The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10827002474670), received between Feb 24, 2022 and May 1, 2024 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Entrapment of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Entrapment of Device2
- Off-Label Use1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Vascular Dissection2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 1, 2024
WHILE UNSHEATHING GRAFT TO BEGIN FENESTRATIONS, SURGEON HAD DIFFICULTY GETTING GRAFT TO COME OUT. WHEN GRAFT FINALLY UNSHEATHED, SURGEON NOTICED THAT PART OF DEPLOYMENT SYSTEM WAS BROKEN RENDERING SYSTEM UNUSABLE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19220094 ↗InjuryJun 16, 2022Entrapment of Device; Entrapment of Device
DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: PHYSICIAN MODIFIED GRAFT AND CREATED 4 FENESTRATIONS FOR THE SYLIAC SMA LEFT AND RIGHT RENAL. THEY DEPLOYED THE GRAFT AND WERE ABLE TO GET OUT THE FENESTRATIONS AND WERE ABLE TO PLACE STENTS OUT THE FENESTRATIONS. BUT THEY HAD A SMALL DISECTION IN THE SILLIAC WHICH THEY WERE ABLE TO STENT AND FIX. THEN ANOTHER DISECTION IN THE SMA (SUPERIOR MESENTERIC ARTERY) WHICH THEY WERE NOT ABLE TO FIX RIGHT AWAY AND HAD TO GO IN AND DO A BYPASS ON. PATIENT OUTCOME: THE EVENT DID REQUIRE AN ADDITIONAL PROCEDURE DUE TO THIS OCCURRENCE: BYPASS THE PATIENT DID NOT REPORT ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. THE PATIENT WAS HOSPITALIZED OR WAS THERE PROLONGED HOSPITALIZATION DUE THIS OCCURRENCE: THERE PROBABLY WILL BE A LONGER THAN EXPECTED HOSPITAL STAY.
Patient problems: Vascular Dissection
MAUDE report 14712212 ↗InjuryJun 1, 2022Entrapment of Device; Off-Label Use; Entrapment of Device; Off-Label Use
DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: PHYSICIAN MODIFIED GRAFT AND CREATED 4 FENESTRATIONS FOR THE SYLIAC SMA LEFT AND RIGHT RENAL. THEY DEPLOYED THE GRAFT AND WERE ABLE TO GET OUT THE FENESTRATIONS AND WERE ABLE TO PLACE STENTS OUT THE FENESTRATIONS. BUT THEY HAD A SMALL DISSECTION IN THE SILLIAC WHICH THEY WERE ABLE TO STENT AND FIX. THEN ANOTHER DISSECTION IN THE SMA (SUPERIOR MESENTERIC ARTERY) WHICH THEY WERE NOT ABLE TO FIX RIGHT AWAY AND HAD TO GO IN AND DO A BYPASS ON. PATIENT OUTCOME: THE EVENT DID REQUIRE AN ADDITIONAL PROCEDURE DUE TO THIS OCCURRENCE: BYPASS THE PATIENT DID NOT REPORT ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. THE PATIENT WAS HOSPITALIZED OR WAS THERE PROLONGED HOSPITALIZATION DUE THIS OCCURRENCE: THERE PROBABLY WILL BE A LONGER THAN EXPECTED HOSPITAL STAY.
Patient problems: Vascular Dissection
MAUDE report 14563966 ↗MalfunctionFeb 24, 2022Physical Resistance/Sticking; Physical Resistance/Sticking
DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: THE DELIVERY SYSTEM WAS PLACED IN AORTIC ARCH. THE PIN AND PULL STEP WAS PERFORMED WITHOUT EVENT. THE WHITE THUMB SCREW WAS UNSCREWED WITHOUT ISSUE. THE GREEN TRIGGER WIRE MECHANISM WAS ATTEMPTED TO BE PULLED TO RELEASE THE GRAFT. THE GREEN TRIGGER WIRE MECHANISM WOULD NOT MOVE AT ALL AND SEIZED ONTO THE DILATOR. MULTIPLE ATTEMPTS WERE MADE TO REMOVE THE GREEN TRIGGER WIRE MECHANISM WITH DIFFERENT USES OF LEVERAGE TO NO AVAIL. THEY THEN ATTEMPTED TO ROTATE THE GREEN TRIGGER WIRE MECHANISM TO RELEASE IT FROM THE DILATOR WITH NO SUCCESS. THE DILATOR WAS THEN UNSCREWED AND PULLED APART TO REVEAL THE TRIGGER WIRES. THE TRIGGER WIRES WERE THEN PULLED MANUALLY AND THE GRAFT WAS RELEASED BUT NOT BEFORE THE GRAFT WAS PULLED DISTALLY ABOUT 1 CM. A SECOND GRAFT WAS PLACED TO EXTEND THE SEAL PROXIMALLY. THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT. (B)(6) 2022 ADD INFO: PATIENT WAS HEPARINIZED. NO SIGNIFICANT CALCIFICATION OR TORTUOSITY. PATIENT OUTCOME: ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13589469 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.