Advance G52353Recalls & reported problems
Cook Incorporated · Advance, 35LP Low Profile PTA Balloon Dilatation Catheter · Class 2
All models of Advance (206)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10827002523538), received between Nov 7, 2016 and Nov 9, 2023 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture6
- Burst Container or Vessel1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- No Code Available1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 9, 2023Material Rupture; Material Rupture
ADDITIONAL INFORMATION RECEIVED 07NOV2023 STATING THAT AN UNSPECIFIED INFLATION DEVICE WAS USED. THE BALLOON "EXPLODED" UPON FIRST INFLATION TO APPROXIMATELY 7 ATMOSPHERES. THE PATIENT'S ANATOMY WAS NOT STENOSED, TORTIOUS, OR CALCIFIED, HOWEVER, THE ARTERIOVENOUS SHUNT WAS APPROXIMATELY 50% STENOSED. THE PROCEDURE WAS COMPLETED WITH AN UNSPECIFIED COMPETITOR'S BALLOON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18099894 ↗MalfunctionFeb 24, 2021Material Rupture; Material Rupture
NO NEW PATIENT OR EVENT INFORMATION TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11372687 ↗MalfunctionAug 12, 2019Material Rupture; Material Rupture
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8883783 ↗InjuryMay 16, 2019Material Rupture; Material Rupture
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Code Available
MAUDE report 8616521 ↗MalfunctionOct 22, 2018Material Rupture; Material Rupture
IT WAS REPORTED, DURING AN ANGIOPLASTY OF A FISTULA, THE ADVANCE 35 LP LOW PROFILE BALLOON CATHETER WOULD NOT HOLD CONTRAST WHICH WAS A 50/50 MIXTURE OF VISIPAQUE AND SALINE. IT COULD BE SEEN LEAKING OUT UNDER X-RAY. THE BALLOON WAS NEVER INFLATED. THE COMPLAINT DEVICE WAS REMOVED AND ANOTHER OF THE SAME TYPE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. PATIENT OUTCOME WAS REPORTED AS GOOD. PER THE COMPLAINANT THE COMPLAINT DEVICE WAS PROBABLY BROKEN UPON ENTRY. THE PATIENT'S ANATOMY WAS NOT CALCIFIED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7993737 ↗MalfunctionJul 26, 2017Material Rupture; Material Rupture
IT IS ALLEGED THAT AN ADVANCE 35 LP LOW PROFILE BALLOON CATHETER WAS BEING USED AS A PHYSICIAN WAS ATTEMPTING TO PLACE A STENT IN A DIALYSIS FISTULA LOCATED IN THE PATIENT'S RIGHT ARM WHEN THE BALLOON RUPTURED CIRCUMFERENTIALLY DURING INFLATION OF THE BALLOON. IT WAS REPORTED THAT A 50/50 MIXTURE OF SALINE AND ISOVIEW WAS USED DURING THE PROCEDURE, AND THE INFLATION PRESSURE IS UNKNOWN. IT WAS ALSO REPORTED THAT THERE WAS SLIGHT ANGULATION INSIDE OF THE VESSEL BUT NO CALCIFICATION. THERE WAS NO PART OF THE DEVICE LEFT IN THE PATIENT. NO ADVERSE EFFECTS HAVE BEEN REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6745479 ↗InjuryNov 7, 2016Burst Container or Vessel; Burst Container or Vessel
THE BALLOON RUPTURED CIRCUMFERENTIALLY. UPON PULLING THE BALLOON OUT, PIECES OF THE BALLOON WERE LEFT INSIDE THE PATIENT AND HAD TO BE REMOVED USING A SNARE. HOWEVER, ONE BALLOON MARKER HAD GONE DOWN TO PARENTERAL, AND THE USER WAS UNABLE TO RETRIEVE THIS PIECE. NOT FLOW LIMITING, THEREFORE THE PIECE REMAINED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A NEW DEVICE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6081711 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page