Advance G50236Recalls & reported problems
Cook Incorporated · Advance, 18LP Low Profile PTA Balloon Dilatation Catheter · Class 2
All models of Advance (206)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002502366), received between Sep 19, 2017 and Feb 18, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak1
- Hole In Material1
- Material Rupture1
- Material Separation1
Patient problems
- Foreign Body In Patient1
- No Code Available1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 18, 2020Material Rupture; Material Separation; Material Rupture; Material Separation
AS REPORTED, DURING ANGIOPLASTY AND STENTING OF THE ILIAC ARTERIES AT THE AORTIC BIFURCATION, AN ADVANCE 18 LP LOW PROFILE BALLOON CATHETER RUPTURED CIRCUMFERENTIALLY AND SEPARATED UPON THE FIRST INFLATION. THE ANATOMY WAS REPORTED TO BE VERY SEVERELY CALCIFIED WITH ANGULATION OVER THE AORTIC BIFURCATION, WHERE THE 100% OCCLUDED LESION WAS LOCATED. AN UNKNOWN 6 FRENCH SHEATH WAS UTILIZED DURING THE PROCEDURE. AN UNKNOWN INFLATION DEVICE AND A 50:50 RATIO OF OMNIPAQUE 300 SOLUTION WERE USED TO INFLATE THE BALLOON, WHICH RUPTURED AT ELEVEN ATMOSPHERES. A THREE-CENTIMETER PORTION OF THE BALLOON THEN SEPARATED FROM THE SHAFT. THE BALLOON CATHETER WAS REMOVED OVER AN UNKNOWN WIRE GUIDE, MAINTAINING NEGATIVE PRESSURE AFTER ALLOWING THE DEVICE TO RETURN TO AMBIENT PRESSURE, UTILIZING A COUNTERCLOCKWISE ROTATION. THE USER ATTEMPTED TO REMOVE THE SEPARATED PORTION OF THE BALLOON FROM THE ILIAC ARTERY; HOWEVER, THIS WAS UNSUCCESSFUL. THE PHYSICIAN THEN SUCCESSFULLY STENTED THE SEPARATED PORTION OF THE BALLOON AGAINST THE VESSEL WALL, USING AN UNKNOWN STENT. THE BALLOON WAS NOT INFLATED WITHIN A STENT PRIOR TO RUPTURE. BLOOD WAS NOTED IN THE INFLATION DEVICE AT THE TIME OF RUPTURE. ACCORDING
Patient problems: Foreign Body In Patient; No Code Available
MAUDE report 9721203 ↗MalfunctionSep 19, 2017Fluid/Blood Leak; Hole In Material; Fluid/Blood Leak; Hole In Material
AN ADVANCE 18 LP LOW PROFILE BALLOON CATHETER WAS USED IN AN UNSPECIFIED PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY. IT WAS REPORTED, DURING INFLATION, CONTRAST LEAKED FROM THE TIP OF THE BALLOON. THE BALLOON LEAK, WHICH THE PHYSICIAN BELIEVES TO BE A PINHOLE, WAS NOTICED PRIOR TO INFLATION AT NOMINAL PRESSURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6875839 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page