Advance G50227Recalls & reported problems

Cook Incorporated · Advance, 18LP Low Profile PTA Balloon Dilatation Catheter · Class 2

All models of Advance (206)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002502274), received between Aug 7, 2018 and Jun 20, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Fluid/Blood Leak1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 20, 2024Break; Break

    AS REPORTED, DURING A BALLOON DILATION OF THE "LEFT PYELO-URETERAL SCHEDULE", AN ADVANCE 18 LP LOW PROFILE BALLOON CATHETER SEPARATED. THE BALLOON WAS ADVANCED OVER ANOTHER MANUFACTURER'S GUIDEWIRE AND THROUGH ANOTHER MANUFACTURER'S CYSTOSCOPE THROUGH THE URETHRA. ANOTHER MANUFACTURER'S CATHETER WAS USED FOR CONTRAST INTRODUCTION. THE BALLOON WAS INFLATED IN THE LEFT SKYLE-URETERAL JUNCTION TO FOURTEEN ATMOSPHERES ONE TIME FOR THREE MINUTES. ONCE THE BALLOON WAS COMPLETELY DEFLATED AND RETURNED TO AMBIENT PRESSURE, A COUNTERCLOCKWISE ROTATION WAS USED TO REMOVE THE BALLOON WHILE MAINTAINING NEGATIVE PRESSURE. THE REMOVAL WAS CONTROLLED WITH MICROSCOPY (X-RAY). UPON REACHING THE URETERO-VESICAL JUNCTION, THE USER ENCOUNTERED PROBLEMS/DIFFICULTY WITH REMOVAL. THE USER MADE NEGATIVE PRESSURE WITH THE NANOMETER AND TURNED THE BALLOON. THE CYSTOSCOPE WAS REMOVED FROM THE PATIENT, FOLLOWED BY THE BALLOON CATHETER. "FORCE" WAS USED TO REMOVE THE BALLOON AND ONCE REMOVED, THE BALLOON CATHETER WAS BROKEN. THE CYSTOSCOPE WAS REINSERTED, AND REMAINS OF THE BALLOON WERE FOUND. THE REMAINS WERE SUCCESSFULLY REMOVED WITH FORCEPS/TWEEZERS. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19575903 ↗
  • MalfunctionJan 16, 2019Material Rupture; Material Rupture

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT. THIS INFORMATION IS DETAILED IN SECTION H10.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8253006 ↗
  • MalfunctionAug 7, 2018Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED, THE ADVANCE 18 LP LOW PROFILE BALLOON CATHETER LEAKED DURING AN UNSPECIFIED PROCEDURE. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT HAS NOT BEEN MADE AVAILABLE. NO ADVERSE EFFECTS HAVE BEEN REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7758762 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →