Zenith Alpha G35367Recalls & reported problems
William Cook Europe ApS · Zenith Alpha, Thoracic Endovascular Graft, Proximal Component · Class 3 · Implant
All models of Zenith Alpha (67)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002353678), received between Mar 6, 2018 and Apr 8, 2020 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Loss of or Failure to Bond” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (P140016); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Loss of or Failure to Bond2
- Activation Failure1
Patient problems
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 8, 2020Loss of or Failure to Bond; Loss of or Failure to Bond
DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: A (B)(6) YEAR OLD MALE PATIENT UNDERWENT A TVAR PROCEDURE IN WHICH THE ZENITH ALPHA THORACIC ENDOVASCULAR GRAFT PROXIMAL COMPONENTS, G35367, WAS USED. THE LAST GRAFT PIECE, G35367, WAS USED FOR BRIDGING BETWEEN TWO GRAFTS AS THE PROXIMAL GRAFT WAS TOO SHORT WITH THE DISTAL PIECE. WHEN THE PHYSICIAN FINISHED AND ROTATED THE DELIVERY SYSTEM THE NOSE CONE OF THE DEVICE INNER WHITE CANNULA CAME OFF AND STAYED ON THE WIRE END OF THE GRAFT. LEAVING THE NOSE CONE AND INNER CANNULA IN THE PATIENT. THE PHYSICIAN THEN ADVANCED A LONGER AND LARGER SHEATH AND SNARED THE SHEATH TO REMOVE FROM THE PATIENT. ADDITIONAL INFORMATION RECEIVED 20MAR2020: AFTER RELEASING THE STENT GRAFT, THE PHYSICIAN RETRIEVED THE INNER SYSTEM AND LEFT THE OUTER SHEATH IN ORDER TO PERFORM BALLOONING. WHEN THE INNER SYSTEM WAS TAKEN OUT, IT WAS NOTED THAT THE SYSTEM ONLY CONTAINED THE GREY POSITIONER ALONG WITH THE HANDLE AND THE TIP ASSEMBLY (DILATOR TIP AND CANNULA) WAS LEFT IN PATIENT. THERE WAS VERY TORTUROUS ANATOMY AND WE WERE WELL IN THE ARCH AROUND ZONE 2. SOME DIFFICULTY RETRIEVING DEVICE OVER THE ARCH BUT AS EXPECTED WITH THE ANATOMY AND SIZE OF ANEURYSM. PA
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9937424 ↗InjuryApr 1, 2020Loss of or Failure to Bond; Loss of or Failure to Bond
DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: A (B)(6) OLD MALE PATIENT UNDERWENT A TVAR PROCEDURE IN WHICH THE ZENITH ALPHA THORACIC ENDOVASCULAR GRAFT PROXIMAL COMPONENTS, G35367, WAS USED. THE LAST GRAFT PIECE, (B)(4), WAS USED FOR BRIDGING BETWEEN TWO GRAFTS AS THE PROXIMAL GRAFT WAS TOO SHORT WITH THE DISTAL PIECE. WHEN THE PHYSICIAN FINISHED AND ROTATED THE DELIVERY SYSTEM THE NOSE CONE OF THE DEVICE INNER WHITE CANNULA CAME OFF AND STAYED ON THE WIRE END OF THE GRAFT. LEAVING THE NOSE CONE AND INNER CANNULA IN THE PATIENT. THE PHYSICIAN THEN ADVANCED A LONGER AND LARGER SHEATH AND SNARED THE SHEATH TO REMOVE FROM THE PATIENT. ADDITIONAL INFORMATION RECEIVED 20MAR2020: AFTER RELEASING THE STENT GRAFT, THE PHYSICIAN RETRIEVED THE INNER SYSTEM AND LEFT THE OUTER SHEATH IN ORDER TO PERFORM BALLOONING. WHEN THE INNER SYSTEM WAS TAKEN OUT, IT WAS NOTED THAT THE SYSTEM ONLY CONTAINED THE GREY POSITIONER ALONG WITH THE HANDLE AND THE TIP ASSEMBLY (DILATOR TIP AND CANNULA) WAS LEFT IN PATIENT. THERE WAS VERY TORTUROUS ANATOMY AND WE WERE WELL IN THE ARCH AROUND ZONE 2. SOME DIFFICULTY RETRIEVING DEVICE OVER THE ARCH BUT AS EXPECTED WITH THE ANATOMY AND SIZE OF ANEURYSM. PATIENT
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9909215 ↗MalfunctionMar 6, 2018Activation Failure; Activation Failure
DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: THEY PUT THE DEVICE IN THE PATIENT AND DECIDED THEY WANTED A LONGER PIECE AND TOOK IT OUT. THEY THOUGHT IT WOULD BE GOOD TO USE THE DEVICE AS A TEACHING OPPORTUNITY. THE PHYSICIAN DEPLOYED PRODUCT ON THE BACK TABLE. AFTER ALL TRIGGER WIRES WERE REMOVED, THE DEVICE REMAINED CONSTRAINED. THERE WAS A SUTURE THAT SHOULD PRE-FORM; HOWEVER, IT DID NOT EXPAND. IT WAS CONSTRAINED. PATIENT OUTCOME: THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7315672 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P140016 on Sep 15, 2015. Submitted by Cook Medical Incorporated as “Zenith Alpha Thoracic Endovascular Graft”. The FDA review took 398 days. The FDA has since approved 18 supplements (changes to the device or its labeling), most recently on Aug 4, 2026. FDA record ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Recalls under the same FDA submission
These recalls involve devices cleared under P140016. One submission often covers many models, so check the product and lot details in each record.
- Z-0043-2026 · Sep 18, 2025 · Open, ClassifiedZenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W ZTA2-P-40-217-W ZTA2-P-42-121 ZTA2-P-42-121-W
- Z-2099-2017 · Mar 22, 2017 · TerminatedZenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with isolated lesions of the descend
Common questions about the Zenith Alpha G35367
Has the Zenith Alpha G35367 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10827002353678), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the Zenith Alpha G35367?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Mar 6, 2018 and Apr 8, 2020. In those reports, the most common specific device problems are “Loss of or Failure to Bond” (2) and “Activation Failure” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Zenith Alpha G35367?
None of the reports are classified as deaths, 2 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Zenith Alpha G35367 FDA approved?
It was approved through premarket approval (PMA) P140016 on Sep 15, 2015.
Is the Zenith Alpha G35367 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Zenith Alpha G35367, and is it still sold?
The FDA's device database lists William Cook Europe ApS as the company responsible for it (brand Zenith Alpha, model G35367). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths253 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths212 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths137 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths43 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page