Skip to content

Zilver 518 G43793Recalls & reported problems

Cook Ireland Ltd · Zilver 518 Vascular Stent · Class 3 · Implant

All models of Zilver 518 (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002437934), received between Oct 31, 2019 and Dec 13, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem2
  • Fracture1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 13, 2023Fracture; Fracture

    AS REPORTED TO CUSTOMER RELATIONS VIA COMPLAINT EMAIL "I AM REPORTING A DEFECTIVE PRODUCT FOR UNIVERSITY OF (B)(6). THE EVENT TOOK PLACE ON 11/13 AND I WAS INFORMED ABOUT IT ON 11/16. THEY WILL NOT BE REPORTING THIS TO THE FDA, THEY DO NOT NEED A REPLACEMENT PRODUCT, YOU CAN SEND AN ACKNOWLEDGEMENT LETTER TO ME, AND I¿LL FORWARD IT ON TO THE ACCOUNT. THE PRODUCT IS NOT AVAILABLE FOR RETURN. (B)(6) G43798/C2073988 PHYSICIAN: DR. (B)(6). DESCRIPTION: ¿THE STENT BROKE¿ IF YOU CAN SEND ME THE FOLLOW-UP QUESTIONS, I WILL ASK THE CUSTOMER. I WAS NOT GIVEN MUCH DETAIL ABOUT THIS INCIDENT. ARE IMAGES OF THE DEVICE OR PROCEDURE AVAILABLE? N/A, YES, NO AT WHAT STAGE OF THE PROCEDURE DID THE COMPLAINT OCCUR? UNPACKING OR PREPARATION OF THE DEVICE, INSERTION, STENT PLACEMENT, REMOVING THE INTRODUCER. DETAILS OF ACCESS SHEATH USED (NAME, FR SIZE, LENGTH)? WHAT WAS THE TARGET LOCATION FOR THE STENT? WAS THE PRODUCT INSPECTED FOR KINKS OR DAMAGE BEFORE USE? N/A, YES, NO WAS THE DEVICE USED PERCUTANEOUSLY? N/A, YES, NO WAS THE DEVICE FLUSHED THROUGH BOTH FLUSHING PORT BEFORE THE PROCEDURE, AS PER IFU? N/A, YES, NO WAS PRE-DILATION PERFORMED AHEAD OF PLACEMENT OF THE STENT? N/A, YES, NO WAS POST-DI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18319680 ↗
  • MalfunctionFeb 27, 2020Use of Device Problem; Use of Device Problem

    THIS FOLLOW UP IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION. THE DELIVERY SYSTEM WAS ADVANCED FROM THE RADIAL ARTERY TO THE ILIAC ARTERY. HOWEVER, BECAUSE TORTUOSITY OF THE DESCENDING AORTA PREVENTED IT FROM ADVANCEMENT, IT COULD NOT REACH TO THE TARGET ILIAC ARTERY AND THE STENT COULD NOT BE PLACED. THE PROCEDURE WAS COMPLETED WITHOUT ANY FURTHER TREATMENT. THERE HAVE BEEN NO ADVERSE EFFECTS TO THE PATIENT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9763711 ↗
  • MalfunctionOct 31, 2019Use of Device Problem; Use of Device Problem

    DISTAL RADIAL APPROACH WAS GAINED. COMMAND WIRE GUIDE / ABBOTT AND PARENT PLUS 45 SHEATH / MEDIKIT WERE INSERTED AND THEN, ZIV5-18-125-10-80/ LOT C1309827 ADVANCED OVER THE WIRE GUIDE. DURING THE ADVANCEMENT, THE USER FELT RESISTANCE AND SUSPECTED STUCK OF DELIVERY SYSTEM. THE USER REMOVED THE DELIVERY SYSTEM AND THE WIRE GUIDE TOGETHER FOR PATIENT'S SAFETY. ANOTHER DEVICE WAS USED INSTEAD. THERE HAVE BEEN NO ADVERSE EFFECTS TO THE PATIENT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9261133 ↗

Other STENT, ILIAC devices

All STENT, ILIAC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page