Skip to content

Zenith Alpha G35356Recalls & reported problems

William Cook Europe ApS · Zenith Alpha, Thoracic Endovascular Graft, Proximal Component · Class 3 · Implant

All models of Zenith Alpha (67)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002353562), received between Sep 9, 2016 and Apr 2, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).

2 other recalls cover devices cleared under the same FDA submission (P140016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation, Positioning or Separation Problem1
  • Improper or Incorrect Procedure or Method1
  • Off-Label Use1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 2, 2024Off-Label Use; Improper or Incorrect Procedure or Method; Off-Label Use; Improper or Incorrect Procedure or Method

    DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: "UPON DEPLOYMENT OF THE DEVICE, IT WAS MOVING BACKWARDS AND THEN UPON SURGEON MANIPULATING IT JUMPED FORWARD AND ENCROACHED ON THE ARCH VESSELS (INNOMINATE & LEFT COMMON CAROTID ARTERY)LCCA)). THE LEFT SUBCLAVIAN ARTERY (LSA) WAS PLANNED FOR COVERAGE AS A LCCA-LSA BYPASS WAS DONE PRIOR. UPON AN ANGIO THE LCCA HAD A FAIR AMOUNT OF COVERAGE THAT REQUIRED STENTING TO ENSURE THE VESSEL REMAINS OPEN. FINAL RESULT OF THE CASE WAS AS PLANNED WITH NO ENDOLEAK AND ALL VESSELS PRESERVED. NO PREEXISTING CONDITIONS OR COMORBIDITIES PRESENT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19018944 ↗
  • MalfunctionSep 9, 2016Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem

    DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: DURING DEPLOYMENT, THERE WAS A LOT OF DIFFICULTY RETRACTING THE SHEATH. ONCE THE SHEATH GOT TO THE FIRST PART OF THE BARE STENT, THERE WAS EXTREME DIFFICULTY RETRACTING THE SHEATH ANY MORE. THEY HEARD A SNAP. THEY BELIEVE THAT THE SHEATH'S NITINOL CORE BROKE. BECAUSE THE SHEATH WOULD NOT RETRACT FROM THAT POINT ON, THERE WAS FEAR TO TRY TO COMPLETE DEPLOYMENT ANY FURTHER. IT WAS DECIDED TO REMOVE THE DELIVERY SYSTEM AND REPLACE WITH A NEW GRAFT. THEY COMPLETED THE PROCEDURE SUCCESSFULLY WITH A SHORTER GRAFT (36 X 161). THE PHYSICIAN AND REP SUSPECT THAT THE TIGHT AORTIC ARCH MADE DEPLOYMENT DIFFICULT. PATIENT OUTCOME: ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5938700 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P140016 on Sep 15, 2015. Submitted by Cook Medical Incorporated as “Zenith Alpha Thoracic Endovascular Graft”. The FDA review took 398 days. The FDA has since approved 18 supplements (changes to the device or its labeling), most recently on Aug 4, 2026. FDA record ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Recalls under the same FDA submission

These recalls involve devices cleared under P140016. One submission often covers many models, so check the product and lot details in each record.

  • Z-0043-2026 · Sep 18, 2025 · Open, Classified
    Zenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W ZTA2-P-40-217-W ZTA2-P-42-121 ZTA2-P-42-121-W
  • Z-2099-2017 · Mar 22, 2017 · Terminated
    Zenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with isolated lesions of the descend

Common questions about the Zenith Alpha G35356

Has the Zenith Alpha G35356 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10827002353562), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Zenith Alpha G35356?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Sep 9, 2016 and Apr 2, 2024. In those reports, the most common specific device problems are “Activation, Positioning or Separation Problem” (1), “Improper or Incorrect Procedure or Method” (1) and “Off-Label Use” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Zenith Alpha G35356?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Zenith Alpha G35356 FDA approved?

It was approved through premarket approval (PMA) P140016 on Sep 15, 2015.

Is the Zenith Alpha G35356 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Zenith Alpha G35356, and is it still sold?

The FDA's device database lists William Cook Europe ApS as the company responsible for it (brand Zenith Alpha, model G35356). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page