COOK G47456Recalls & reported problems
William Cook Europe ApS · Zenith TX2, Dissection Endovascular Graft with Pro-form, and the Z-Trak Plus Introduction System-Straight Component · Class 3 · Implant
All models of COOK (26)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002474564), received between May 21, 2020 and May 25, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Malposition of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Malposition of Device2
- Difficult to Remove1
- Mechanical Problem1
- Physical Resistance/Sticking1
Patient problems
- Failure of Implant1
- Hemorrhage/Blood Loss/Bleeding1
- No Code Available1
- Stroke/CVA1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 25, 2020Mechanical Problem; Malposition of Device; Physical Resistance/Sticking; Mechanical Problem; Malposition of Device; Physical Resistance/Sticking
ADDITIONAL INFORMATION RECEIVED FROM MEDWATCH (B)(4) ON (B)(6)2020 : DURING DEPLOYMENT OF THE DEVICE, THE GRAFT FULLY DEPLOYED BUT THE TRIGGER WIRE RELEASE CAP WOULD NOT COME FREE. THE SHEATH HAD TO BE CUT WHICH RESULTED IN LARGE VOLUME BLOOD LOSS WHILE THE TRIGGER WIRES WERE MANUALLY REMOVED TO RELEASE THE GRAFT FROM THE DELIVERY SYSTEM. THE WIRE BECAME STUCK IN THE DEVICE AND WIRE ACCESS TO THE VESSEL WAS LOST. THE LEFT GROIN CUT DOWN WAS PERFORMED AND PROXIMAL AND DISTAL CONTROL WAS ACHIEVED.
Patient problems: Failure of Implant; Hemorrhage/Blood Loss/Bleeding; Stroke/CVA
MAUDE report 10085009 ↗InjuryMay 21, 2020Difficult to Remove; Malposition of Device; Difficult to Remove; Malposition of Device
AS REPORTED "THE GREEN TRIGGER WIRE MECHANISM WAS UNABLE TO BE REMOVED FROM THE DELIVERY SYSTEM FOLLOWING DEPLOYMENT. THIS PREVENTED THE PHYSICIAN FROM BEING ABLE TO REMOVE THE DELIVERY SYSTEM AND REQUIRED THE TEAM TO TAKE MANUAL ACTION TO ACCESS AND REMOVE THE TRIGGER WIRES IN ORDER TO REMOVE THE DELIVERY SYSTEM. THE PHYSICIAN NEEDED TO CUT THE GREY INNER CANNULA IN ORDER TO ACCESS AND MANUALLY REMOVE THE INDIVIDUAL TRIGGER WIRES. THIS INTERVENTION WAS EFFECTIVE BUT INDUCED BLOOD LOSS THROUGH THE DELIVERY SYSTEM. WHILE ATTEMPTING TO REMOVE THE GREEN TRIGGER WIRE FROM THE TX2 DELIVERY SYSTEM, THE GRAFT WAS INADVERTENTLY REPOSITIONED SEVERAL MILLIMETERS RESULTING IN AN INADEQUATE PROXIMAL LANDING. AS A RESULT, AN ADDITIONAL ZDEG STENT GRAFT WAS REQUIRED TO ACHIEVE THE INTENDED PROXIMAL LANDING POSITION. THE DEPLOYMENT SYSTEM WAS SECURED IN A BIO-HAZARD BAG AND IT CURRENTLY BEING HELD AT THE HOSPITAL FACILITY."
Patient problems: No Code Available
MAUDE report 10080239 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P180001 on Dec 31, 2018. Submitted by William Cook Europe Aps as “Zenith Dissection Endovascular System”. The FDA review took 363 days. The FDA has since approved 17 supplements (changes to the device or its labeling), most recently on Aug 4, 2026. FDA record ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Common questions about the COOK G47456
Has the COOK G47456 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10827002474564), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the COOK G47456?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between May 21, 2020 and May 25, 2020. In those reports, the most common specific device problems are “Malposition of Device” (2), “Difficult to Remove” (1) and “Mechanical Problem” (1); the most common patient problems are “Failure of Implant” (1), “Hemorrhage/Blood Loss/Bleeding” (1) and “Stroke/CVA” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the COOK G47456?
None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the COOK G47456 FDA approved?
It was approved through premarket approval (PMA) P180001 on Dec 31, 2018.
Is the COOK G47456 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the COOK G47456, and is it still sold?
The FDA's device database lists William Cook Europe ApS as the company responsible for it (brand COOK, model G47456). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths253 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths212 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths137 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths43 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page