Skip to content

Zilver 518 G43783Recalls & reported problems

Cook Ireland Ltd · Zilver 518 Vascular Stent · Class 3 · Implant

All models of Zilver 518 (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10827002437835), received between Oct 28, 2021 and May 23, 2025 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Off-Label Use” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Off-Label Use8
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Foreign Body In Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 23, 2025Off-Label Use; Off-Label Use

    DEVICE WOULDN¿T TRACK OVER WIRE TO DESIRED LOCATION. THEN IT DID NOT FIT THROUGH A 071 ID BENCH TOP CATHETER. NO PATIENT IMPACT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22077321 ↗
  • MalfunctionJan 23, 2025Off-Label Use; Off-Label Use

    IT WAS REPORTED THAT A 8 BY 80 ZILVER 518 STENT DID NOT DEPLOY CORRECTLY DURING AN OFF-LABEL SINUS STENT AS PER THE DR. IT WAS STATED NEURO STARTED TO DEPLOY THE STENT, SHE MET RESISTANCE AT THE STENOSIS AS THE ARMADILLO KEPT PUSHING UP MAKING IT DIFFICULT TO CONTINUE TO DEPLOY. THERE WAS A LOT OF TENSION. SHE STOPPED DEPLOYING TO SEE SITUATION. BUT AS THE STENT CONTINUED TO UNSHEATH, IT STARTED TO STRETCH OUT THE STENT BEFORE IT WAS FULLY DEPLOYED AT DISTAL END. NOT FULLY DEPLOYED CORRECTLY. IT WAS REPORTED THAT "THE PATIENT IS FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21217721 ↗
  • MalfunctionJun 27, 2024Off-Label Use; Off-Label Use

    DESCRIPTION OF EVENT: THE PROXIMAL END OF THE STENT WAS NOT DEPLOYING AS IT SHOULD, THE HANDLE WAS BACK AS FAR AS IT COULD ON THE METAL CANNULA. THE PHYSICIAN WAS FINALLY ABLE TO GET THE DEVICE TO DEPLOY BY USING FORWARD AND BACKWARD MOTION OF THE SHEATH. PATIENT OUTCOME: NO PATIENT IMPACT REPORTED. SG 30MAY2024 PATIENT/EVENT INFO - NOTES: THE FOLLOWING INFORMATION HAS BEEN REQUESTED & RECEIVED VIA EMAIL ON 30MAY2024. SG 30MAY2024 PLEASE SEE AVAILABLE ANSWERS BELOW. ALL OTHERS UNKNOWN. 1. WILL THE FACILITY BE REPORTING THIS TO A GOVERNMENT AGENCY (FDA, COMPETENT AUTHORITY, ETC.)? NO. 2. CONFIRM LOT: C2143471 CORRECT . 3. WILL THE PRODUCT BE RETURNED? PARTIALLY. 1. IF YES PLEASE USE THE FEDEX RETURN LABEL ALREADY PROVIDED. ONLY THE DELIVERY DEVICE OF PATIENT #3 WILL BE RETURNED . 4. CAN YOU PROVIDE ANY PATIENT INFORMATION (AGE, WEIGHT, GENDER, PRE-EXISTING CONDITIONS)? NO. 5. CONFIRM THE DESCRIPTION OF EVENT: THE PROXIMAL END OF THE STENT WAS NOT DEPLOYING AS IT SHOULD, THE HANDLE WAS BACK AS FAR AS IT COULD ON THE METAL CANNULA. THE PHYSICIAN WAS FINALLY ABLE TO GET THE DEVICE TO DEPLOY BY USING FORWARD AND BACKWARD MOTION OF THE SHEATH. CORRECT . 6. DID ANY UNINTENDED SECTION OF T

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19622153 ↗
  • MalfunctionJun 6, 2024Off-Label Use; Off-Label Use

    A SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION ON 11-SEP-2024.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19481065 ↗
  • MalfunctionJun 5, 2024Off-Label Use; Off-Label Use

