Zenith Alpha G38180Recalls & reported problems
William Cook Europe ApS · Zenith Alpha, Thoracic Endovascular Graft, Proximal Tapered Component · Class 3 · Implant
All models of Zenith Alpha (67)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10827002381800), received between Apr 20, 2016 and Oct 22, 2024 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (P140016); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Failure2
- Physical Resistance/Sticking2
- Difficult or Delayed Activation1
- Failure to Advance1
- Improper or Incorrect Procedure or Method1
- Leak/Splash1
- Off-Label Use1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 22, 2024Improper or Incorrect Procedure or Method; Difficult or Delayed Activation; Improper or Incorrect Procedure or Method; Difficult or Delayed Activation
DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: A FEMALE PATIENT OF UNDISCLOSED AGE UNDERWENT A THORACIC ENDOVASCULAR AORTIC REPAIR (TEVAR) DESCENDING THORACIC AORTA PROCEDURE IN WHICH THE ZENITH ALPHA THORACIC ENDOVASCULAR GRAFT PROXIMAL TAPERED COMPONENTS, G38180, WAS USED. AFTER UNSHEATHING GRAFT AND DISTAL TRIGGER WIRE RELEASE FROM GRAFT, DISTAL END OF GRAFT DID NOT OPEN COMPLETELY. BALLOON WAS PLACED IN DISTAL PART OF GRAFT AND INFLATED TO OPEN GRAFT COMPLETELY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20503523 ↗InjuryFeb 2, 2024Activation Failure; Activation Failure
DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER. A PATIENT OF UNDISCLOSED GENDER AND AGE UNDERWENT A TEVAR (THORACIC ENDOVASCULAR AORTIC REPAIR) PROCEDURE IN WHICH THE ZENITH ALPHA THORACIC ENDOVASCULAR GRAFT PROXIMAL TAPERED COMPONENTS, WAS USED. WHEN THE GRAFT WAS UNSHEATHED IN THE PATIENT THE THIRD STENT FROM THE TOP WOULD NOT OPEN FULLY. IT WAS CONSTRAINED ABOUT FIFTY PERCENT. THE SURGEON USED A CODA BALLOON TO OPEN THE STENT SUCCESSFULLY. IT WAS A DISSECTION, PATIENT HAD SOME TEARS IN THEIR AORTA PATIENT OUTCOME: DISTRICT MANAGER CONTACTED THE SURGEON ONE DAY POST-OP AND SURGEON CONFIRMED THE PATIENT WAS DOING WELL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18630473 ↗InjuryJan 30, 2024Off-Label Use; Activation Failure; Off-Label Use; Activation Failure
NO ADDITIONAL INFORMATION, REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED, SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18605230 ↗MalfunctionMar 18, 2021Physical Resistance/Sticking; Physical Resistance/Sticking
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11515020 ↗MalfunctionMar 12, 2020Use of Device Problem; Failure to Advance; Physical Resistance/Sticking; Use of Device Problem; Failure to Advance; Physical Resistance/Sticking
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 9823143 ↗Apr 20, 2016Leak/Splash
DESCRIPTION OF EVENT ACCORDING TO COMPLAINANT: "PROXIMAL ZENITH ALPHA DEVICE IMPLANTED WAS A ZTA-PT-38-34-167-W, DISTAL COMPONENT WAS A ZTA-P-24-105-W. DISTAL COMPONENT IMPLANTED FIRST TO COMPENSATE FOR DIAMETERS ON OVERLAP. APPROXIMATE OVERLAP BETWEEN THE TWO GRAFTS WAS 7CM, OR A LITTLE MORE THAN 3 STENTS. UPON IMPLANTATION, CODA BALLOONING WITH A 40MM CODA BALLOON WAS PERFORMED TO A NOMINAL PRESSURE THROUGHOUT THE DEVICE, IN ALL SEGMENTS. UPON COMPLETION OF BALLOONING, A DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) WAS PERFORMED WITH A PIGTAIL FLUSH CATHETER, FLUSHPORTS JUST PROXIMAL TO THE FABRIC. AN EARLY DETECTION OF AN ENDOLEAK WAS PRESENT IN THE ANEURYSM SAC, HOWEVER ITS SOURCE WAS INDETERMINATE. WE REBALLOONED THE ENTIRE DEVICE, THIS TIME APPLYING MORE THAN NOMINAL PRESSURE IN THE MOST PROXIMAL ZONE, OVERLAP JUNCTION, AND DISTAL ZONE, AS THOSE WERE THE MOST LIKELY CAUSES OF AN EARLY ONSET ENDOLEAK. ADDITIONAL DSA REVEALED THE SAME ENDOLEAK. CONFIDENT THAT THE LEAK WAS NOT A 1A ENDOLEAK AS THE DEVICE WAS SITTING NICELY IN THE PROXIMAL SEGMENT, WE MOVED THE PIGTAIL CATHETER INTO THE MIDDLE OF THE DEVICE, AS WE THOUGHT THE SOURCE COULD BE A TYPE 3 DUE TO THE OVERLAP JUNCTION, PRIMARI
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5590097 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P140016. One submission often covers many models, so check the product and lot details in each record.
- Z-0043-2026 · Sep 18, 2025 · Open, ClassifiedZenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W ZTA2-P-40-217-W ZTA2-P-42-121 ZTA2-P-42-121-W
- Z-2099-2017 · Mar 22, 2017 · TerminatedZenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with isolated lesions of the descend
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page