Advance G50237Recalls & reported problems
Cook Incorporated · Advance, 18LP Low Profile PTA Balloon Dilatation Catheter · Class 2
All models of Advance (206)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002502373), received between Jun 27, 2016 and Feb 22, 2017 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Material Rupture1
- Sticking1
Patient problems
- No Code Available1
- No Known Impact Or Consequence To Patient1
- Thrombus1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 22, 2017Sticking; Sticking
IT WAS REPORTED, DURING A PERIPHERAL LOWER LEG CASE, THE ADVANCE 18 LP LOW PROFILE BALLOON CATHETER WAS BEING REMOVED OVER A V18 BOSTON SCIENTIFIC WIRE, AND THE BALLOON BECAME STUCK, AS IF THE WIRE WAS STICKY. THE WIRE AND BALLOON HAD TO BE REMOVED FROM THE PATIENT TOGETHER. ALSO DURING THE PROCEDURE, A RUBICON CROSSING CATHETER WAS USED WITH AN ABBOTT WIRE AND THE SAME DIFFICULTY OCCURRED. THE PROCEDURE WAS COMPLETED AND THE PATIENT REMAINED UNHARMED. THE USER FACILITY REPORTED THEY WERE UNSURE WHAT CAUSED THE DEVICES TO BECOME STUCK AND WHY THE WIRE BECAME STICKY. NO ADDITIONAL DETAILS HAVE BEEN RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6349154 ↗InjuryJun 27, 2016Material Rupture; Detachment of Device or Device Component; Material Rupture; Detachment of Device or Device Component
DURING A RIGHT FEMORAL ANGIOPLASTY PROCEDURE, FEMORAL PUNCTURE WAS CONDUCTED FROM THE IPSILATERAL APPROACH (RIGHT) . THE PHYSICIAN REPORTED DIFFICULT PUNCTURE DUE TO BODY HABITUS. A 4FR SHORT INTRODUCER WAS IN SITU. FOCAL CALCIFIED LESIONS NOTED ON INITIAL IMAGING THROUGH SFA. 0.014" WIRE (ANOTHER MANUFACTURER) WAS USED TO CROSS LESIONS. PRIMARY DILATION WITH 4MM BALLOON (COOK PTA4). SECONDARY DILATION WITH 5MM PTA4 BALLOON. BALLOON SEEN TO RUPTURE DURING PLASTY OF LESION IN MID SFA AREA. MAXIMUM PRESSURE USED OF 10ATM. A SMALL AMOUNT OF CONTRAST WAS SEEN TO REMAIN WITHIN THE RUPTURED BALLOON. NEGATIVE PRESSURE APPLIED WITH INFLATION DEVICE TO PROMOTE COMPLETE DEFLATION OF BALLOON. BALLOON WITHDRAWN OVER GUIDE WIRE UNTIL RESISTANCE FELT WHEN BALLOON REACHED PROXIMAL SFA. PHYSICIAN REPORTS APPLYING ADDITIONAL TENSION AT THIS POINT AS HE FELT THAT THE BALLOON WAS PARTIALLY WITHDRAWN INTO FEMORAL ACCESS SHEATH AND WAS PLANNING TO WITHDRAW BOTH THE RUPTURED BALLOON AND THE ACCESS SHEATH TOGETHER. THE PHYSICIAN REPORTS FEELING THE RESISTANCE ON THE BALLOON SUDDENLY GIVE. ON EXAMINATION OF WITHDRAWN SHEATH AND BALLOON, THE DISTAL TIP OF BALLOON CATHETER WAS MISSING. UNDER FLUOROSCOPY, DI
Patient problems: No Code Available; Thrombus
MAUDE report 5751490 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page