Skip to content

Zilver 518 G43780Recalls & reported problems

Cook Ireland Ltd · Zilver 518 Vascular Stent · Class 3 · Implant

All models of Zilver 518 (20)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10827002437804), received between Feb 8, 2024 and Feb 8, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Off-Label Use” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Off-Label Use1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 8, 2024Off-Label Use; Off-Label Use

    FORESHORTENED 30 TO 20, DR MEASURED. PATIENT OUTCOME: NO PATIENT IMPACT REPORTED SG ON (B)(6) 2024. PATIENT/EVENT INFO - NOTES: THE FOLLOWING INFORMATION HAS BEEN RECEIVED VIA PHONE CALL ON (B)(6) 2024. SG (B)(6) 2024 3.32 ARE IMAGES OF THE DEVICE OR PROCEDURE AVAILABLE? NO 3.33 AT WHAT STAGE OF THE PROCEDURE DID THE COMPLAINT OCCUR? REMOVING THE INTRODUCER. 3.34 DETAILS OF ACCESS SHEATH USED (NAME, FR SIZE, LENGTH)? BMX PENUMBRA 3.35 WHAT WAS THE TARGET LOCATION FOR THE STENT? INTRACRANIAL VESSEL 3.36 WAS THE PRODUCT INSPECTED FOR KINKS OR DAMAGE BEFORE USE? UNKNOWN 3.37 WAS THE DEVICE USED PERCUTANEOUSLY? YES 3.38 WAS THE DEVICE FLUSHED THROUGH BOTH FLUSHING PORT BEFORE THE PROCEDURE, AS PER IFU? N/A 3.39 WAS PRE-DILATION PERFORMED AHEAD OF PLACEMENT OF THE STENT? N/A 3.40 WAS POST-DILATION PERFORMED AFTER THE PLACEMENT OF THE STENT? N/A 3.41 DETAILS OF THE WIRE GUIDE USED (NAME, DIAMETER, HYRDOPHYLLIC)? N/A 3.42 DID THE PATIENT EXHIBIT DIFFICULT OR ALTERED ANATOMY (IF ALTERED PLEASE SPECIFY HOW IT IS ALTERED)? FOCAL STENOSIS 3.43 WAS RESISTANCE ENCOUNTERED WHEN ADVANCING THE WIRE GUIDE TO THE TARGET LOCATION? NO 3.44 WAS RESISTANCE ENCOUNTERED WHEN ADVANCING THE DELIVERY SYSTEM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18666957 ↗

Other STENT, ILIAC devices

All STENT, ILIAC devices →