Skip to content

Zenith Alpha G38191Recalls & reported problems

William Cook Europe ApS · Zenith Alpha, Thoracic Endovascular Graft, Proximal Tapered Component · Class 3 · Implant

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002381916), received between Jun 9, 2021 and Mar 14, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (P140016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam2
  • Difficult or Delayed Separation1
  • Failure to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 14, 2024Mechanical Jam; Mechanical Jam

    DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: THE ATTACHED GRAFT HAD ISSUES WITH DEPLOYMENT. WE MOST DEPLOYED THIS GRAFT FIRST AND THE HANDLE DID NOT CLICK WHEN TURNING (HAN SAID THE HANDLE WAS DISASSOCIATED). SO HE HAD TO POP THE BACK OF THE HANDLE OFF AND PULL THE TRIGGER WIRES."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18901617 ↗
  • MalfunctionJun 13, 2022Mechanical Jam; Difficult or Delayed Separation; Mechanical Jam; Difficult or Delayed Separation

    DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: A 74 YEAR OLD FEMALE PATIENT UNDERWENT A TVAR PROCEDURE IN WHICH THE ZENITH ALPHA THROACIC ENDOVASCULAR GRAFT PROXIMAL TAPERED COMPONENTS, G38191, WAS USED. THE PROFORM WIRE DID NOT DETACH FROM IMPLANT. THE DISTRICT MANAGER AND SURGEON BROKE APART THE HANDLE TO MANUALLY RELEASE THE WIRE AND GRAFT. THE GRAFT HAD PULLED BACK FROM INTENDED SPOT OF DEPLOYMENT AS THE PHYSICIAN HAD THOUGHT THE GRAFT HAD DETACHED FROM THE DELIVERY SYSTEM. AT THIS TIME THERE IS NO FOLLOW UP PROCEDURE BEING SCHEDULED TO PLACE A GRAFT IN THE AREA ABOVE THE IMPLANTED GRAFT, HOWEVER, THE SURGEON STATED AT SOME POINT IN THE FUTURE ANOTHER PROCEDURE WILL NEED TO BE DONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14674165 ↗
  • MalfunctionJun 9, 2021Failure to Advance; Failure to Advance

    ACCORDING TO INITIAL REPORTER: I DID A CASE WITH A PHYSICIAN AND OUR ALPHA THORACIC DEVICE WOULD NOT TRACK DUE TO THE LENGTH OF THE NOSE CONE BEING PROBLEMATIC, SPECIFICALLY IN THIS PATIENT¿S ANATOMY. IT WAS THE 2ND DEVICE THAT WOULDN¿T TRACK. ADDITIONAL INFORMATION RECEIVED ON 01JUN2021: LANDED AN ALPHA GRAFT AT THE LEFT SUBCLAVIAN ARTERY. THIS 2ND PIECE WOULD NOT TRACK THROUGH THE FIRST GRAFT DUE TO THE NOSE CONE LENGTH COMBINED WITH THE PATIENT'S TORTUOUS ANATOMY. ADDITIONAL INFORMATION RECEIVED ON 02JUN2021: THE FIRST DEVICE MENTIONED IN THE INITIAL E-MAIL WAS THE OTHER DEVICE INVOLVED IN THIS PROCEDURE, AND IT HAD NO PROBLEM. TO CLARIFY, ONLY THE SECOND DEVICE DID NOT TRACK. ADDITIONAL INFORMATION RECEIVED 03JUN2021: THE FIRST DEVICE WAS IMPLANTED WITH NO ISSUE AT THE LEFT SUBCLAVIAN ARTERY. THE 2ND DEVICE WOULD NOT TRACK THROUGH THE FIRST DEVICE DUE TO THE NOSE CONE IN COMBINATION WITH THE PATIENTS ANATOMY. THE ASCENDING AORTA WAS SHORT, THE ARCH WAS TIGHT AND THE FIRST STENT BEING IN PLACE TOOK AWAY THE ABILITY TO ADVANCE THE DEVICE. PATIENT HAD 2CM SEAL ZONE DISTAL TO THE LEFT SUBCLAVIAN ARTERY AND THEN A LARGE ANEURYSM IMMEDIATELY AFTER. PLAN WAS TO SEAL PROXIMALLY WITH A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11964316 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P140016. One submission often covers many models, so check the product and lot details in each record.

  • Z-0043-2026 · Sep 18, 2025 · Open, Classified
    Zenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W ZTA2-P-40-217-W ZTA2-P-42-121 ZTA2-P-42-121-W
  • Z-2099-2017 · Mar 22, 2017 · Terminated
    Zenith Alpha Thoracic Endovascular Graft The Zenith Alpha" Thoracic Endovascular Graft is indicated for the endovascular treatment of patients with isolated lesions of the descend

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.