Skip to content

Advance G52348Recalls & reported problems

Cook Incorporated · Advance, 35LP Low Profile PTA Balloon Dilatation Catheter · Class 2

All models of Advance (206)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002523484), received between Jul 13, 2017 and Dec 26, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 26, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, DURING A PERIPHERAL ANGIOPLASTY PROCEDURE VIA A CONTRALATERAL APPROACH, THE HUB SEPARATED FROM AN ADVANCE 35 LP LOW PROFILE BALLOON CATHETER. A TWENTY-MILLILITER SYRINGE WAS USED TO FLUSH THE DEVICE DURING PREPARATION. PER THE REPORTER, MULTIPLE CATHETERS AND BALLOONS USED DURING THE PROCEDURE WERE DIFFICULT TO TRACK THROUGH ANOTHER MANUFACTURER¿S SHEATH, OVER AN UNSPECIFIED ROSEN WIRE GUIDE. THE COMPLAINT DEVICE WAS DIFFICULT TO ADVANCE UP AND OVER A TIGHT BIFURCATION, THROUGH THE OTHER MANUFACTURER¿S SHEATH. PER THE REPORTER, THE PHYSICIAN MENTIONED THAT THE OTHER MANUFACTURER¿S SHEATH MIGHT HAVE MIGRATED INTO THE AORTA, MAKING THE ¿BEND¿ MORE ACUTE. AFTER THE BALLOON WAS USED, ¿A LOT OF¿ RESISTANCE WAS ENCOUNTERED UPON ATTEMPTED REMOVAL OF THE DEVICE OVER A WIRE GUIDE. WITH THE USE OF FORCE TO PULL THE CATHETER, THE HUB SEPARATED. THE BALLOON CATHETER WAS THEN REMOVED WITH THE WIRE GUIDE AS A UNIT, AT WHICH POINT THE USER NOTED THAT THE CATHETER WAS STRETCHED BUT OTHERWISE INTACT. PER THE REPORTER, THE PHYSICIAN ¿DOES NOT THINK THE BALLOON CATHETER WAS THE ISSUE.¿ THE OTHER MANUFACTURER¿S SHEATH WAS EXCHANGED FOR ANOTHER DEVICE, AFTER WHICH THE SURGEON REPORTEDLY FO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21020585 ↗
  • MalfunctionJul 13, 2017Burst Container or Vessel

    THE INFORMATION RECEIVED INDICATED THAT THE BALLOON BURST BEFORE OPTIMUM PRESSURE. THE BALLOON WAS REMOVED, A NEW BALLOON WAS OPENED AND USED. A PIECE OF THE DEVICE DID NOT REMAIN IN THE PATIENT, THE EVENT DID NOT REQUIRE AN ADDITIONAL PROCEDURE AND THERE WERE NO REPORTED ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE ISSUE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6711066 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page