COOK G47468Recalls & reported problems
William Cook Europe ApS · Zenith TX2, Dissection Endovascular Graft with Pro-form, and the Z-Trak Plus Introduction System-Tapered Component · Class 3 · Implant
All models of COOK (26)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10827002474687), received between Jun 21, 2023 and Jun 21, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Off-Label Use” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Off-Label Use1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 21, 2023Off-Label Use; Physical Resistance/Sticking; Off-Label Use; Physical Resistance/Sticking
DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: ZDEG TRIGGER WIRE RELEASE MECHANISM WOULD NOT RELEASE FROM DEVICE. TRIGGER WIRES COULD NOT INITIALLY BE REMOVED AFTER FULLY UNSHEATHING ZDEG INSIDE THE THORACIC AORTA. UNABLE TO CONFIRM NOR DENY THAT THE TRIGGER WIRE RELEASE MECHANISM HAD NOT BEEN ROTATED PRIOR TO ATTEMPTING TO REMOVE FROM TRIGGER WIRES. PHYSICIAN WAS ABLE TO PRY MECHANISM APART ENOUGH TO RELEASE WIRES AND SUCCESSFULLY DEPLOY THE DEVICE. ADDITIONAL INFORMATION RECEIVED (B)(6)2023 : THE PHYSICIANS DID USE (B)(4) SETS OF FORCEPS TO CORRECT THE STUCK RELEASE MECHANISM, BUT NOT TO PHYSICALLY REMOVE THE TRIGGER WIRES. THE TEAM ACTUALLY USED THE FORCEPS TO TRY AND SPREAD THE GREEN TRIGGER MECHANISM APART IN AN EFFORT TO ACCESS THE WIRES. THEY BROKE BOTH FORCEPS IN THE PROCESS. ALTHOUGH THE SCRUB TEAM SAID THE DOCTOR ALSO ¿BEAT THE DEVICE A LOT¿ MEANING THE GREEN MECHANISM. THE PHYSICAL MANIPULATION OF THE MECHANISM WAS ENOUGH TO RELEASE ENOUGH TENSION AND THE TEAM WAS ABLE TO REMOVE THE MECHANISM WITH ALL WIRES INTACT. THE ONLY WIRE THAT LOOKED DAMAGED WAS THE STAINLESS STEEL WIRE, WHICH WAS BENT IN A 90 DEGREE ANGLE. PATIENT OUTCOME: THE GRAFT WAS FULLY DEPLOYED AND LA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17168969 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page