Evolution G48033Recalls & reported problems
Cook Ireland Ltd · EVOLUTION ESOPHAGEAL, STENT SYSTEM - PARTIALLY COVERED · Class 2 · Implant
All models of Evolution (21)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 10827002480336), received between Apr 12, 2016 and Apr 6, 2026 — classified as 4 injuries, 15 malfunctions.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Activation, Positioning or Separation Problem9
- Improper or Incorrect Procedure or Method5
- Physical Resistance/Sticking3
- Activation Failure2
- Difficult to Remove1
- Fracture1
- Insufficient Device Problem Information1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- No Known Impact Or Consequence To Patient7
- Dysphagia/ Odynophagia1
- Excessive Tear Production1
- Insufficient Information1
- No Consequences Or Impact To Patient1
- Obstruction/Occlusion1
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 6, 2026Physical Resistance/Sticking
THE STENT ABRUPTLY OPENED AND COULD NOT BE RETRIEVED BACK. THE PROCEDURE WAS COMPLETED BY COMPETITOR DEVICE. WHAT WAS THE TARGET LOCATION? ESOPHGEAL. WHAT WAS THE DIAMETER OF THE WIRE GUIDE USED? 0.035X480 CMS. DID THE PATIENT EXHIBIT DIFFICULT ANATOMY (N/A, TORTUOUS, ALTERED)? NO. WERE ANY OTHER DEFECTS (OTHER THAN THE COMPLAINT ISSUE) OBSERVED ON THE DELIVERY SYSTEM (E.G., KINKS) PRIOR TO RETURN? NO. WAS RESISTANCE ENCOUNTERED WHEN ADVANCING THE DELIVERY SYSTEM TO THE TARGET LOCATION? NO. WAS PRE-DILATION/POST-DILATION PERFORMED? NO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24792624 ↗MalfunctionDec 20, 2024Material Twisted/Bent; Activation Failure; Material Twisted/Bent; Activation Failure
A SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO COMPLETION OF THE INVESTIGATION ON 29 MAY 2025.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20985887 ↗MalfunctionApr 24, 2024Improper or Incorrect Procedure or Method; Physical Resistance/Sticking; Improper or Incorrect Procedure or Method; Physical Resistance/Sticking
THE STENT COULD NOT BE DEPLOYED, AS THE TRIGGER GOT JAMMED. PATIENT/EVENT INFO - NOTES: QUERY REPLY: GENERAL QUESTIONS: 1. AT WHAT STAGE OF THE PROCEDURE DID THE COMPLAINT OCCUR? [WHEN UNPACKING OR PREPARING THE EVOLUTION, WHILE INSERTING THE EVOLUTION IN THE PATIENT, DURING STENT PLACEMENT, WHILE REMOVING THE INTRODUCER, OR DURING STENT REPOSITIONING/REMOVAL] ¿ DURING STENT PLACEMENT. 2. WHAT ENDOSCOPE TYPE AND CHANNEL SIZE WAS USED? 3. WHAT WAS THE POSITION OF THE ELEVATOR? [N/A, OPEN, CLOSED]. 4. DETAILS OF THE WIRE GUIDE USED (DIAMETER, TYPE, MAKE)? 0.035 INCH ¿ SGW GUIDE WIRE 5. WAS THE ZIP PORT FACING UPWARDS AND SLIGHTLY CURVED WHEN BACKLOADING THE WIRE GUIDE? [N/A, YES, NO]. 6. DID ANY PART OF THE STENT CONTACT THE PATIENT¿S ANATOMY WHEN THE COMPLAINT OCCURRED? [YES]. 7. PLEASE ADVISE THE ANATOMICAL LOCATION OF THE INTENDED TARGET SITE. ESOPHAGUS. 8. HOW LONG WAS THE STENT IN THE PATIENT BY THE TIME THIS COMPLAINT OCCURRED? THE CUSTOMER COULD NOT RECALL THE TIME, BUT SAID MINIMAL TIME. 9. FOR DEVICES WHERE THE IFU STATES FOR LONGER TERM PATENCY HAS NOT BEEN ESTABLISHED, WAS PERIODIC EVALUATION COMPLETED? [N/A]. 10. IF YES, HOW OFTEN WAS THIS COMPLETED? 11. DID THE PATIENT R
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19173182 ↗MalfunctionJun 9, 2023Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION ON THE 06-JUL-2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17096986 ↗InjuryMay 2, 2023Migration; Migration
MDR- 2054 106-007 AE1. MDR- 2054 106-007 AE1 - GASTROINTESTINAL: OTHER ; STENT DISLOCATION DISTAL, RELATIONSHIP: DEVICE , RELATED ; STENT DISLOCATION NEEDS TO REPOSITION, PROCEDURE: NON RELATED, PRE-EXISTING,: NO , DEFICIENCY: NO. 2 DAYS POST PROCEDURE. STENT DISLOCATION DISTAL - STENT REPOSITIONED ENDOSCOPICALLY. STATUS: RESOLVED, TREATMENT: ENDOSCOPIC, STUDY STENT REPOSITIONED, DEATH: NO.
