Zenith Alpha G35404Recalls & reported problems
William Cook Europe ApS · Zenith Alpha, Thoracic Endovascular Graft, Distal Component · Class 3 · Implant
All models of Zenith Alpha (67)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10827002354040), received between Sep 14, 2017 and Sep 14, 2017 — classified as 1 malfunction.
The most frequently reported specific device problem is “Difficult or Delayed Positioning” (1 report).
2 other recalls cover devices cleared under the same FDA submission (P140016); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Difficult or Delayed Positioning1
- Kinked1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2017Difficult or Delayed Positioning; Kinked; Difficult or Delayed Positioning; Kinked
DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: "THE TRIGGER WIRE AT THE TRI-FOLD OF THE DISTAL PIECE, DID NOT RELEASE. UPON REMOVING THE DELIVERY SYSTEM THE DEVICE IN-FOLDED. PERFORMED A FOLLOW UP WITH THE PHYSICIAN ON (B)(6) 2017 AND THE PHYSICIAN INDICATED THE PATIENT IS DOING WELL." ADDITIONAL INFORMATION RECEIVED 01SEP2017: "THE ISSUE WAS NOT WITH THE BARE STENT OF THE DISTAL GRAFT AT ALL. AFTER THE BARE STENT IS DEPLOYED, THE BLUE HANDLE THEN IS ABLE TO ROTATE AGAIN TO RELEASE THE REMAINING TRIGGER WIRES AT THE TOP OF THE GRAFT (THEY HOLD THE TOP OF THE DISTAL GRAFT IN A CONSTRAINED FORM). THE SURGEON ROTATED THE HANDLE AND SAID HE FELT A STOP. HE CHOSE NOT TO VIEW THIS STEP UNDER FLUOROSCOPY. THE NEXT STEP THEN IS TO REMOVE THE DILATOR. WHEN THE SURGEON BEGAN REMOVING THE DILATOR HE DID NOT REALIZE THE TRIGGER WIRE WAS NOT RELEASED; EFFECTIVELY PULLING THE TOP OF THE GRAFT TO MEET THE BOTTOM OF THE GRAFT. ONCE HE SAW THE GRAFT HAD BEEN INFOLDED, HE WAS ABLE TO TURN THE HANDLE MORE. THIS ACTION RELEASED THE TRIGGER WIRES, AND WE WERE ABLE TO REMOVE THE DILATOR SO TROUBLESHOOTING TO REMOVE THE TRIGGER WIRES WAS NOT NECESSARY." PATIENT OUTCOME: THE PATIENT DID NOT REQUIRE A
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6864165 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P140016. One submission often covers many models, so check the product and lot details in each record.
- Z-0043-2026 · Sep 18, 2025 · Open, ClassifiedZenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W ZTA2-P-40-217-W ZTA2-P-42-121 ZTA2-P-42-121-W
- Z-2099-2017 · Mar 22, 2017 · TerminatedZenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with isolated lesions of the descend
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page