    DESCRIPTION OF EVENT: THE PROXIMAL END OF THE STENT WAS NOT DEPLOYING AS IT SHOULD, THE HANDLE WAS BACK AS FAR AS IT COULD ON THE METAL CANNULA. THE PHYSICIAN WAS FINALLY ABLE TO GET THE DEVICE TO DEPLOY BY USING FORWARD AND BACKWARD MOTION OF THE SHEATH. PATIENT OUTCOME: NO PATIENT IMPACT REPORTED. (B)(6) 2024. PATIENT/EVENT INFO - NOTES: THE FOLLOWING INFORMATION HAS BEEN RECEIVED VIA PHONE CALL ON 09MAY2024. (B)(6) 2024. 3.32 ARE IMAGES OF THE DEVICE OR PROCEDURE AVAILABLE? NO. 3.33 AT WHAT STAGE OF THE PROCEDURE DID THE COMPLAINT OCCUR? STENT PLACEMENT. 3.34 DETAILS OF ACCESS SHEATH USED (NAME, FR SIZE, LENGTH)? ROUTE 92 MEDICAL -8FR BASE .CAMP SHEATH SYSTEM. 3.35 WHAT WAS THE TARGET LOCATION FOR THE STENT? VENOUS SINUS. 3.36 WAS THE PRODUCT INSPECTED FOR KINKS OR DAMAGE BEFORE USE? YES. 3.37 WAS THE DEVICE USED PERCUTANEOUSLY? YES. 3.38 WAS THE DEVICE FLUSHED THROUGH BOTH FLUSHING PORT BEFORE THE PROCEDURE, AS PER IFU? YES. 3.39 WAS PRE-DILATION PERFORMED AHEAD OF PLACEMENT OF THE STENT? NO. 3.40 WAS POST-DILATION PERFORMED AFTER THE PLACEMENT OF THE STENT? NO. 3.41 DETAILS OF THE WIRE GUIDE USED (NAME, DIAMETER, HYRDOPHYLLIC)? BOSTON SCIENTIFIC V18 WIRE. 3.42 DID THE PATIENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19467342 ↗
  • MalfunctionMay 10, 2024Off-Label Use; Off-Label Use

    SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION ON 27-AUG-2024

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19284368 ↗
  • MalfunctionAug 18, 2022Off-Label Use; Off-Label Use

    SUPPLEMENTAL MDR BEING SUBMITTED DUE TO COMPLETION OF LAB EVALUATION ON 25-AUG-2022: (B)(4). USED. VISUAL INSPECTION - RED SAFETY TAB NOT RETURNED, THE BLUE OTHER SHEATH IS SEPERATED AT THE HANDLE. FUNCTIONAL - DEVICE FLUSHES WITHOUT ISSUE. A 0.018INCH WIRE GUIDE PASSES WITH NO ISSUE. (B)(4). USED. VISUAL INSPECTION - RED SAFETY TAB NOT RETURNED. BLUE SHEATH IS SEPERATED AT THE HANDLE. FUNCTIONAL - DEVICE FLUSHES WITHOUT ISSUE, A 0.018 INCH WIRE GUIDE PASSES WITH NO ISSUES. (B)(4). UNUSED. RETURNED IN ORIGINAL PACKAGING. VISUAL - RED SAFETY TAB RETURNED IN PLACE ON DEVICE. NO DAMAGE OBSERVED. FUNCTIONAL - DEVICE FLUSHES WITH NO ISSUE. 0.018INCH WIRE GUIDE PASSES WITH NO ISSUES. STENT DEPLOYS WITH NO ISSUE. (B)(4). UNUSED. RETURNED IN ORIGINAL PACKAGING. VISUAL - RED SAFETY TAB RETURNED IN PLACE ON DEVICE. NO DAMAGE OBSERVED. FUNCTIONAL - DEVICE FLUSHES WITH NO ISSUE. 0.018INCH WIRE GUIDE PASSES WITH NO ISSUES. STENT DEPLOYS WITH NO ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15248846 ↗
  • InjuryNov 12, 2021Break; Break

