Zenith Alpha G35348Recalls & reported problems
William Cook Europe ApS · Zenith Alpha, Thoracic Endovascular Graft, Proximal Component · Class 3 · Implant
All models of Zenith Alpha (67)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 2
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002353487), received between Oct 5, 2020 and Oct 14, 2020 — classified as 2 deaths.
The most frequently reported specific device problem is “Failure to Advance” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (P140016); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Failure to Advance2
- Difficult to Advance1
- Off-Label Use1
Patient problems
- Cardiac Arrest2
- Death2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
DeathOct 14, 2020Failure to Advance; Failure to Advance
FURTHER INFORMATION PROVIDE STATES: LOCATION OF THE PRIMARY ENTRY TEAR DISTAL TO THE LEFT SUBCLAVIAN ARTERY (LSA) DISSECTION EXTENDED FROM JUST DISTAL TO LSA DOWN INTO THE ILIACS, WITH A KNOWN RUPTURE IN THE THORACIC AORTA. ZDEG WAS PLACED PROXIMALLY AT THE DISTAL EDGE OF THE LEFT COMMON CAROTID ARTERY AND DISTALLY IN THE MID-DESCENDING THORACIC AORTA (6-8CM ABOVE THE CELIAC). ZDES WAS PLACED INSIDE THE ZDEG PROXIMALLY AND EXTENDED INTO THE MID ABDOMINAL AORTA. THE RUPTURE THAT WAS TREATED IN THE FIRST CASE, BUT NOT SEEN ON FINAL ANGIO AT THE END OF THE FIRST CASE, WAS AGAIN SHOWN TO HAVE EXTRAVASATION BEING FED FROM FALSE LUMEN FILLING DISTAL TO THE ZDEG. THE ZDES ACCORDIONED OCCURRED WHEN ATTEMPTING TO DELIVER THE ALPHA DEVICE. IT SEEMED TO GET CAUGHT ON THE BARE STENT AND WHEN RECOGNIZED THAT WAS THE RESISTANCE, THE STENT HAD KINKED MOVED OUT AND UP. ZENITH ALPHA DEVICE WS PLACED PROXIMALLY INTO THE ZDEG, DISTALLY BELOW THE ZDES AND BELIEVED TO BE ABOUT THE LEVEL OF THE CELIAC. CHOSE TO COVER AND TRY TO EXCLUDE RUPTURE AS QUICKLY AS POSSIBLE AS CPR HAD BEEN ON GOING FOR AN EXTENDED PERIOD OF TIME, SO NO CONFORMATION ANGIO WAS DONE PRIOR TO PLACEMENT.
Patient problems: Cardiac Arrest; Death
MAUDE report 10680241 ↗DeathOct 5, 2020Off-Label Use; Failure to Advance; Difficult to Advance; Off-Label Use; Failure to Advance; Difficult to Advance
ADDITIONAL INFORMATION RECEIVED ON 16OCT2020: LOCATION OF THE PRIMARY ENTRY TEAR DISTAL TO THE LEFT SUBCLAVIAN ARTERY (LSA) DISSECTION EXTENDED FROM JUST DISTAL TO LSA DOWN INTO THE ILIACS, WITH A KNOWN RUPTURE IN THE THORACIC AORTA. ZDEG WAS PLACED PROXIMALLY AT THE DISTAL EDGE OF THE LEFT COMMON CAROTID ARTERY AND DISTALLY IN THE MID-DESCENDING THORACIC AORTA (6-8CM ABOVE THE CELIAC). ZDES WAS PLACED INSIDE THE ZDEG PROXIMALLY AND EXTENDED INTO THE MID ABDOMINAL AORTA. THE RUPTURE THAT WAS TREATED IN THE FIRST CASE, BUT NOT SEEN ON FINAL ANGIO AT THE END OF THE FIRST CASE, WAS AGAIN SHOWN TO HAVE EXTRAVASATION BEING FED FROM FALSE LUMEN FILLING DISTAL TO THE ZDEG. THE ZDES ACCORDIONED OCCURRED WHEN ATTEMPTING TO DELIVER THE ALPHA DEVICE. IT SEEMED TO GET CAUGHT ON THE BARE STENT AND WHEN RECOGNIZED THAT WAS THE RESISTANCE, THE STENT HAD KINKED MOVED OUT AND UP. ZENITH ALPHA DEVICE WAS PLACED PROXIMALLY INTO THE ZDEG, DISTALLY BELOW THE ZDES AND BELIEVED TO BE ABOUT THE LEVEL OF THE CELIAC. CHOSE TO COVER AND TRY TO EXCLUDE RUPTURE AS QUICKLY AS POSSIBLE AS CPR HAD BEEN ON GOING FOR AN EXTENDED PERIOD OF TIME, SO NO CONFORMATION ANGIO WAS DONE PRIOR TO PLACEMENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Cardiac Arrest; Death
MAUDE report 10626041 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P140016. One submission often covers many models, so check the product and lot details in each record.
- Z-0043-2026 · Sep 18, 2025 · Open, ClassifiedZenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W ZTA2-P-40-217-W ZTA2-P-42-121 ZTA2-P-42-121-W
- Z-2099-2017 · Mar 22, 2017 · TerminatedZenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with isolated lesions of the descend
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page