Advance G50224Recalls & reported problems
Cook Incorporated · Advance, 18LP Low Profile PTA Balloon Dilatation Catheter · Class 2
All models of Advance (206)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002502243), received between Jul 13, 2017 and Jan 28, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
- Hole In Material1
- Inflation Problem1
- Material Puncture/Hole1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 28, 2022Material Puncture/Hole; Material Puncture/Hole
AS REPORTED, DURING AN UNSPECIFIED PROCEDURE A ADVANCE 18 LP LOW PROFILE BALLOON CATHETER WAS FOUND NOT TO INFLATE DUE TO A LEAK. THE BALLOON CATHETER WAS ADVANCED TO THE POPLITEAL ARTERY VIA A COOK GUIDING SHEATH. THE USER BEGAN INFLATING WITH A BALLOON EXPANSION PRESSURE PUMP, BUT UPON INFLATING THEY FOUND THAT PRESSURE COULD NOT BE MAINTAINED AND THE CONTRAST SUBSTANCE WAS LEAKING FROM THE BALLOON. NO ANGULATION OR VESSEL TORTUOSITY WAS NOTED. THE BALLOON WAS THEN REMOVED WITH THE SHEATH. ANOTHER NEW BALLOON CATHETER WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13384697 ↗MalfunctionAug 24, 2017Hole In Material; Inflation Problem; Hole In Material; Inflation Problem
AN ADVANCE 18 LP LOW PROFILE BALLOON CATHETER WAS USED IN A TIBIAL ANGIOPLASTY. IT WAS REPORTED, THE BALLOON HAD A HOLE IN IT AND WAS UNABLE TO BE INFLATED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. IT HAS NOT BEEN REPORTED THAT THE PATIENT REQUIRED ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6816705 ↗MalfunctionJul 13, 2017Burst Container or Vessel
THE INFORMATION RECEIVED INDICATED THAT WHEN THE VASCULAR SURGEON INFLATED THE BALLOON TO RATED BURST PRESSURE OF 14 ATM IN THE POSTERIOR TIBIAL, THE BALLOON BURST. THE BALLOON WASN'T KEPT. A PIECE OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT, THE PATIENT DID NOT REQUIRE ADDITIONAL PROCEDURES AND THERE WERE NO ADVERSE EFFECTS TO THE PATIENT ASSOCIATED WITH THE ISSUE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6710684 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page