Zenith TX2 Pro-Form G53435Recalls & reported problems
William Cook Europe ApS · Zenith TX2, TAA Endovascular Graft with Pro-Form, Proximal Component · Class 3 · Implant
All models of Zenith TX2 Pro-Form (15)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002534350), received between Dec 13, 2016 and Dec 3, 2019 — classified as 1 death, 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture1
- Kinked1
- Migration1
- Migration or Expulsion of Device1
Patient problems
- Death1
- No Consequences Or Impact To Patient1
- No Information1
- Paraplegia1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 3, 2019Fracture; Migration; Fracture; Migration
DESCRIPTION OF EVENT ACCORDING TO STUDY: ON (B)(6) 2014, THE PATIENT RECEIVED A ZTEG-2P-38-202-PF-US (LOT# E3187738) AND A ZTEG-2P-38-152-PF-US (LOT# E31849901). THERE WERE NO DIFFICULTIES DEPLOYING THE DEVICE. A MOLDING BALLOON FROM ANOTHER MANUFACTURER WAS USED WITHOUT DIFFICULTY. NO PATIENT-RELATED ADVERSE EVENTS WERE OBSERVED DURING THE PROCEDURE. PER SITE, THE DEVICE WAS PATENT WITH NO KINK OR ENDOLEAK ON THE COMPLETION ANGIOGRAM. ANALYSIS NOTED A PATENT GRAFT WITH NO KINK. A POSSIBLE ENDOLEAK WAS NOTED AT THE OVERLAP OF THE TWO DEVICES AND AT THE MID/DISTAL ASPECT OF THE DISTAL COMPONENT. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL ON (B)(6) 2014 (FOUR DAYS POST-PROCEDURE). (B)(6) 2014 NON-CONTRAST CT SCAN (65 DAYS POST-PROCEDURE): ANALYSIS: NO KINK NOTED. ANEURYSM DIAMETER = 67.1 MM. NO OTHER ASSESSMENTS WERE MADE DUE TO NON-CONTRAST EXAM OF POOR QUALITY. (B)(6) 2015 CT SCAN (198 DAYS POST-PROCEDURE): ANALYSIS: GRAFTS PATENT, INTACT WITH NO KINK. A SMALL DISTAL ENDOLEAK OF UNKNOWN ORIGIN WAS NOTED. MAXIMUM ANEURYSM DIAMETER = 59.5 MM. (B)(6) 2015 CT SCAN (366 DAYS POST-PROCEDURE): ANALYSIS: GRAFTS PATENT, INTACT WITH NO MIGRATION OR KINK. AN ENDOLEAK OF UNKNOWN TYPE WAS NOT
Patient problems: No Information
MAUDE report 9400637 ↗DeathJun 25, 2018Migration or Expulsion of Device; Migration or Expulsion of Device
DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: "PATIENT DIED, CAUSE OF DEATH UNKNOWN." TVAR COMPONENTS SEPARATED, ZTA IN (B)(4) IMPLANTED TO REPAIR OR RECONNECT OR MITIGATE THIS. TVAR COMPONENTS HAD SEPARATED AFTER SEVERAL YEARS. ADDITIONAL INFORMATION RECEIVED 12JUN2018: "PATIENT EXPIRED ON (B)(6) 2018. "PATIENT EXPERIENCED SPINAL SHOCK FROM PARAPLEGIA, BUT THIS WAS NOT DEVICE-SPECIFIC OR RELATED. THE PHYSICIAN ASSERTS THAT THE PROCEDURE RESULTED IN THE PARAPLEGIA COMPLICATION, AND THAT IT WAS NOT DEVICE SPECIFIC OR RELATED." "PRODUCT INFORMATION: A) (B)(6) 2011 INDEX TEVAR: ZTEG-2D-34-186-PF-US, LOT E2463077----AND ZTEG-2P-38-202-PF-US, LOT E2776443. B) (B)(6) 2014 TEVAR (PROXIMAL EXTENSION FOR DISEASE PROGRESSION) ZTEG-2P-38-127-PF-US, LOT E3078098" PATIENT OUTCOME: PATIENT IS DECEASED.
Patient problems: Death; Paraplegia
MAUDE report 7633027 ↗InjuryDec 13, 2016Kinked; Kinked
DESCRIPTION OF EVENT ACCORDING TO STUDY: ON (B)(6) 2014, THE PATIENT RECEIVED A ZTEG-2P-38-202-PF-US (LOT # E3187738) AND A ZTEG-2P-38-152-PF-US (LOT # E31849901). THERE WERE NO DIFFICULTIES DEPLOYING THE DEVICE. A MOLDING BALLOON WAS USED WITHOUT DIFFICULTY. NO PATIENT-RELATED ADVERSE EVENTS WERE OBSERVED DURING THE PROCEDURE. PER SITE, THE DEVICE WAS PATENT WITH NO KINK OR ENDOLEAK ON THE COMPLETION ANGIOGRAM. ANALYSIS NOTED A PATENT GRAFT WITH NO KINK. A POSSIBLE ENDOLEAK WAS NOTED AT THE OVERLAP OF THE TWO DEVICES AND AT THE MID/DISTAL ASPECT OF THE DISTAL COMPONENT. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL ON (B)(6) 2014 (FOUR DAYS POST-PROCEDURE). ON (B)(6) 2014, NON-CONTRAST CT SCAN (65 DAYS POST-PROCEDURE): ANALYSIS: NO KINK NOTED. ANEURYSM DIAMETER = 67.1 MM. NO OTHER ASSESSMENTS WERE MADE DUE TO NON-CONTRAST EXAM OF POOR QUALITY. ON (B)(6) 2015, CT SCAN (198 DAYS POST-PROCEDURE): ANALYSIS: GRAFTS PATENT, INTACT WITH NO KINK. A SMALL DISTAL ENDOLEAK OF UNKNOWN ORIGIN WAS NOTED. MAXIMUM ANEURYSM DIAMETER = 59.5 MM. ON (B)(6) 2015, CT SCAN (366 DAYS POST-PROCEDURE): ANALYSIS: GRAFTS PATENT, INTACT WITH NO MIGRATION OR KINK. AN ENDOLEAK OF UNKNOWN TYPE WAS NOTED ((B)(4)).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6168726 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page