Advance G52348Recalls & reported problems
Cook Incorporated · Advance, 35LP Low Profile PTA Balloon Dilatation Catheter · Class 2
All models of Advance (206)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002523484), received between Jul 13, 2017 and Dec 26, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Burst Container or Vessel1
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 26, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING A PERIPHERAL ANGIOPLASTY PROCEDURE VIA A CONTRALATERAL APPROACH, THE HUB SEPARATED FROM AN ADVANCE 35 LP LOW PROFILE BALLOON CATHETER. A TWENTY-MILLILITER SYRINGE WAS USED TO FLUSH THE DEVICE DURING PREPARATION. PER THE REPORTER, MULTIPLE CATHETERS AND BALLOONS USED DURING THE PROCEDURE WERE DIFFICULT TO TRACK THROUGH ANOTHER MANUFACTURER¿S SHEATH, OVER AN UNSPECIFIED ROSEN WIRE GUIDE. THE COMPLAINT DEVICE WAS DIFFICULT TO ADVANCE UP AND OVER A TIGHT BIFURCATION, THROUGH THE OTHER MANUFACTURER¿S SHEATH. PER THE REPORTER, THE PHYSICIAN MENTIONED THAT THE OTHER MANUFACTURER¿S SHEATH MIGHT HAVE MIGRATED INTO THE AORTA, MAKING THE ¿BEND¿ MORE ACUTE. AFTER THE BALLOON WAS USED, ¿A LOT OF¿ RESISTANCE WAS ENCOUNTERED UPON ATTEMPTED REMOVAL OF THE DEVICE OVER A WIRE GUIDE. WITH THE USE OF FORCE TO PULL THE CATHETER, THE HUB SEPARATED. THE BALLOON CATHETER WAS THEN REMOVED WITH THE WIRE GUIDE AS A UNIT, AT WHICH POINT THE USER NOTED THAT THE CATHETER WAS STRETCHED BUT OTHERWISE INTACT. PER THE REPORTER, THE PHYSICIAN ¿DOES NOT THINK THE BALLOON CATHETER WAS THE ISSUE.¿ THE OTHER MANUFACTURER¿S SHEATH WAS EXCHANGED FOR ANOTHER DEVICE, AFTER WHICH THE SURGEON REPORTEDLY FO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21020585 ↗MalfunctionJul 13, 2017Burst Container or Vessel
THE INFORMATION RECEIVED INDICATED THAT THE BALLOON BURST BEFORE OPTIMUM PRESSURE. THE BALLOON WAS REMOVED, A NEW BALLOON WAS OPENED AND USED. A PIECE OF THE DEVICE DID NOT REMAIN IN THE PATIENT, THE EVENT DID NOT REQUIRE AN ADDITIONAL PROCEDURE AND THERE WERE NO REPORTED ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE ISSUE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6711066 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page