Advance G52331Recalls & reported problems
Cook Incorporated · Advance, 35LP Low Profile PTA Balloon Dilatation Catheter · Class 2
All models of Advance (206)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002523316), received between Jul 12, 2017 and Jul 6, 2020 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Rupture2
- Unsealed Device Packaging1
Patient problems
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 6, 2020Unsealed Device Packaging; Unsealed Device Packaging
AS REPORTED, PRIOR TO USE DURING AN ANGIOGRAM, THE SEAL WAS NOTED TO BE MISSING FROM AN ADVANCE 35 LP LOW PROFILE BALLOON CATHETER PACKAGE. AS THE USER OPENED THE PACKAGE, THE SEAL WAS NOTED TO BE MISSING AND THE BALLOON WAS MISSING FROM THE SHAFT. THE DEVICE DID NOT MAKE CONTACT WITH ANY PATIENT. A NEW DEVICE OF THE SAME TYPE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Patient Involvement
MAUDE report 10233454 ↗MalfunctionJul 6, 2020Material Rupture; Material Rupture
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10233324 ↗MalfunctionJul 12, 2017Material Rupture
THE DEVICE WAS USED TO TREAT A LESION WITH CALCIFICATION. THE BALLOON RUPTURED DURING THE PROCEDURE. ANOTHER MANUFACTURER'S DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE HAVE BEEN NO ADVERSE EFFECTS TO THE PATIENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6708046 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page