Advance G52265Recalls & reported problems
Cook Incorporated · Advance, 35LP Low Profile PTA Balloon Dilatation Catheter · Class 2
All models of Advance (206)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10827002522654), received between Dec 22, 2016 and Apr 12, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture4
- Deflation Problem1
- Material Puncture/Hole1
- Unsealed Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 12, 2024Material Rupture; Material Rupture
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19097496 ↗MalfunctionMay 17, 2022Unsealed Device Packaging; Unsealed Device Packaging
AS REPORTED, PRIOR TO AN UNKNOWN PROCEDURE, AN ADVANCE 35 LP LOW PROFILE BALLOON CATHETER'S STERILE PACKAGE WAS FOUND TO BE UNSEALED ON ONE SIDE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14427009 ↗MalfunctionApr 18, 2022Material Puncture/Hole; Material Rupture; Material Puncture/Hole; Material Rupture
AS REPORTED, DURING AN AORTIC-ILIAC-FEMORAL INTERVENTION, AN ADVANCE 35 LP LOW PROFILE BALLOON CATHETER RUPTURED. A COOK 6 FRENCH X 55-CENTIMETER HI-FLEX STRAIGHT ANSEL SHEATH WAS USED DURING THE PROCEDURE, AS WELL AS A COOK SPHERE INFLATION DEVICE. ACCESS WAS GAINED IN THE RIGHT ILIAC ARTERY AND A CONTRALATERAL APPROACH WAS USED. THE VESSELS WERE REPORTEDLY TORTUOUS, CALCIFIED, AND VERY SMALL. UPON THE INITIAL INFLATION TO PRE-DILATE THE 85% OCCLUDED LESION, PRIOR TO PLACING A STENT, THE BALLOON REPORTEDLY RUPTURED ON A PIECE OF CALCIUM WITHIN THE DISTAL LEFT SUPERFICIAL FEMORAL ARTERY. BLOOD WAS NOTED IN THE INFLATION DEVICE AT THE TIME OF RUPTURE. THE BALLOON WAS REMOVED BY ITSELF, OVER AN UNKNOWN WIRE GUIDE. ANOTHER MANUFACTURER'S BALLOON WAS THEN USED; HOWEVER, THIS BALLOON ALSO RUPTURED. A THIRD (OTHER MANUFACTURER'S) BALLOON WAS USED TO COMPLETE THE INTENDED PROCEDURE. THE COMPLAINT DEVICE WAS NOT INFLATED WITHIN A STENT PRIOR TO RUPTURING. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14136979 ↗MalfunctionApr 23, 2018Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING A PERIPHERAL ARTERIOGRAM, THE ADVANCE 35 LP LOW PROFILE BALLOON CATHETER DEVICE RUPTURED AT 8 ATMOSPHERES (ATM), PRIOR TO REACHING NOMINAL PRESSURE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING ANOTHER ADVANCE 35 LP LOW PROFILE BALLOON CATHETER. ACCORDING TO THE INITIAL REPORTER, A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7451284 ↗MalfunctionMar 9, 2017Material Rupture; Material Rupture
IT WAS REPORTED, DURING AN UNSPECIFIED PROCEDURE, AN ADVANCE 35 LP LOW PROFILE BALLOON CATHETER WAS USED TO TREAT SEVERE CALCIFIED LESION OF THE EXTERNAL ILIAC ARTERY, BY WAY OF THE BRACHIAL ARTERY, HOWEVER THE BALLOON RUPTURED. ANOTHER ADVANCE 35 LP WAS USED INSTEAD TO COMPLETE THE PROCEDURE. THERE HAVE BEEN NO ADVERSE EFFECTS TO THE PATIENT REPORTED. THE DEVICE WILL NOT BE RETURNED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6393015 ↗MalfunctionDec 22, 2016Deflation Problem; Deflation Problem
RECOIL OF THE PATIENT'S STENOSIS WAS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6200327 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page