Patient problems: No Clinical Signs, Symptoms or Conditions; Obstruction/Occlusion
MAUDE report 16852579 ↗MalfunctionSep 2, 2022Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
A SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO COMPLETION OF THE INVESTIGATION ON 28-FEB-2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15348054 ↗InjuryAug 15, 2022Improper or Incorrect Procedure or Method; Activation, Positioning or Separation Problem; Improper or Incorrect Procedure or Method; Activation, Positioning or Separation Problem
DOCTOR WAS DEPLOYING 15CM ESOPHAGEAL EVOLUTION PARTIALLY COVERED METAL STENT,AFTER DEPLOYMENT WHEN DOCTOR PULLED THE DELIVERY DEVICE THE PROXIMAL PART OF THE STENT BEND INSIDE THE STENT.DOCTOR REMOVED THE STENT WITH THE HELP OF THE RAT TOOTH FORCEP AND DEPLOYED ANOTHER STENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15224371 ↗MalfunctionJan 14, 2022Insufficient Device Problem Information; Activation Failure; Insufficient Device Problem Information; Activation Failure
AS PER SOURCE DOC: "CUSTOMER INDICATED THAT THEY COULDN'T PLACE THE PRODUCT".
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 13257073 ↗MalfunctionMay 12, 2021Activation, Positioning or Separation Problem; Physical Resistance/Sticking; Activation, Positioning or Separation Problem; Physical Resistance/Sticking
SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11812799 ↗MalfunctionFeb 5, 2021Structural Problem; Activation, Positioning or Separation Problem; Structural Problem; Activation, Positioning or Separation Problem
SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 11284883 ↗MalfunctionJun 10, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
ISSUES WITH A ESOPHAGEAL STENT. SURGEON WANTED A 20 FR STENT WHICH THEY TOOK FROM THE ENDOSCOPY UNIT DOWNSTAIRS. THE SURGEON HAD NEVER USED THE COOK STENTS BUT HE SAID WAS FAMILIAR WITH THE BOSTON SCIENTIFIC ONES. WHILE HE WAS DEPLOYING IT HE DID EXPERIENCE SOME DIFFICULTY TOWARDS THE LAST PART OF IT BUT THEN HE THOUGHT ¿.MAY BE THAT¿S HOW IT IS AND THEN WHILE HE WAS PULLING IT BACK THE WHOLE STENT ALSO CAME ALONG WITH IT AND COULD NOT BE USED ANYMORE. THE RELEASE PIN WAS NOT PULLED BEFORE CATHETER REMOVAL. PATIENT OUTCOME: DID ANY PIECE OF THE DEVICE REMAIN INSIDE THE PATIENT¿S BODY? NO. PLEASE DESCRIBE THE OBJECT AND HOW IT WAS RETRIEVED (IF APPLICABLE): DID THE PATIENT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE? NO. DID THE PRODUCT CAUSE OR CONTRIBUTE TO THE NEED FOR ADDITIONAL PROCEDURE(S)? NO. PLEASE SPECIFY ADDITIONAL PROCEDURE(S) AND PROVIDE DETAILS: DID THE COMPLAINANT INFORM OF ANY ADVERSE EFFECT(S) ON THE PATIENT DUE TO THIS OCCURRENCE? NO. DID THE COMPLAINANT INFORM THAT THE PRODUCT CAUSED OR CONTRIBUTED TO THE ADVERSE EFFECT(S)? NO.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10137288 ↗MalfunctionJan 29, 2020Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
