Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002522548), received between Jul 13, 2018 and Oct 19, 2023 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture3
- Difficult to Remove1
- Inflation Problem1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- Perforation of Vessels1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 19, 2023Difficult to Remove; Material Rupture; Difficult to Remove; Material Rupture
AS REPORTED, DURING ANGIOPLASTY, AN ADVANCE 35 LP LOW PROFILE BALLOON CATHETER'S BALLOON RUPTURED. A COOK SHEATH AND WIRE GUIDE WERE USED DURING THE PROCEDURE. THE ANATOMY WAS REPORTEDLY NOT TORTUOUS OR CALCIFIED. VENOPLASTY WAS PERFORMED FOR RIGHT BRACHIOCEPHALIC STENOSIS. THE BALLOON WAS INFLATED TO AN UNKNOWN PRESSURE; HOWEVER, THE BALLOON REPORTEDLY RUPTURED. THE BALLOON RETURNED TO AMBIENT PRESSURE AND NEGATIVE PRESSURE WAS MAINTAINED UPON ATTEMPTED REMOVAL OF THE DEVICE; HOWEVER, A COUNTERCLOCKWISE ROTATION WAS NOT CONDUCTED. THE BALLOON CATHETER WAS UNABLE TO BE REMOVED OVER THE WIRE; THEREFORE, THE BALLOON CATHETER AND THE WIRE GUIDE WERE REMOVED TOGETHER. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17965168 ↗InjurySep 11, 2018Material Rupture; Material Rupture
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: Perforation of Vessels
MAUDE report 7865404 ↗MalfunctionJul 13, 2018Inflation Problem; Material Rupture; Material Integrity Problem; Inflation Problem; Material Rupture; Material Integrity Problem
IT WAS REPORTED, DURING TREATMENT OF FEMORAL ARTERY OCCLUSIONS OF THE PATIENT'S LOWER EXTREMITIES, THE ADVANCE 35 LP LOW PROFILE BALLOON CATHETER COULD NOT BE INFLATED. THE PHYSICIAN ADVANCED THE BALLOON THROUGH THE CALCIFIED ARTERY TO THE TARGET SITE. AFTER DISCOVERING THE COMPLAINT DEVICE WOULD NOT INFLATE, IT WAS WITHDRAWN. A SPLIT WAS FOUND IN THE BALLOON. A 6 FRENCH SHEATH WAS BEING UTILIZED DURING THE PROCEDURE. SUBSEQUENTLY, THE COMPLAINT DEVICE WAS CHANGED FOR A NEW BALLOON CATHETER AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7686640 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page