Advance G52280Recalls & reported problems
Cook Incorporated · Advance, 35LP Low Profile PTA Balloon Dilatation Catheter · Class 2
All models of Advance (206)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10827002522807), received between Apr 27, 2018 and Dec 19, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash2
- Material Puncture/Hole2
- Inflation Problem1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 19, 2023Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18358171 ↗MalfunctionDec 1, 2020Material Puncture/Hole; Material Puncture/Hole
NO NEW PATIENT OR EVENT INFORMATION SINCE THE LAST REPORT WAS SUBMITTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10926832 ↗MalfunctionNov 14, 2018Material Rupture; Material Rupture
IT WAS REPORTED, DURING A PERIPHERAL ANGIOPLASTY PROCEDURE TWO ADVANCE 35 LP LOW PROFILE BALLOON CATHETER DEVICES RUPTURED. ACCORDING TO THE INITIAL REPORTER, THE RUPTURES OCCURRED ON THE BACK TABLE, DURING PREP, AND THE DEVICES NEVER MADE PATIENT CONTACT. IT WAS REPORTED THAT THE BALLOONS WERE INFLATED USING A MIX OF 2CC OF CONTRAST WITH 8 CC OF SALINE. IT WAS STATED THAT BOTH OF THE BALLOONS BURST AT 2-3 ATMS. IT IS UNKNOWN WHICH DIRECTION THE RUPTURES OCCURRED. ADDITIONALLY, A SIMILAR DEVICE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. SEE THE MANUFACTURER REPORT NUMBER 1820334-2018-03230 FOR THE SECOND BALLOON. THIS REPORT IS FOR THE FIRST BALLOON.
Patient problems: No Patient Involvement
MAUDE report 8072241 ↗MalfunctionApr 27, 2018Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash
IT WAS REPORTED THAT DURING A TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT (TIPS) PROCEDURE, THE ADVANCE 35 LP LOW PROFILE BALLOON CATHETER WOULD ONLY INFLATE TO 2 ATMOSPHERES. THE PHYSICIAN WITHDREW THE COMPLAINT DEVICE AND EXCHANGED IT FOR A NEW ONE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. ACCORDING TO THE COMPLAINANT, A LEAK WAS FOUND NEAR THE VALVE OF THE DEVICE. ACCORDING TO THE INITIAL REPORTER, A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7468625 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page