Zenith G32562Recalls & reported problems
William A. Cook Australia Pty. Ltd. · Zenith Fenestrated, AAA Endovascular Graft, Distal Body · Class 3 · Implant
All models of Zenith (32)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10827002325620), received between Jun 29, 2015 and Feb 13, 2026 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
111 other recalls cover devices cleared under the same FDA submission (P020018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak2
- Material Split, Cut or Torn2
- Appropriate Device Problem Term/Code Not Available1
- Difficult to Remove1
- Leak/Splash1
- Mechanical Problem1
Patient problems
- Clinical Endoleak2
- Hemorrhage/Blood Loss/Bleeding1
- No Code Available1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryFeb 13, 2026Fluid/Blood Leak; Material Split, Cut or Torn
ACCORDING TO THE INITIAL REPORTER: DURING THE ORIGINAL IMPLANT ON (B)(6) 2025, AN ENDOLEAK WAS OBSERVED, BUT ITS CAUSE HAD NOT BEEN CONFIRMED AT THE TIME. A FOLLOW-UP CT SCAN ON (B)(6) 2025, SHOWED THAT THE ANEURYSM HAD GROWN BY 0.5 CM. ANOTHER CT SCAN ON (B)(6) 2025, REVEALED AN ADDITIONAL 1.0 CM OF GROWTH. ON (B)(6) 2025, THE PATIENT UNDERWENT AN ANGIOGRAM AND BILATERAL LIMB COMPONENT BALLOONING, BUT THE PROCEDURE DID NOT RESOLVE THE ANEURYSM, AND NO FURTHER INTERVENTIONS WERE PERFORMED AT THAT TIME. A SUBSEQUENT CT SCAN ON (B)(6) 2026, SHOWED AN ADDITIONAL 0.5 CM OF GROWTH. ON (B)(6) 2026, THE PATIENT HAD AN INTERVENTION IN WHICH A GORE EXCLUDER WAS IMPLANTED TO RESOLVE THE ISSUE. DURING THE IMPLANT PROCEDURE, THE DEVICE IFU WAS FOLLOWED, AND THE DEVICE FUNCTIONED AS EXPECTED WHILE FOLLOWING THE IFU. THE ENDOLEAK WAS DISCOVERED ON THE FINAL ANGIOGRAM DURING THE DEPLOYMENT PROCEDURE. THERE WERE NO ADVERSE EVENTS, BUT THE PATIENT WAS AT RISK FOR PENDING RUPTURE. THE PATIENT OUTCOME WAS A SUCCESSFUL RE-INTERVENTION. NO PART OF THE PROCEDURE WAS PERFORMED OFF-LABEL OR OUTSIDE THE PATIENT SELECTION CRITERIA. THERE WERE NO DIFFICULTIES EXPERIENCED DURING DEPLOYMENT. THE PATIENT¿S ANAT
Patient problems: Clinical Endoleak
MAUDE report 24335292 ↗InjuryFeb 10, 2026Fluid/Blood Leak; Material Split, Cut or Torn
SUSPECTED FABRIC TEAR AT CROTCH OF GRAFT:DURING ORIGINAL IMPLANT ((b)(6) 2025), AN ENDOLOEAK WAS OBSERVED BUT THE CAUSE HAD NOT BEEN CONFIRMED.(b)(6) 2025 - A FOLLOW UP CT SHOWED THAT THE ANEURYSM GREW 0.5 CM.(b)(6) 2025 - AN ADDITIONAL FOLLOW UP CT SHOWED THAT THE ANEURYSM GREW AN ADDITIONAL 1.0 CM.(b)(6) 2025 - THE PATIENT WAS TAKEN IN FOR AN ANGIOGRAM AND BILATERAL LIMB COMPONENT BALLOONING. IT DID NOT RESOLVE THE ANEURYSM. THEY DID NOT DO ANYTHING ELSE.(b)(6) 2026 - ANOTHER CT SHOWED AN ADDITIONAL 0.5 CM OF GROWTH.(b)(6) 2026 - THE PATIENT HAD AN INTERVENTION THAT PUT IN A GORE EXCLUDER TO RESOLVE THE ISSUE.ADDITIONAL INFORMATION RECEIVED:DURING THE IMPLANT PROCEDURE, THE DEVICE IFU WAS FOLLOWED, AND THE DEVICE FUNCTIONED AS EXPECTED WHILE FOLLOWING THE IFU. ENDOLEAK WAS DISCOVRED ON THE FINAL ANGIOGRAM DURING DEPLOYMENT PROCEDURE.THERE WERE NO ADVERSE EVENTS, BUT THE PATIENT WAS AT RISK FOR PENDING RUPTURE.THE PATIENT OUTCOME WAS SUCCESSFUL RE-INTERVENTION.NO PART OF THE PROCEDURE WAS PERFORMED OFF-LABEL OR OUT OF THE PATIENT SELECTION CRITERIA.NO DIFFICULTY WAS EXPERIENCED DURING DEPLOYMENT.THE PATIENT'S ANATOMY WAS A LITTLE TORTUOUS, BUT NOT BAD.THE PHYSICIAN THINKS A FABRIC
Patient problems: Clinical Endoleak
MAUDE report 24310803 ↗InjuryOct 1, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
FOLLOWING THE PROCEDURE FOR IMPLANTATION, THERE WAS A LEAK AT THE DISTAL EDGE OF THE OVERLAP BETWEEN THE TWO DEVICES ON COMPLETION ANGIOGRAM. THE SOURCE OF THE LEAK COULD NOT BE DETERMINED. THE PHYSICIAN IMPLANTED TWO ADDITIONAL COVERED STENTS EXTENDING SLIGHTLY ABOVE THE FLOW DIVIDER AND POST DILATED THEM. ANOTHER ANGIOGRAM FOLLOWING THIS SHOWED THE LEAK TO BE REPAIRED BUT THE EXACT SOURCE OF THE LEAK WAS NOT IDENTIFIED.
