Advance G52372Recalls & reported problems
Cook Incorporated · Advance, 35LP Low Profile PTA Balloon Dilatation Catheter · Class 2
All models of Advance (206)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10827002523729), received between Dec 5, 2014 and Mar 9, 2022 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Rupture2
- Balloon rupture1
- Failure to Advance1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Clinical Signs, Symptoms or Conditions1
- Nonresorbable materials, unretrieved in body1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 9, 2022Material Rupture; Material Rupture
AS REPORTED, DURING A PROCEDURE INVOLVING PLACEMENT OF A STENT WITHIN THE COMMON ILIAC VEIN, AN ADVANCE 35 LP LOW PROFILE BALLOON CATHETER RUPTURED LONGITUDINALLY UPON THE FIRST INFLATION WITHIN A COOK 16-MILLIMETER STENT. THE PATIENT HAD A HISTORY OF VENOUS DISEASE, AND THE ANATOMY WAS CALCIFIED (PER INTRAVENOUS ULTRASOUND). AN UNSPECIFIED 8 FRENCH, 13-CENTIMETER COOK SHEATH WAS USED DURING THE PROCEDURE. THE COMPLAINT DEVICE WAS INFLATED TO ELEVEN ATMOSPHERES, USING AN UNKNOWN INFLATION DEVICE, TO PRE-DILATE THE COMMON ILIAC VEIN (WITHIN THE STENT) AT THE TIME OF RUPTURE. BLOOD WAS NOTED IN THE INFLATION DEVICE AT THE TIME OF RUPTURE. THE BALLOON WAS REMOVED THROUGH THE SHEATH, AND ANOTHER MANUFACTURER'S 10-MILLIMETER BALLOON WAS USED TO SUCCESSFULLY FINISH THE PROCEDURE AND PLACE AN UNSPECIFIED COOK 10-MILLIMETER STENT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13710185 ↗MalfunctionDec 19, 2016Material Rupture; Material Rupture
IT WAS REPORTED DURING A FISTULA PROCEDURE THE LOW PROFILE CATHETER BALLOON POPPED WHEN BALLOONING THE FISTULA. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6186862 ↗MalfunctionAug 8, 2016Failure to Advance; Failure to Advance
IT WAS REPORTED THAT THE LOW PROFILE BALLOON CATHETER SECTION THAT CONNECTED THE PART OF BALLOON AND CATHETER WAS ROUGH AND FAILED TO GO THROUGH THE VESSEL SHEATH DURING THE PROCEDURE. THE BALLOON WAS WITHDRAWN FROM THE PATIENT AND THE OPERATION WAS CANCELLED AS ANOTHER DEVICE WAS NOT AVAILABLE TO COMPLETE THE PROCEDURE. IT IS UNKNOWN AT THE TIME OF THIS REPORT IF ANOTHER PROCEDURE WAS PERFORMED AT A LATER DATE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES NOR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5852947 ↗InjuryDec 5, 2014Balloon rupture
A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES NOR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: Nonresorbable materials, unretrieved in body
MAUDE report 4335984 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page