Zilver 635 G43821Recalls & reported problems
Cook Ireland Ltd · Zilver 635 Vascular Stent · Class 3 · Implant
All models of Zilver 635 (55)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002438214), received between Nov 10, 2020 and Aug 18, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method1
- Off-Label Use1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 18, 2023Off-Label Use; Off-Label Use
ON (B)(6) 2023 AT THE (B)(6), ONE OF OUR CLINICIANS EXPERIENCED A DEVICE MALFUNCTION DURING DEPLOYMENT. NO PATIENT HARM OCCURRED, ACCORDING TO THE PHYSICIAN. WE ARE CONDUCTING AN INTERNAL INVESTIGATION INTO THIS INCIDENT AND CAN RETURN THE ITEM IN QUESTION UPON COMPLETION FOR YOUR CORROBORATING INVESTIGATION. I NOTICED THERE WAS A RECALL FOR THESE ITEMS IN APRIL 2013, BUT NO ACTIVE RECALLS FOR THIS ITERATION OF THE ITEM. PLEASE LET US KNOW IF YOU REQUIRE ANY ADDITIONAL INFORMATION, OR IF WE NEED TO COMPLETE ANY DOCUMENTATION. PER EMAIL ON 25JUL2023: DURING A LOWER EXTREMITY ANGIOGRAM, WE POSITIONED A STENT ACROSS A LESION AND DEPLOYED IT. WHEN WE WENT TO REMOVE THE DEPLOYMENT DEVICE, WE NOTED THE STENT MOVED WITH IT AND WAS NOT COMPLETELY DEPLOYED. WE IDENTIFIED THAT THE OUTER SHEATH THAT CONSTRAINS THE STENT AND IS SUPPOSED TO ROLL BACK WITH THE DEPLOYMENT HANDLE HAD SEPARATED FROM THE HANDLE, AND THE STENT WAS STILL CONSTRAINED. WE MANUALLY PULLED THE SHEATH BACK WITH A HEMOSTAT TO RELEASE THE STENT. WE THEN REMOVED THE ENTIRE DEPLOYMENT DEVICE OUT OF THE PATIENT. THE STENT REMAINED IN THE PATIENT, AS IT SHOULD. THE FOLLOWING INFORMATION HAS BEEN RECEIVED VIA EMAIL ON 25JUL2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17570947 ↗MalfunctionNov 10, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
SUPPLEMENTAL FOLLOW-UP REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION AND AN UPDATE TO THE INVESTIGATION CONCLUSIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10820545 ↗
Other STENT, ILIAC devices
- Cordis S.M.A.R.T. Control Vascular Stent System C10040MLCordis US Corp. · STENT, ILIAC1 recall3 reports
- Absolute Pro 1012534-100Abbott Vascular Inc. · STENT, ILIAC51 reports
- Protégé™ GPS™ SERB65-09-60-80Ev3, Inc · STENT, ILIAC1 recall0 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C09060MLCordis US Corp. · STENT, ILIAC1 recall0 reports
- Absolute Pro 1012535-100Abbott Vascular Inc. · STENT, ILIAC1 death41 reports
- Omnilink Elite 1012631-39Abbott Vascular Inc. · STENT, ILIAC27 reports
- Omnilink Elite 1012632-29Abbott Vascular Inc. · STENT, ILIAC25 reports
- Absolute Pro 1012536-100Abbott Vascular Inc. · STENT, ILIAC21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page