROXIMAL BODY, G32536, WAS USED. THE GRAFT WAS INSERTED INTO TORTUOUS ANATOMY AND WAS ROTATED (ROTATING THE GRAY POSITIONER AND SHEATH IN UNISON) TO CORRECT POSITION. THE GRAFT WAS UNSHEATHED AND CANDY-WRAPPING WAS NOTICED. THE GRAFT WAS THEN ROTATED IN THE OPPOSITE DIRECTION TO "UNWRAP". RENAL ARTERIES WERE CANNULATED AND ANSEL SHEATHS WERE ADVANCED INTO POSITION. WHEN THE GOLD TRIGGER WIRE WAS PULLED AND RELEASED, THE PHYSICIAN NOTICED SOME TENSION. THE BLACK TRIGGER WIRE WAS RELEASED AND PULLED WITH SOME TENSION AS WELL. THE PIN VISE WAS LOOSENED, THE TOP CAP WAS ADVANCED, AND THE PIN VISE TIGHTENED. THE WHITE TRIGGER WIRE WAS PULLED, AND THE WIRE BROKE OFF, FLUSH WITH THE TRIGGER MECHANISM. ADVANCED TROUBLE SHOOTING TECHNIQUES WERE FOLLOWED TO EXPOSE THE TRIGGER WIRE. THE TRIGGER WIRE WAS PULLED AND EXTREMELY TIGHT. EVENTUALLY, THE WIRE WAS REMOVED AND THE CASE WAS COMPLETED. ADDITIONAL INFORMATION WAS RECEIVED: THE DEVICE WAS TWISTED DUE TO THE TORTUOUS ILIAC OF THE PATIENT. THERE WAS SOME EVIDENCE TO SUGGEST THE GRAFT MAY HAVE BEEN TWISTED PRIOR TO INSERTION

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24602063 ↗
  • InjuryApr 2, 2019Coagulation in Device or Device Ingredient

    THE ORIGINAL IMPLANT ON (B)(6) 2019 WAS SUCCESSFUL. THE FOLLOW UP CT SHOWED NO ISSUES, ALL VESSELS WERE NOTED TO BE PATENT. THE PATIENT WAS RELEASED FROM HOSPITAL. 19 DAYS POST PROCEDURE, ON (B)(6) 2019 THE PATIENT PRESENTED WITH THROMBOSED LEFT RENAL AND SUPERIOR MESENTERIC ARTERY (SMA) STENTS. THE DOCTOR PERFORMED AN EMERGENT SUPERIOR MESENTERIC ARTERY (SMA) BYPASS.

    Patient problems: Occlusion

    MAUDE report 8475693 ↗
  • InjuryAug 6, 2017Occlusion Within Device

    THE PATIENT'S FENESTRATED GRAFT PROCEDURE HAPPENED ON THE (B)(6) 2017. THE PATIENT RECEIVED THE DEVICE AND 2 ADDITIONAL I-CAST STENTS (NOT MANUFACTURED BY COOK MEDICAL). IN THE MIDDLE OF THE AFTERNOON ON (B)(6) 2017 THE PATIENT STARTED COMPLAINING OF BACK PAIN AND HAD NO URINE PRODUCTION (ANURIC). THE PATIENT WAS TAKEN TO THE OPERATING ROOM IN THE AFTERNOON OF (B)(6) FOR AN ANGIOGRAM. THE DOCTOR DID ANGIOS OF THE ABDOMINAL AORTA AND SELECTIVE ANGIOS OF EACH RENAL ARTERY. BOTH RENAL ARTERIES WERE FOUND TO BE OCCLUDED AT THE DISTAL END OF THE RENAL I-CAST STENTS. THROMBECTOMY & PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY WAS PERFORMED TO RESTORE BLOOD FLOW IN THE RENAL ARTERIES.

    Patient problems: Occlusion; Pain; Urinary Retention

    MAUDE report 6769841 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P020018. One submission often covers many models, so check the product and lot details in each record.

  • Z-3152-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-73-ZT, REF G48476, 16Fr ID/ 6.0 mm OD, Sterile.
  • Z-3144-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-16-90-ZT, REF G48463, 16Fr ID/ 6.0 mm OD, Sterile.
  • Z-3135-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-30-82-ZT, REF G48422, 20 Fr ID/ 7.7 mm OD
  • Z-3156-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Main Body Extension, REF ESBE-22-29, REF G32179, 18 Fr. ID/ 6.90 OD, Sterile.
  • Z-3139-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-10-71-ZT, REF G48444, 14 Fr ID/ 5.3 mm OD, Sterile.
  • Z-3137-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-32-96-ZT, REF G48428, 20 Fr ID/ 7.7 mm OD, Sterile.
  • Z-3131-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-24-96-ZT, REF G48408, 18 Fr ID/ 7.1 mm OD, Sterile
  • Z-3158-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Main Body Extension, REF ESBE-32-39, REF G32184, 20 Fr. ID/ 7.7 mm OD, Sterile.
  • Z-3130-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-24-82-ZT, REF G48407, 18 Fr ID/ 7.1 mm OD, Sterile
  • Z-3151-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-22-73-ZT, REF G48474, 16Fr ID/ 6.0 mm OD, Sterile.
  • Z-3149-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-90-ZT, REF G48471, 16Fr ID/ 6.0 mm OD, Sterile.
  • Z-3141-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-14-56-ZT, REF G48455, 16Fr ID/ 6.0 mm OD, Sterile.
  • Z-3136-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-32-82-ZT, REF G48427, 20 Fr ID/ 7.7 mm OD, Sterile.
  • Z-3146-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-18-73-ZT, REF G48466, 16Fr ID/ 6.0 mm OD, Sterile.
  • Z-3145-2011 · Apr 15, 2011 · Terminated
    Cook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-18-56-ZT, REF G48465, 16Fr ID/ 6.0 mm OD, Sterile.

…and 96 more.

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page