Cantata G54531Recalls & reported problems
Cook Incorporated · Cantata, 2.5 Microcatheter · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002545318), received between Jul 28, 2017 and Aug 8, 2019 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Material Separation1
Patient problems
- No Code Available2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 8, 2019Material Separation; Material Separation
NO ADDITIONAL PATIENT OR EVENT INFORMATION HAS BEEN RECEIVED SINCE THE LAST REPORT WAS SUBMITTED ON 29AUG2019.
Patient problems: No Code Available
MAUDE report 8873352 ↗InjuryJul 28, 2017Break; Break
INTERNATIONAL CUSTOMER REPORTED THAT A PATIENT WAS DIAGNOSED BY CT ( COMPUTED TOMOGRAPHY) SCAN WITH AN ABDOMINAL BLEED. THE ETIOLOGY OF THE BLEED WAS NOT IDENTIFIED. THE PATIENT WAS UNDERGOING A PROCEDURE TO LOCATE AND EMBOLIZE THE SOURCE OF THE BLEED. THE PHYSICIAN ATTEMPTED TO LOCATE THE SOURCE OF THE BLEED BY PROBING THE TRUNCUS WITH A COMPETITOR PRODUCT AND PROBING THE LEFT GASTRIC ARTERY WITH A CANTATA DUO 2.5 SUPERSELECTIVE MICROCATHETER. THE PHYSICIAN PULLED THE CANTATA DUO 2.5 SUPERSELECTIVE MICROCATHETER IN ORDER TO PROBE THE AMS WITH THE COMPETITOR PRODUCT. A PORTION OF THE MICROCATHETER REMAINED WITHIN THE AORTA AT THE TIP OF THE COMPETITOR PRODUCT WHILE ATTEMPTING TO REMOVE THE DEVICE. THE BROKEN PORTION WAS RETRIEVED USING A SLING DEVICE AND REMOVED THROUGH A SECOND 4 FR CATHETER. THE DEVICE IS AVAILABLE FOR EVALUATION.
Patient problems: No Code Available
MAUDE report 6751014 ↗
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths980 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths56 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page