Zenith G32592Recalls & reported problems
William A. Cook Australia Pty. Ltd. · Zenith Fenestrated, AAA Endovascular Graft, Distal Body · Class 3 · Implant
All models of Zenith (32)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 2
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10827002325927), received between Mar 11, 2020 and Oct 26, 2020 — classified as 2 deaths, 2 injuries.
The most frequently reported specific device problem is “Coagulation in Device or Device Ingredient” (1 report).
111 other recalls cover devices cleared under the same FDA submission (P020018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
- Coagulation in Device or Device Ingredient1
- Material Split, Cut or Torn1
- Structural Problem1
Patient problems
- Death2
- Occlusion2
- Rupture2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
DeathOct 26, 2020Structural Problem
A (B)(6) YEAR OLD FEMALE PATIENT UNDERWENT A ZENITH FENESTRATED GRAFT PLACEMENT FOR A ABDOMINAL AORTIC ANEURYSM IN WHICH THE ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT WAS USED. THE CASE WENT AS STANDARD. IN THE RIGHT ILIAC LIMB THE PHYSICIAN USED A GORE GRAFT AS INITIALLY PLANNED. AFTER PLACED THE STENT LOOKED LIKE IT WAS NARROWED SO THE PHYSICIAN PUT DOWN KISSING STENTS AND BALLOONED AT THE NARROWING. THE BALLOON WAS A BARD BALLOON AND USED 14MM ON THE RIGHT AND USED 12MM ON THE LEFT. AT THIS TIME THE PATIENT BLOOD PRESSURE STARTED TO DROP. THE PHYSICIAN REALIZED RUPTURED THE AORTA WHEN BALLOONED. A CODA BALLOON WAS PUT DOWN AND SLOWED THE LEAK. PHYSICIAN DID AN ANGIOGRAM AND SAW CONTRAST LEAKING AT THE BIFURCATED AREA. THE PHYSICIAN PLACED ANOTHER GORE GRAFT DOWN TO THE FENESTRATED TO REALIGN THE LEG AND SEAL IT OFF. JUST UNDER THE RENAL STENTS ANOTHER GORE GRAFT WAS PLACED TO RE-STENT LOWEST RIGHT RENAL TO REASSURE NO LEAKING AROUND THERE. THE PATIENT WAS GIVEN A UNIT OF BLOOD AND WOULD STABILIZE, THEN BLOOD PRESSURE WOULD DROP AND PATIENT WAS GIVEN ANOTHER UNIT OF BLOOD TO STABILIZE; WAS DONE DURING PROCEDURE TO KEEP PATIENT STABILIZED. THE DISTRICT MANAGER DUE TO SCHEDULE LEFT
Patient problems: Death; Rupture
MAUDE report 10734485 ↗DeathOct 15, 2020Material Split, Cut or Torn; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn
A 78 YEAR OLD FEMALE PATIENT UNDERWENT A ZENITH FENESTRATED GRAFT PLACEMENT FOR A ABDOMINAL AORTIC ANEURYSM IN WHICH THE ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT WAS USED. THE CASE WENT AS STANDARD. IN THE RIGHT ILIAC LIMB THE PHYSICIAN USED A GORE GRAFT AS INITIALLY PLANNED. AFTER PLACED THE STENT LOOKED LIKE IT WAS NARROWED SO THE PHYSICIAN PUT DOWN KISSING STENTS AND BALLOONED AT THE NARROWING. THE BALLOON WAS A BARD BALLOON AND USED 14MM ON THE RIGHT AND USED 12MM ON THE LEFT. AT THIS TIME THE PATIENT BLOOD PRESSURE STARTED TO DROP. THE PHYSICIAN REALIZED RUPTURED THE AORTA WHEN BALLOONED. A CODA BALLOON WAS PUT DOWN AND SLOWED THE LEAK. PHYSICIAN DID AN ANGIOGRAM AND SAW CONTRAST LEAKING AT THE BIFURCATED AREA. THE PHYSICIAN PLACED ANOTHER GORE GRAFT DOWN TO THE FENESTRATED TO REALIGN THE LEG AND SEAL IT OFF. JUST UNDER THE RENAL STENTS ANOTHER GORE GRAFT WAS PLACED TO RE-STENT LOWEST RIGHT RENAL TO REASSURE NO LEAKING AROUND THERE. THE PATIENT WAS GIVEN A UNIT OF BLOOD AND WOULD STABILIZE, THEN BLOOD PRESSURE WOULD DROP AND PATIENT WAS GIVEN ANOTHER UNIT OF BLOOD TO STABILIZE; WAS DONE DURING PROCEDURE TO KEEP PATIENT STABILIZED. THE DISTRICT MANAGER DUE TO SCHEDULE LEFT DUR
Patient problems: Death; Rupture
MAUDE report 10681898 ↗InjuryMar 12, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
A PATIENT WITH A PRE-EXISTING ENDOGRAFT PLACED AT ANOTHER FACILITY HAD COME IN WITH A TOTAL LIMB OCCLUSION DUE TO THROMBUS. PHYSICIAN LYSED THROMBUS AND STENTED ILIAC LIMB. PATIENT IS DOING WELL. THE LIMB OCCLUSION CAUSED THE PATIENT TO BE SYMPTOMATIC. ON THE CONTRALATERAL SIDE.
Patient problems: Occlusion
MAUDE report 9826144 ↗InjuryMar 11, 2020Coagulation in Device or Device Ingredient
A PATIENT WITH A PRE-EXISTING ENDOGRAFT PLACED AT ANOTHER FACILITY HAD COME IN WITH A TOTAL LIMB OCCLUSION DUE TO THROMBUS. PHYSICIAN LYSED THROMBUS AND STENTED ILIAC LIMB. PATIENT IS DOING WELL. THE LIMB OCCLUSION CAUSED THE PATIENT TO BE SYMPTOMATIC.
Patient problems: Occlusion
MAUDE report 9820969 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020018. One submission often covers many models, so check the product and lot details in each record.
- Z-3135-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-30-82-ZT, REF G48422, 20 Fr ID/ 7.7 mm OD
- Z-3147-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-56-ZT, REF G48469, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3155-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-24-90-ZT, REF G48479, 16 Fr. ID/ 6.0 OD, Sterile.
- Z-3132-2011 · Apr 15, 2011 · TerminatedCook Medical, Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-26-96-ZT, REF G48413, 18 Fr ID/ 7.1 mm OD, Sterile.
- Z-3130-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-24-82-ZT, REF G48407, 18 Fr ID/ 7.1 mm OD, Sterile
- Z-3145-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-18-56-ZT, REF G48465, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3139-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-10-71-ZT, REF G48444, 14 Fr ID/ 5.3 mm OD, Sterile.
- Z-3133-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-26-125-ZT, REF G48415, 18 Fr ID/ 7.1 mm OD, Sterile.
- Z-3153-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-56-ZT, REF G48477, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3142-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-16-56-ZT, REF G48461, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3136-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-32-82-ZT, REF G48427, 20 Fr ID/ 7.7 mm OD, Sterile.
- Z-3141-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-14-56-ZT, REF G48455, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3146-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-18-73-ZT, REF G48466, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3134-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-28-96-ZT, REF G48418, 20 Fr ID/ 7.7 mm OD, Sterile
- Z-3151-2011 · Apr 15, 2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-22-73-ZT, REF G48474, 16Fr ID/ 6.0 mm OD, Sterile.
…and 96 more.
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page