Skip to content

COOK G47451Recalls & reported problems

William Cook Europe ApS · Zenith TX2, Dissection Endovascular Graft with Pro-form, and the Z-Trak Plus Introduction System-Straight Component · Class 3 · Implant

All models of COOK (26)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002474519), received between Jan 27, 2022 and Feb 15, 2022 — classified as 2 injuries.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Physical Resistance/Sticking2

Patient problems

  • Hemorrhage/Blood Loss/Bleeding2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 15, 2022Physical Resistance/Sticking; Physical Resistance/Sticking

    DESCRIPTION OF EVENT ACCORDING TO THE INITIAL REPORTER: PT CAME IN EMERGENTLY DUE TO RUPTURE. (B)(6) (B)(6) 2022 ((B)(4)) THE FIRST DEVICE (G47452, LOT E4201640) DID NOT LAND IN THE RIGHT PLACE DUE TO OPERATOR ERROR. A LACK OF UPWARD FORCE IN HOLDING THE DEVICE INTO POSITION CAUSED THE PROBLEM. ON ((B)(6)) THEY HAD TO USE SECOND DEVICE (G47451, LOT E4202793) BECAUSE THE FIRST DEVICE WAS NOT IN THE PROPER PLACE. WHEN THEY TRIED TO USE THE SECOND DEVICE, THE TRIGGER WIRES COULD NOT BE PULLED OUT. THEY USED AS MUCH FORCE AS THEY COULD USE. THEY THEN INTENTIONALLY BROKE APART THE DEVICE TO GET AHOLD OF THE WIRES TO RELEASE THE GRAFT FROM THE DELIVERY SYSTEM. WHILE RETRACTING THE WIRES, THE GRAFT WAS PUSHED BACK OUT OF PLACE BY BLOOD FLOW FORCES. IT WAS NOT SITTING CORRECTLY WITHIN THE FIRST DEVICE. ((B)(4)) A THIRD DEVICE (G47470, LOT E3923818) WOULD NOT TRACK UP BECAUSE OF THE WAY THAT THE SECOND DEVICE WAS PLACED. THE PROCEDURE WAS THEN ABORTED. PATIENT OUTCOME: ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID EXPERIENCE AN ADVERSE EFFECTS DUE TO THIS OCCURRENCE. THE PATIENT LOST 1.5 UNITS OF BLOOD AND DID NOT ACHIEVE PROXIMAL SEAL. UNKNOWN IF THE PATIENT DID REQUIRE AN ADDITIONAL

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 13534327 ↗
  • InjuryJan 27, 2022Physical Resistance/Sticking; Physical Resistance/Sticking

    DESCRIPTION OF EVENT ACCORDING TO THE INITIAL REPORTER: PT CAME IN EMERGENTLY DUE TO RUPTURE. ON (B)(6) 2022 ((B)(4)) THE FIRST DEVICE ((B)(4), LOT E4201640) DID NOT LAND IN THE RIGHT PLACE DUE TO OPERATOR ERROR. A LACK OF UPWARD FORCE IN HOLDING THE DEVICE INTO POSITION CAUSED THE PROBLEM. ((B)(4)) THEY HAD TO USE SECOND DEVICE ((B)(4), LOT E4202793) BECAUSE THE FIRST DEVICE WAS NOT IN THE PROPER PLACE. WHEN THEY TRIED TO USE THE SECOND DEVICE, THE TRIGGER WIRES COULD NOT BE PULLED OUT. THEY USED AS MUCH FORCE AS THEY COULD USE. THEY THEN INTENTIONALLY BROKE APART THE DEVICE TO GET AHOLD OF THE WIRES TO RELEASE THE GRAFT FROM THE DELIVERY SYSTEM. WHILE RETRACTING THE WIRES, THE GRAFT WAS PUSHED BACK OUT OF PLACE BY BLOOD FLOW FORCES. IT WAS NOT SITTING CORRECTLY WITHIN THE FIRST DEVICE. ((B)(4)) A THIRD DEVICE ((B)(4), LOT E3923818) WOULD NOT TRACK UP BECAUSE OF THE WAY THAT THE SECOND DEVICE WAS PLACED. THE PROCEDURE WAS THEN ABORTED. PATIENT OUTCOME: ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID EXPERIENCE AN ADVERSE EFFECTS DUE TO THIS OCCURRENCE. THE PATIENT LOST 1.5 UNITS OF BLOOD AND DID NOT ACHIEVE PROXIMAL SEAL. UNKNOWN IF THE PATIENT DID REQUIRE AN ADDITIONAL PROCEDU

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 13368796 ↗

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →

Sources and updates