EVENT DESCRIBED BY DR.(B)(6) ,AFTER PLACING AN .035 STIFF WIRE THROUGH THE SCOPE ,UNDER FLORO , THE STENT WAS ADVANCED ON THE STIFF GUIDE WIRE, AFTER POSITIONING IT WELL , THEY STARTED TO DEPLOY THE STENT , UNFORTUNATELY THEY COULD NOT SEE THE STENT OPENING UNDER XRAY, THE RED VISUAL MARKER ON THE TOP OF THE HANDLE WAS COMING BACKWARD , BUT EVEN AFTER IT REACHES POINT OF NO RETURN THE PROXIMAL END OF THE STENT WAS NOT OPENING , THE METAL STYLET WAS REMOVED BUT STILL THEY COULD NOT DEPLOY THE STENTS , THE WHOLE ASSEMBLY WAS TAKEN OUT OF THE PATIENT , EVEN IN THE AIR THE STENT WAS NOT OPENING . THE PROCEDURE WAS COMPLETED USING OTHER COMPANY STENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9642838 ↗InjuryJun 4, 2019Difficult to Remove; Difficult to Remove
AS INITIALLY REPORTED TO CUSTOMER RELATIONS: PER DISTRIBUTORS EMAIL: OUR QA STAFF REPORTED THE FOLLOWING CASE: DOCTORS WERE PERFORMING AN ESOPHAGEAL STRICTURE. THE DOCTOR INSERTED THE STENT AND ONCE READY FOR RELEASE INSIDE THE PATIENT THE SECURE DID NOT COME OUT, THE STENT STARTED TO EXPAND INTO THE PATIENT BUT IT COULD NOT BE RELEASED BECAUSE OF THE STUCKED SAFE CAUSING AN ESOPHAGEAL TEARING. PER THE CUSTOMERS COMPLAINT FORM: AT THE MOMENT OF THE INTRODUCTION AND PROCEEDING WITH THE RELEASED OF THE STENT, IT COULDNT BE RELEASED BECAUSE OF THE SAFE. THE DOCTOR DECIDES TO TAKE THE STENT AWAY CAUSING ESOPHAGEAL TEARING.
Patient problems: Excessive Tear Production
MAUDE report 8665885 ↗MalfunctionFeb 13, 2018Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
AFTER PULLING OUT THE WIRE TO FINALLY PLACE THE STENT, THE STENT COULD NOT BE PLACED IN THE STRICTURE. HENCE THE ENTIRE SYSTEM HAD TO BE PULLED OUT
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7266087 ↗MalfunctionNov 29, 2016Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
A FOLLOW-UP MDR IS BEING SUBMITTED TO INCLUDE THE DEVICE EVALUATION DETAILS. STENT WAS BEING DEPLOYED, THEN WAS RETRACTED TO REPOSITION. STENT WOULD NOT RETRACT PROPERLY WITH SOME OF THE STENT BEING VISIBLE AFTER RETRACTION. ENDOSCOPE AND STENT REMOVED FROM PATIENT AS PROCEDURE COULD NOT CONTINUE WITH COOK STENT. ENDOSCOPE INSERTED BACK INTO PATIENT AND A NEW STENT FROM ANOTHER SUPPLIER WAS USED TO COMPLETE PROCEDURE. A MALFUNCTION PRECEDENCE EXISTS FOR A DEPLOYMENT ISSUE THAT RESULTS IN AN EXPOSED STENT BEING REMOVED FROM THE PATIENT WITH THE DELIVERY SYSTEM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6131639 ↗InjurySep 12, 2016Migration or Expulsion of Device; Activation, Positioning or Separation Problem; Migration or Expulsion of Device; Activation, Positioning or Separation Problem
THE USER ATTEMPTED TO PLACE EVO-20-25-15-E INSIDE AN 8CM BARE STENT PREVIOUSLY PLACED IN THE ESOPHAGUS. HE ALIGNED THE DISTAL END OF THE EVO STENT AT THE DISTAL END OF THE 8CM BARE STENT, AND DEPLOYED THE STENT AS EXTENDING TO THE PROXIMAL SIDE. HOWEVER, IT WAS 'STENT-IN-STENT', SO THE EVO STENT WAS NOT APPRESSED TO THE ESOPHAGUS WALL. THE EVO STENT MIGRATED TO THE PROXIMAL SIDE BECAUSE THE DELIVERY SYSTEM TIP CAUGHT THE EVO STENT DURING REMOVAL. STENTING BECAME LONGER THAN THE USER PLANNED, SO HE REMOVED THE EVO STENT USING A GRASPING FORCEPS AND PLACED A 12.5CM STENT SUCCESSFULLY TO COMPLETE THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5942546 ↗MalfunctionAug 24, 2016Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