Patient problems: No Code Available
MAUDE report 9139180 ↗InjurySep 18, 2019Leak/Splash
FOLLOWING THE PROCEDURE FOR IMPLANTATION, THERE WAS A LEAK AT THE DISTAL EDGE OF THE OVERLAP BETWEEN THE TWO DEVICES ON COMPLETION ANGIOGRAM. THE SOURCE OF THE LEAK COULD NOT BE DETERMINED. THE PHYSICIAN IMPLANTED TWO ADDITIONAL COVERED STENTS EXTENDING SLIGHTLY ABOVE THE FLOW DIVIDER AND POST DILATED THEM. ANOTHER ANGIOGRAM FOLLOWING THIS SHOWED THE LEAK TO BE REPAIRED BUT THE EXACT SOURCE OF THE LEAK WAS NOT IDENTIFIED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 9084244 ↗MalfunctionJun 29, 2015Mechanical Problem; Difficult to Remove
DURING DEPLOYMENT OF A ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT DISTAL BIFURCATED BODY INTO THE RIGHT COMMON ILIAC ARTERY FOR THE TREATMENT OF ILIAC OCCLUSION, THE BLACK TRIGGER WIRE MECHANISM GOT STUCK AND WOULDN'T COME OUT. THE USER TRIED TWISTING IT WITHOUT SUCCESS. THEN PLIERS WERE USED TO PULL THE WIRES OUT. THE WIRES CAME OUT EASILY. THE WHITE TRIGGER WIRE COULDN'T BE REMOVED BECAUSE THE BLACK TRIGGER WIRE MECHANISM WAS STUCK ON THE DELIVERY SYSTEM. THIS WIRE WAS ALSO PULLED OUT WITH PLIERS. THE CASE ENDED UP GOING AS PLANNED FOLLOWING THIS. THE DEVICE WAS SUCCESSFULLY DEPLOYED AND REMAINS IN THE PATIENT AS INTENDED. THE PATIENT DID NOT REQUIRE AN ADDITIONAL PROCEDURE DUE TO THIS OCCURRENCE, HOWEVER IT DID ADD APPROXIMATELY 40 MINUTES TO THE PROCEDURE TIME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 4886693 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020018. One submission often covers many models, so check the product and lot details in each record.
- Z-3154-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-24-73-ZT, REF G48478, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3132-2011 · Apr 15, 2011 · TerminatedCook Medical, Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-26-96-ZT, REF G48413, 18 Fr ID/ 7.1 mm OD, Sterile.
- Z-3136-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-32-82-ZT, REF G48427, 20 Fr ID/ 7.7 mm OD, Sterile.
- Z-3150-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-56-ZT, REF G48473, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3145-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-18-56-ZT, REF G48465, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3151-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-22-73-ZT, REF G48474, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3134-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-28-96-ZT, REF G48418, 20 Fr ID/ 7.7 mm OD, Sterile
- Z-3148-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-73-ZT, REF G48470, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3137-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-32-96-ZT, REF G48428, 20 Fr ID/ 7.7 mm OD, Sterile.
- Z-3146-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-18-73-ZT, REF G48466, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3157-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Main Body Extension, REF ESBE-30-39, REF G32183, 20 Fr. ID/ 7.7 OD, Sterile.
- Z-3152-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-73-ZT, REF G48476, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3140-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-12-56-ZT, REF G48449, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3156-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Main Body Extension, REF ESBE-22-29, REF G32179, 18 Fr. ID/ 6.90 OD, Sterile.
- Z-3153-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-56-ZT, REF G48477, 16Fr ID/ 6.0 mm OD, Sterile.
…and 96 more.
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page