Skip to content

Advance G50223Recalls & reported problems

Cook Incorporated · Advance, 18LP Low Profile PTA Balloon Dilatation Catheter · Class 2

All models of Advance (206)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10827002502236), received between May 14, 2021 and May 14, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Rupture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2021Material Rupture; Material Rupture

    ADDITIONAL INFORMATION RECEIVED ON 07JUN2021: THE BALLOON WAS REMOVED BY PULLING. A 5/6FR CORDIS SHEATH WAS UTILIZED. AN INFLATION DEVICE WAS USED. THE RUPTURE OCCURRED DURING THE FIRST INFLATION OF THE BALLOON, AND IT WAS INFLATED TO LESS THAN NOMINAL PRESSURE. THE BALLOON WAS NOT INFLATED INSIDE OF THE STENT PRIOR TO RUPTURE. THERE WAS NO NOTED TO BE SIGNIFICANT ANGULATION OR VIRTUOSITY AND LOCATION OF THE LESION WAS THE AT ORIGIN. THE VESSEL WAS NOTED TO BE MODERATELY/HEAVILY CALCIFIED. THERE WAS NO NOTED OCCLUSION. THE BALLOON MARKERS STARTED SEPARATING ON THE WIRE WHEN ATTEMPTING TO REMOVE. USING ROTATION AND TENSION, THE DEVICE WAS ABLE TO BE REMOVED. THE VESSEL OCCLUDED DUE TO THIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11824587 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates