Zenith G32541Recalls & reported problems
William A. Cook Australia Pty. Ltd. · Zenith Fenestrated, AAA Endovascular Graft, Proximal Body · Class 3 · Implant
All models of Zenith (32)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 2
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002325415), received between Aug 14, 2022 and Aug 25, 2022 — classified as 2 deaths.
The most frequently reported specific device problem is “Break” (2 reports).
111 other recalls cover devices cleared under the same FDA submission (P020018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
- Activation Problem1
- Entrapment of Device1
- Material Twisted/Bent1
Patient problems
- Failure of Implant2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
DeathAug 25, 2022Break; Break
THE DEVICE WOULD ROTATE 180 DEGREES EVERY TIME IT WAS ADVANCED THROUGH THE GROIN. ROTATION WAS EVENTUALLY CORRECTED AND GRAFT DEPLOYMENT WAS BEGUN. AFTER OPENING THE FIRST TWO STENTS, THE ORIENTATION ROTATED ON THE GRAFT. ROTATION OF THE ENTIRE SYSTEM WAS PERFORMED. WHEN THE GRAFT WAS NOW CORRECTLY ORIENTED, THE DISTAL END OF THE GRAFT WAS OPENED. UPON, OPENING THE DISTAL END, IT WAS NOTED THAT THE GRAFT WAS WRAPPED AROUND ITSELF (LIKE A CANDY WRAPPER). CANNULATION OF THE FENESTRATIONS WAS PERFORMED FOR THE ARM THROUGH THE SCALLOP AS THE PHYSICIAN COULD NOT CANNULATE THROUGH THE WRAPPED GRAFT (FROM BELOW). THEREFORE, CANNULATION WAS DONE THROUGH THE ARM. UPON REMOVING, THE GOLD TRIGGER WIRE, THE WIRE SNAPPED. HEMOSTATS WERE THEN USED TO GRAB THE REMAINDER OF THE WIRE THAT SNAPPED FROM THE GOLD KNOB. THE BLACK TRIGGER WIRE ALSO SNAPPED AT THE HUB AND WAS UNABLE TO BE PULLED FROM THE DEVICE. THEREFORE, THE TOP CAP COULD NOT BE DEPLOYED. THE OPERATION WAS CONVERTED TO AN OPEN PROCEDURE. THE PATIENT EXPIRED IN THE ICU WHILE AT THE HOSPITAL. NOTE: THE PATIENT HAD AN EXISTING ENDOLOGICS GRAFT SITTING APPROX. 2 CM'S BELOW THE PATIENTS RENALS.
Patient problems: Failure of Implant
MAUDE report 15294596 ↗DeathAug 14, 2022Break; Entrapment of Device; Material Twisted/Bent; Activation Problem; Material Twisted/Bent
THE DEVICE WOULD ROTATE 180 DEGREES EVERY TIME IT WAS ADVANCED THROUGH THE GROIN. ROTATION WAS EVENTUALLY CORRECTED AND GRAFT DEPLOYMENT WAS BEGUN. AFTER OPENING THE FIRST TWO STENTS, THE ORIENTATION ROTATED ON THE GRAFT. ROTATION OF THE ENTIRE SYSTEM WAS PERFORMED. WHEN THE GRAFT WAS NOW CORRECTLY ORIENTED, THE DISTAL END OF THE GRAFT WAS OPENED. UPON, OPENING THE DISTAL END, IT WAS NOTED THAT THE GRAFT WAS WRAPPED AROUND ITSELF (LIKE A CANDY WRAPPER). CANNULATION OF THE FENESTRATIONS WAS PERFORMED FOR THE ARM THROUGH THE SCALLOP AS THE PHYSICIAN COULD NOT CANNULATE THROUGH THE WRAPPED GRAFT (FROM BELOW). THEREFORE, CANNULATION WAS DONE THROUGH THE ARM. UPON REMOVING, THE GOLD TRIGGER WIRE, THE WIRE SNAPPED. HEMOSTATS WERE THEN USED TO GRAB THE REMAINDER OF THE WIRE THAT SNAPPED FROM THE GOLD KNOB. THE BLACK TRIGGER WIRE ALSO SNAPPED AT THE HUB AND WAS UNABLE TO BE PULLED FROM THE DEVICE. THEREFORE, THE TOP CAP COULD NOT BE DEPLOYED. THE OPERATION WAS CONVERTED TO AN OPEN PROCEDURE. THE PATIENT EXPIRED IN THE ICU WHILE AT THE HOSPITAL.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant
MAUDE report 15223055 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020018. One submission often covers many models, so check the product and lot details in each record.
- Z-3152-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-73-ZT, REF G48476, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3141-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-14-56-ZT, REF G48455, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3132-2011 · Apr 15, 2011 · TerminatedCook Medical, Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-26-96-ZT, REF G48413, 18 Fr ID/ 7.1 mm OD, Sterile.
- Z-3142-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-16-56-ZT, REF G48461, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3145-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-18-56-ZT, REF G48465, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3153-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-56-ZT, REF G48477, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3133-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-26-125-ZT, REF G48415, 18 Fr ID/ 7.1 mm OD, Sterile.
- Z-3149-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-90-ZT, REF G48471, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3130-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-24-82-ZT, REF G48407, 18 Fr ID/ 7.1 mm OD, Sterile
- Z-3137-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-32-96-ZT, REF G48428, 20 Fr ID/ 7.7 mm OD, Sterile.
- Z-3158-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Main Body Extension, REF ESBE-32-39, REF G32184, 20 Fr. ID/ 7.7 mm OD, Sterile.
- Z-3144-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-16-90-ZT, REF G48463, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3147-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-56-ZT, REF G48469, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3135-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-30-82-ZT, REF G48422, 20 Fr ID/ 7.7 mm OD
- Z-3140-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-12-56-ZT, REF G48449, 16Fr ID/ 6.0 mm OD, Sterile.
…and 96 more.
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page