    "SOME PORTION OF THE DELIVERY SYSTEM BROKE OFF INTO THE PATIENT. WHEN IT BROKE OFF, THE DOCTOR FELT COMFORTABLE THAT THE PATIENT "WOULD BE FINE" AND DIDN'T ATTEMPT RETRIEVAL DURING THE PROCEDURE." LAB EVALUATION COMPLETED ON 26-NOV-2020, THE FOLLOWING OBSERVATIONS WERE NOTED: DEVICE COULD NOT BE FLUSHED. 0.018 INCH WIRE GUIDE ADVANCED THROUGH DEVICE WITH RESISTANCE. COULD NOT ADVANCE WIRE GUIDE BEYOND APPROX. 107CM. HANDLE MOVED TO HUB WITH A LOT OF RESISTANCE. INNER CATHETER PROTRUDES THROUGH THE SHEATH AT APPROX. 2.5CM FROM DISTAL END OF SHEATH WHEN HANDLE WAS MOVED TO HUB. PATIENT OUTCOME: DID ANY UNINTENDED SECTION OF THE DEVICE REMAIN INSIDE THE PATIENT¿S BODY? YES. IF YES, PLEASE DESCRIBE. SEE EVENT DESCRIPTION. WAS THE PATIENT HOSPITALIZED OR WAS THERE PROLONGED HOSPITALIZATION DUE TO THIS OCCURRENCE? IF YES, PLEASE DESCRIBE. DID THE PATIENT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE? IF YES, PLEASE DESCRIBE. DID THE PRODUCT CAUSE OR CONTRIBUTE TO THE NEED FOR ADDITIONAL PROCEDURES? IF YES, PLEASE SPECIFY ADDITIONAL PROCEDURES AND PROVIDE DETAILS. HAS THE COMPLAINANT REPORTED ANY ADVERSE EFFECTS ON THE PATIENT DUE TO THIS OCCURRENCE? HAS THE COMPLAINANT REPORTE

    Patient problems: Foreign Body In Patient

    MAUDE report 12802071 ↗
  • MalfunctionOct 28, 2021Off-Label Use; Off-Label Use

    THE CATHETER PORTION OF THE FLEXOR CAME DETACHED FROM THE DEPLOYMENT DEVICE. WHEN HE WENT TO PULL BACK TO DEPLOY IT, IT DETACHED. THE STENT WAS NOT ABLE TO BE DEPLOYED. THEY PULLED THE DELIVERY CATHETER OUT OVER THE WIRE. THE STENT WAS STILL IN THE CATHETER. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER DEVICE OF THE SAME TYPE. OFF LABEL USE: ATTEMPTED DELIVERY IN THE RIGHT SFA. PATIENT OUTCOME: DID ANY UNINTENDED SECTION OF THE DEVICE REMAIN INSIDE THE PATIENT¿S BODY? -NO IF YES, PLEASE DESCRIBE. WAS THE PATIENT HOSPITALIZED OR WAS THERE PROLONGED HOSPITALIZATION? -NO DID THE PATIENT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE? -NO IF YES, PLEASE DESCRIBE. DID THE PRODUCT CAUSE OR CONTRIBUTE TO THE NEED FOR ADDITIONAL PROCEDURES? -NO IF YES, PLEASE SPECIFY ADDITIONAL PROCEDURES AND PROVIDE DETAILS. HAS THE COMPLAINANT REPORTED ANY ADVERSE EFFECTS ON THE PATIENT DUE TO THIS OCCURRENCE? -NO HAS THE COMPLAINANT REPORTED THAT THE PRODUCT CAUSED OR CONTRIBUTED TO THE ADVERSE EFFECTS? -NO PLEASE SPECIFY ADVERSE EFFECTS AND PROVIDE DETAILS. GENERAL QUESTIONS FOR COMPLAINT OCCURRING DURING USE, REQUEST THE FOLLOWING: WHERE WAS THE ACCESS SITE? -LEFT GROIN WHAT WAS THE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12710554 ↗

Other STENT, ILIAC devices

All STENT, ILIAC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page