Advance G52276Recalls & reported problems
Cook Incorporated · Advance, 35LP Low Profile PTA Balloon Dilatation Catheter · Class 2
All models of Advance (206)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002522760), received between Oct 17, 2018 and Nov 19, 2018 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Material Rupture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Rupture2
- Material Separation1
Patient problems
- No Code Available1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 19, 2018Material Rupture; Material Rupture
IT WAS REPORTED, THE ADVANCE 35 LP LOW PROFILE BALLOON CATHETER RUPTURED DURING AN UNSPECIFIED PROCEDURE. THIS INFORMATION WAS GATHERED FROM (B)(4) (REPORT REFERENCE NUMBER 1820334-2018-03152). WHEN THE COMPLAINT DEVICE WAS RECEIVED BY THE MANUFACTURER, IT WAS NOTED THERE WERE TWO INCOMPLETE DEVICES INCLUDED IN THE RETURN PACKAGING. THE COMPLAINT DEVICE FROM (B)(4) WAS IDENTIFIED BY ITS SEPARATED STATE, AND LACK OF CATHETER. THE SECOND ADVANCE 35 LP LOW PROFILE BALLOON CATHETER RECEIVED CONSISTED OF THE A RUPTURED BALLOON STILL IN-TACT WITH THE CATHETER PORTION OF THE DEVICE (REFERENCE THIS REPORT). UPON REACHING OUT TO THE FACILITY, THE PHYSICIAN CONFIRMED THAT ONLY ONE BALLOON ((B)(4)) HAD RUPTURED DURING THE PROCEDURE, REQUIRING A SURGICAL CUT-DOWN TO REMOVE THE SEPARATED DISTAL PORTION OF THE DEVICE. THE TREATING PHYSICIAN WAS UNABLE PROVIDE ANY DETAILS SURROUNDING THE SECOND DEVICE. NEITHER THE PHYSICIAN NOR THE AREA REPRESENTATIVE WERE ABLE TO EXPLAIN WHY AN EXTRA BALLOON DEVICE WAS INCLUDED IN THE PACKAGING. THEREFORE, NO PATIENT HARM OR ADVERSE EFFECTS HAVE BEEN REPORTED REGARDING THE DEVICE CAPTURED IN THIS REPORT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8087956 ↗InjuryOct 17, 2018Material Rupture; Material Separation; Material Rupture; Material Separation
IT WAS REPORTED, AN ADVANCE 35 LP LOW PROFILE BALLOON CATHETER RUPTURED UPON INFLATION TO RATED BURST PRESSURE IN A HEAVILY CALCIFIED LESION IN THE SUPERFICIAL FEMORAL ARTERY. A MIXTURE OF CONTRAST TO SALINE WAS USED. PROPORTIONS NOT SPECIFIED. THE TIP OF THE BALLOON BROKE OFF AT SOME POINT WHEN THE USER WAS REMOVING THE BALLOON. THE PHYSICIAN HAD TO DO A CUT DOWN TO REMOVE THE BALLOON TIP WHICH WAS FOUND RIGHT AT THE COMMON FEMORAL ARTERY. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Code Available
MAUDE report 7976609 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page