Skip to content

Zenith Alpha G35355Recalls & reported problems

William Cook Europe ApS · Zenith Alpha, Thoracic Endovascular Graft, Proximal Component · Class 3 · Implant

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10827002353555), received between Jan 7, 2020 and Oct 22, 2021 — classified as 4 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (P140016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available2
  • Device Dislodged or Dislocated2
  • Off-Label Use1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • No Code Available2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 22, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated

    DESCRIPTION OF EVENT ACCORDING TO THE INITIAL REPORTER: COMPONENT SEPARATION. OVERLAP SECTION NOT INCLUDED IN SCAN RANGE. COMPONENT SEPARATION, ON (B)(6) 2018: THORACIC ENDOGRAFT GORE 40X150. COMPONENT SEPARATION REPAIR. PATIENT OUTCOME: ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12680696 ↗
  • InjuryOct 18, 2021Off-Label Use; Device Dislodged or Dislocated; Off-Label Use; Device Dislodged or Dislocated

    IMPLANT PROCEDURE: ON (B)(6) 2017 WITH ZTA A FEN-THORACO-ABDOMINAL-GRAFT (CMD). COMPONENT SEPARATION: (B)(6) 2018.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12645647 ↗
  • InjuryJan 8, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    INITIAL REPORTER EVENT DESCRIPTION: ALPHA DEVICE USED FOR A THORACIC ANEURYSM. LOOKS LIKE A TYPE 3 ENDOLEAK AT STEEPEST CURVE OF THE ARCH. REINTERVENTION ON (B)(6) 2019 TO IMPLANT ANOTHER ALPHA (ZTA-P-38-167-W) TO SEAL OFF LEAK. THE PATIENT NO LONGER HAS A LEAK. FIRST IMPLANT WAS TUESDAY, (B)(6) 2019. THE DOCTOR SCANNED THE PATIENT (B)(6) 2019 AND SAW AN ENDOLEAK. HE WASN¿T SURE IF IT WAS A TYPE 1A OR TYPE 3. HE LINED THE GRAFT WITH A NEW ALPHA ON (B)(6) 2019. THE PATIENT DOES NOT HAVE A LEAK ANY MORE. THE GRAFT LANDED IN ZONE 2, JUST DISTAL TO THE CAROTID ARTERY. PATIENT OUTCOME: THE PATIENT DID REQUIRE ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE: AN ADDITIONAL IMPLANT OF AN ALPHA DEVICE WAS SCHEDULED FOR (B)(6) 2019 TO SEAL OFF ENDOLEAK. A ZTA-P-38-167-W WAS IMPLANTED ON (B)(6) 2019. DOCTOR IS HAPPY WITH FINAL RESULT.

    Patient problems: No Code Available

    MAUDE report 9566050 ↗
  • InjuryJan 7, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: ALPHA DEVICE USED FOR A THORACIC ANEURYSM. LOOKS LIKE A TYPE 3 ENDOLEAK AT STEEPEST CURVE OF THE ARCH. REINTERVENTION ON (B)(6) 2019 TO IMPLANT ANOTHER ALPHA (ZTA-P-38-167-W) TO SEAL OFF LEAK. THE PATIENT NO LONGER HAS A LEAK. FIRST IMPLANT WAS TUESDAY, (B)(6) 2019. THE DOCTOR SCANNED THE PATIENT (B)(6) 2019 AND SAW AN ENDOLEAK. HE WASN¿T SURE IF IT WAS A TYPE 1A OR TYPE 3. HE LINED THE GRAFT WITH A NEW ALPHA ON (B)(6) 2019. THE PATIENT DOES NOT HAVE A LEAK ANY MORE. THE GRAFT LANDED IN ZONE 2, JUST DISTAL TO THE CAROTID ARTERY. PATIENT OUTCOME: THE PATIENT DID REQUIRE ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE: AN ADDITIONAL IMPLANT OF AN ALPHA DEVICE WAS SCHEDULED FOR (B)(6) 2019 TO SEAL OFF ENDOLEAK. A ZTA-P-38-167-W WAS IMPLANTED ON (B)(6) 2019. DOCTOR IS HAPPY WITH FINAL RESULT. HAS THE COMPLAINANT REPORTED THAT THE PRODUCT CAUSED OR CONTRIBUTED TO THE ADVERSE EFFECTS: NO.

    Patient problems: No Code Available

    MAUDE report 9556580 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P140016. One submission often covers many models, so check the product and lot details in each record.

  • Z-0043-2026 · Sep 18, 2025 · Open, Classified
    Zenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W ZTA2-P-40-217-W ZTA2-P-42-121 ZTA2-P-42-121-W
  • Z-2099-2017 · Mar 22, 2017 · Terminated
    Zenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with isolated lesions of the descend

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →