Zilver PTX G38479Recalls & reported problems
Cook Ireland Ltd · Zilver PTX, Drug-Eluting Peripheral Stent · Class 3 · Implant
All models of Zilver PTX (28)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10827002384795), received between Aug 2, 2017 and Apr 13, 2022 — classified as 5 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture3
- Activation, Positioning or Separation Problem2
- Delivery System Failure2
- Structural Problem2
- Adverse Event Without Identified Device or Use Problem1
- Deformation Due to Compressive Stress1
- Device Contamination with Chemical or Other Material1
- Difficult to Remove1
Patient problems
- No Consequences Or Impact To Patient5
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient2
- Disability1
- Hypovolemia1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryApr 13, 2022Fracture; Structural Problem; Fracture; Structural Problem
NOTED DURING ROUTINE FOLLOW UP ULTRASOUND: PATIENT IS ASYMPTOMATIC. STENT APPEARED FRACTURED; CAT SCAN SHOWS SEPARATED FRAGMENTS WHICH HAVE CREATED A PSEUDOANEURYSM IN THE PROXIMAL SFA. STENT WAS PLACED IN 2019 DURING A AAA (STENT GRAFT). IMAGINE REVIEW COMPLETED ON 06-APR-2022: IMPRESSION: THE IMAGES CONFIRM A TYPE III STENT FRACTURE AND PSEUDOANEURYSM. THE IMAGES DO NOT CONFIRM THAT THE FRACTURE CAUSED THE PSEUDOANEURYSM BECAUSE THE PSEUDOANEURYSM PRECEDED THE FRACTURE. THE PROXIMAL RIGHT SFA STENT IMPLANTATION LOCATION WOULD HAVE SUBJECTED THE STENT TO REPETITIVE STRESS FROM HIP FLEXION. BY REMOVING VESSEL WALL SUPPORT FROM THE STENT, THE PSEUDOANEURYSM INCREASED THE REPETITIVE STRESS EXERTED ON THE STENT DURING HIP FLEXION. WAS THE DEVICE FLUSHED BEFORE THE PROCEDURE, AS PER IFU? N/A, YES, NO. WERE THERE ANY ISSUES WITH FLUSHING OF THE DEVICE? N/A, YES, NO. DETAILS OF THE ACCESS SHEATH USED (NAME, FR SIZE,LENGTH)? DETAILS OF THE WIRE GUIDE USED (NAME, DIAMETER, HYDROPHYLIC)? WHAT APPROACH WAS USED TO ACCESS THE TARGET SITE? CONTRALATERAL, IPSILATERAL, ANTEGRADE, RETROGRADE, OTHER. PLEASE SPECIFY FOR OTHER: IF CONTRALATERAL, WAS THE BIFURCATION ANGLE STEEP? N/A, YES, NO. WHAT WA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14102365 ↗InjuryFeb 25, 2022Fracture; Fracture
REPORTED BY THE DM VIA PHONE CALL THE PHYSICIAN DID A ROUTINE FOLLOW UP ON THE PATIENT, DURING THE ULTRASOUND IT WAS DISCOVERED HE DEVICE WAS FRACTURED IN HALF. THE PHYSICIAN HAS INTENTIONS TO COVER WITH A NEW COVERED STENT. THE FOLLOWING HAS BEEN ANSWERED- (B)(6), 28JAN2022. DID ANY UNINTENDED SECTION OF THE DEVICE REMAIN INSIDE THE PATIENT¿S BODY? NO IF YES, PLEASE DESCRIBE. DID THE PATIENT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE? NOT YET DID THE PRODUCT CAUSE OR CONTRIBUTE TO THE NEED FOR ADDITIONAL PROCEDURES? YES IF YES, PLEASE SPECIFY ADDITIONAL PROCEDURES AND PROVIDE DETAILS. ADDITIONAL PROCEDURE INFORMATION IS NOT AVAILABLE CURRENTLY. HAS THE COMPLAINANT REPORTED ANY ADVERSE EFFECTS ON THE PATIENT DUE TO THIS OCCURRENCE? NO. HAS THE COMPLAINANT REPORTED THAT THE PRODUCT CAUSED OR CONTRIBUTED TO THE ADVERSE EFFECTS? NO. PLEASE SPECIFY ADVERSE EFFECTS AND PROVIDE DETAILS. WAS THE PATIENT HOSPITALIZED OR WAS THERE PROLONGED HOSPITALIZATION DUE THIS OCCURRENCE? NO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13602506 ↗InjuryFeb 22, 2022Fracture; Structural Problem; Fracture; Structural Problem
REPORTED BY THE DM VIA PHONE CALL THE PHYSICIAN DID A ROUTINE FOLLOW UP ON THE PATIENT, DURING THE ULTRASOUND IT WAS DISCOVERED HE DEVICE WAS FRACTURED IN HALF. THE PHYSICIAN HAS INTENTIONS TO COVER WITH A NEW COVERED STENT. THE FOLLOWING HAS BEEN ANSWERED: (B)(6), (B)(6) 2022. DID ANY UNINTENDED SECTION OF THE DEVICE REMAIN INSIDE THE PATIENT¿S BODY? NO. IF YES, PLEASE DESCRIBE. DID THE PATIENT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE? NOT YET. DID THE PRODUCT CAUSE OR CONTRIBUTE TO THE NEED FOR ADDITIONAL PROCEDURES? YES. IF YES, PLEASE SPECIFY ADDITIONAL PROCEDURES AND PROVIDE DETAILS. ADDITIONAL PROCEDURE INFORMATION IS NOT AVAILABLE CURRENTLY. HAS THE COMPLAINANT REPORTED ANY ADVERSE EFFECTS ON THE PATIENT DUE TO THIS OCCURRENCE? NO. HAS THE COMPLAINANT REPORTED THAT THE PRODUCT CAUSED OR CONTRIBUTED TO THE ADVERSE EFFECTS? NO. PLEASE SPECIFY ADVERSE EFFECTS AND PROVIDE DETAILS. WAS THE PATIENT HOSPITALIZED OR WAS THERE PROLONGED HOSPITALIZATION DUE THIS OCCURRENCE? NO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13578290 ↗MalfunctionApr 24, 2019Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
EVENT IS FDA MDR REPORTABLE BASED ON THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF ¿STENT SHORTENING/ COMPRESSION DURING DEPLOYMENT'. AS REPORTED TO CUSTOMER RELATIONS: "PHYSICIAN TOLD ME THE PATIENT WAS BROUGHT BACK TO THE FACILITY AFTER A FIRST PROCEDURE--THE FIRST TIME, THEY PLACED A 7MM ZILVER PTX IN THE SFA. THERE WAS A DISSECTION THAT HAD ALREADY OCCURRED THAT THE PHYSICIAN WANTED TO FIX, SO HE PUT A BALLOON IN, PRIOR TO PUTTING IN A 6X100 PTX. WANTED TO LAY ANOTHER 6X60 AFTER THAT, AND WHEN HE WENT TO DEPLOY THE 6X60 HE SAID THE STENT BUNCHED UP ON ITSELF, LEADING TO A SHORTENING OF ABOUT 2/3 THE LENGTH (SO 6X60 LOOKED LIKE A 6X20).
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8544512 ↗InjuryFeb 14, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2018 AN ARTERIOGRAM OF THE LOWER LEG WITH BALLOON ANGIOPLASTY AND STENT PLACEMENT WAS PERFORMED AND 5 COOK STENTS WERE PLACED. NO FLOW WAS NOTED, AND THE PHYSICIAN INDICATED THEY DID NOT BELIEVE IT WAS DUE TO THE DEVICES. ON (B)(6) 2019 THE DM FOLLOWED UP TO CHECK ON THE PATIENT, AND THE PHYSICIAN INFORMED THE DM THAT THE PATIENT HAD TO HAVE AMPUTATION (UNKNOWN LOCATION). PATIENT IS STABLE AND HEALING WELL.
Patient problems: Disability; Hypovolemia
MAUDE report 8339320 ↗InjuryJan 10, 2019Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
DURING DEPLOYMENT OF THE STENT IN THE PROXIMAL POPLITEAL, WHEN THE STENT WAS APPROXIMATELY A QUARTER OF THE WAY DEPLOYED, THE THUMB WHEEL BROKE. THE USER ATTEMPTED TO CRACK OPEN THE HANDLE TO MANUALLY DEPLOY THE STENT, BUT THIS WAS UNSUCCESSFUL. THE USER PULLED BACK ON THE SYSTEM AND THE STENT DEPLOYED. THE STENT DEPLOYED INSIDE PREVIOUSLY PLACED STENT, WHICH WAS NOT THE INTENDED TARGET LOCATION FOR THE STENT PLACEMENT. THE USER BALLOONED THE STENT TO SECURE IT AND THEN PROCEED TO PLACE A ZILVER 518 TO COMPLETE THE INTENDED PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8235354 ↗MalfunctionDec 29, 2017Delivery System Failure; Delivery System Failure
THIS FOLLOW UP REPORT IS BEING SUBMITTED DUE TO THE EVALUATION OF THE DEVICE INVOLVED IN THIS EVENT AND THE CONCLUSION OF THIS INVESTIGATION. INITIAL REPORT DETAILS:THE PHYSICIAN WAS DEPLOYING THE STENT AND AS IT WAS ABOUT 2/3 DEPLOYED, THE THUMBWHEEL STOPPED TURNING. HE WAS ABLE TO BREAK OPEN THE HANDLE AND PULLED ON SOME OF THE WIRES AND DEPLOYED THE REST OF THE WAY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7153824 ↗MalfunctionSep 4, 2017Delivery System Failure; Activation, Positioning or Separation Problem; Delivery System Failure; Activation, Positioning or Separation Problem
AS THE FACILITY WAS TURNING THE WHEEL AND RETRACTING THE COVER OF THE STENT, THE CABLE BROKE. ONLY ABOUT 2CM'S OF THE DEVICE WAS DEPLOYED IN THE PATIENT. AT THIS POINT THE WHEEL WOULD TURN FREELY AS IT FELT LIKE IT WAS NOT CONNECTED TO ANYTHING. TO GET THE STENT OFF THE DEPLOYMENT SYSTEM, THE PHYSICIAN PULLED THE UNIT BACK AND THE STENT DEPLOYED ON ITS OWN. THE STENT DID ELONGATE PASSED THE OSTEUM OF THE SFA INTO THE PATIENT'S COMMON FEMORAL.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6841017 ↗MalfunctionAug 23, 2017Difficult to Remove; Difficult to Remove
THIS REPORT FORMS PART OF A RETROSPECTIVE REVIEW OF OPEN COMPLAINTS FOR A NEW MALFUNCTION PRECEDENCE ESTABLISHED FOR THIS DEVICE FAMILY. THE CUSTOMER WAS USING 6 FR BALKAN SHEATH WITH EXTRA STIFF AMPLATZ WIRE WITH A 6 FR ZILVER PTX. WHEN THEY PUT THE PTX OVER THE WIRE, IT WAS QUITE STIFF. ONCE THEY DEPLOYED THE STENT AND TRIED TO GET IT OFF THE WIRE, IT WAS STUCK/JAMMED ON THE WIRE. THEY HAD TO TAKE EVERYTHING OUT. THEY WERE ABLE TO SUCCESSFULLY COMPLETE THE PROCEDURE WITH OTHER DEVICES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6812949 ↗MalfunctionAug 2, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
UPON OPENING THE PACKAGE, A CONTAMINANT WAS FOUND WITHIN THE STERILE PACKAGING. THE DEVICE WAS NOT USED / DID NOT MAKE PATIENT CONTACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6759720 ↗
Other Stent, superficial femoral artery, drug-eluting devices
- ELUVIA™ H74939294601210Boston Scientific Corporation · Stent, superficial femoral artery, drug-eluting1 death80 reports
- Zilver PTX G24890Cook Ireland Ltd · Stent, superficial femoral artery, drug-eluting1 recall24 reports
- ELUVIA™ H74939294701210Boston Scientific Corporation · Stent, superficial femoral artery, drug-eluting58 reports
- Zilver PTX G38482Cook Ireland Ltd · Stent, superficial femoral artery, drug-eluting57 reports
- Zilver PTX G24895Cook Ireland Ltd · Stent, superficial femoral artery, drug-eluting1 recall3 reports
- ELUVIA™ H74939294601510Boston Scientific Corporation · Stent, superficial femoral artery, drug-eluting50 reports
- Zilver PTX G38483Cook Ireland Ltd · Stent, superficial femoral artery, drug-eluting32 reports
- ELUVIA™ H74939294701510Boston Scientific Corporation · Stent, superficial femoral artery, drug-eluting31 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page