Advance G52301Recalls & reported problems
Cook Incorporated · Advance, 35LP Low Profile PTA Balloon Dilatation Catheter · Class 2
All models of Advance (206)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002523019), received between Jul 13, 2017 and Dec 3, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Hole In Material1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 3, 2024Break; Break
AS REPORTED, DURING A PROCEDURE INVOLVING BALLOON DILATION OF A STENOSED ILIAC VEIN, AN ADVANCE 35 LP LOW PROFILE BALLOON CATHETER SEPARATED. THE ANATOMY WAS REPORTEDLY SLIGHTLY STENOSED. A COOK 10-FRENCH SHEATH WAS ALSO USED DURING THE PROCEDURE. PER THE REPORTER, AN UNSPECIFIED INFLATION DEVICE WAS USED TO INFLATE THE BALLOON TO FIVE ATMOSPHERES. THE BALLOON WAS INFLATED TWO TIMES, FOR TWO MINUTES EACH TIME, TO GRADUALLY DILATE THE LESION, WHICH WAS REPORTEDLY SEVENTY PERCENT OCCLUDED. DURING DILATION, THE BALLOON CATHETER SUDDENLY ¿BROKE¿ WITHIN THE VESSEL. THE USER IS REPORTEDLY UNSURE IF THE BALLOON RUPTURED PRIOR TO SEPARATION OF THE CATHETER; HOWEVER, BLOOD WAS NOT NOTED IN THE INFLATION DEVICE. THE BALLOON WAS NOT INFLATED WITHIN A STENT. ONCE DEFLATED, THE BALLOON WAS ALLOWED TO RETURN TO AMBIENT PRESSURE, AND NEGATIVE PRESSURE WAS MAINTAINED UPON ATTEMPTED REMOVAL OF THE DEVICE. THE USER ATTEMPTED TO REMOVE THE BALLOON CATHETER AND SHEATH AS A UNIT. A COUNTERCLOCKWISE ROTATION OF THE BALLOON WAS NOT CONDUCTED UPON REMOVAL. THE SEPARATED CATHETER FRAGMENTS WERE SUCCESSFULLY REMOVED WITH AN UNSPECIFIED FILTER RETRIEVAL DEVICE, VIA A JUGULAR APPROACH TO THE INFERIOR VENA CAV
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20835847 ↗MalfunctionJul 13, 2017Hole In Material; Leak/Splash; Hole In Material; Leak/Splash
WHILE PERFORMING INTERVENTIONAL VENOUS STENTING, IT WAS REPORTED THAT A PINHOLE LEAK WAS OBSERVED IN THE PROXIMAL PART OF THE BALLOON MATERIAL OF THE LOW PROFILE BALLOON CATHETER. IT WAS PROVIDED THAT THE PATIENT'S VENOUS ANATOMY WAS NOT CALCIFIED, AND A 50/50 MIXTURE OF CONTRAST AND SALINE WAS USED IN THE BALLOON. NO ADVERSE EFFECTS HAVE BEEN REPORTED, AND THERE WAS NO NEED FOR ADDITIONAL PROCEDURES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6710229 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page