INFORMATION INITIALLY RECEIVED INDICATED THE EVOLUTION STENT FAILED TO DEPLOY. NOTE THE STENT WAS RETURNED FULLY DEPLOYED WITH DISPLACED LASSO LOOPS. ADDITIONAL INFORMATION RELATING TO THIS COMPLAINT HAS BEEN REQUESTED BUT TO DATE CLARIFICATION HAS NOT BEEN RECEIVED. FDA MDR REPORTABLE BASED UPON A REPORTING PRECEDENCE ESTABLISHED FOR THIS DEVICE FAMILY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5899593 ↗MalfunctionMay 11, 2016Fracture; Fracture
DR. SHOWED ME A VIDEO OF A ENDOSCOPY WHERE A PIECE (~ 3CMS) OF THE EVOLUTION DISTAL END (OF THE STENT) WAS TOTALLY BROKEN, THIS MEAN IN TWO PIECES ONE PART IMPLANTED AND THE OTHER IN THE STOMACH. PATIENT RETURNED WITH APHAGIA - THAT IS WHEN THEY CONFIRMED THE ISSUE. A NEW DEVICE WAS IMPLANTED TO RECTIFY THE ISSUE. THE STENT HAD BEEN IMPLANTED IN THE PATIENT FOR 3 YEARS PRIOR TO THIS OCCURRENCE.
Patient problems: Dysphagia/ Odynophagia
MAUDE report 5646580 ↗MalfunctionApr 12, 2016Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
THE EVOLUTION STENT COULDN'T BE RELEASED - EVEN WITH THE SAFETY WIRE PULLED. WHEN THE USER REMOVED THE INTRODUCER THE FULLY EXPANDED STENT WAS ALSO REMOVED OUT OF THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5570344 ↗
FDA clearance and classification
The FDA regulates product code ESW (prosthesis, esophageal) under 21 CFR 878.3610 as a Class II device. 21 CFR 878.3610 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Evolution G48033
Has the Evolution G48033 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10827002480336), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Evolution G48033?
The FDA's MAUDE database lists 19 adverse event reports for this device, received between Apr 12, 2016 and Apr 6, 2026. In those reports, the most common specific device problems are “Activation, Positioning or Separation Problem” (9), “Improper or Incorrect Procedure or Method” (5) and “Physical Resistance/Sticking” (3); the most common patient problems are “Dysphagia/ Odynophagia” (1), “Excessive Tear Production” (1) and “Obstruction/Occlusion” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Evolution G48033?
None of the reports are classified as deaths, 4 as injuries and 15 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Evolution G48033 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Evolution G48033, and is it still sold?
The FDA's device database lists Cook Ireland Ltd as the company responsible for it (brand Evolution, model G48033). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other PROSTHESIS, ESOPHAGEAL devices
- WallFlex™ Esophageal M00516730Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall29 reports
- WallFlex™ Esophageal M00516740Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall1 death22 reports
- WallFlex™ Esophageal M00516750Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall1 death19 reports
- WallFlex™ Esophageal M00516700Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall18 reports
- WallFlex™ Esophageal M00516910Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall15 reports
- WallFlex™ Esophageal M00516710Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall13 reports
- WallFlex™ Esophageal M00516930Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall13 reports
- WallFlex™ Esophageal M00516940Boston Scientific Corporation · PROSTHESIS, ESOPHAGEAL1 recall12 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page