Zenith Spiral-Z G55226Recalls & reported problems
Cook Incorporated · Zenith, Spiral-Z AAA Iliac Leg · Class 3 · Implant
All models of Zenith Spiral-Z (29)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 1
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10827002552262), received between Jun 8, 2017 and Jul 3, 2024 — classified as 1 death, 5 injuries.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak4
- Improper or Incorrect Procedure or Method2
- Leak/Splash1
- Material Puncture/Hole1
- Occlusion Within Device1
Patient problems
- Clinical Endoleak3
- No Code Available3
- Ruptured Aneurysm1
- Thrombosis1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
DeathJul 3, 2024Fluid/Blood Leak; Fluid/Blood Leak
ADDITIONAL INFORMATION WAS PROVIDED ON 23JUL2024 IN THE FORM OF OFFICIAL TRANSLATION OF PREVIOUSLY PROVIDED REPORTS. THE PATIENT UNDERWENT ENDOVASCULAR AORTIC REPAIR (EVAR) ON 11MAR2013 FOR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM (AAA). THE FOLLOWING COOK DEVICES WERE IMPLANTED DURING THE PROCEDURE: COOK ZENITH LOW PROFILE AAA ENDOVASCULAR GRAFT MAIN BODY (RPN: ZALB-28-108). COOK ZENITH FLEX WITH SPIRAL-Z TECHNOLOGY AAA ENDOVASCULAR GRAFT ILIAC LEG (RPN: ZSLE-11-74-ZT). COOK ZENITH FLEX WITH SPIRAL-Z TECHNOLOGY AAA ENDOVASCULAR GRAFT ILIAC LEG (RPN: ZSLE-13-56-ZT). COOK ZENITH FLEX WITH SPIRAL-Z TECHNOLOGY AAA ENDOVASCULAR GRAFT ILIAC LEG (RPN: ZSLE-13-74-ZT). THE PATIENT PRESENTED TO THE EMERGENCY ROOM OF (B)(6) ON (B)(6) 2024. HE WAS CONSCIOUS AND RATED A 15 ON THE GLASCOW COMA SCALE. HE HAD A SUDDEN ONSET OF LOWER BACK PAIN THAT BEGAN ON (B)(6) 2024 AROUND 17:00 AFTER CARRYING HEAVY LOADS. HIS PAIN WAS DESCRIBED AS CRAMP-LIKE AND WAS LOCATED IN THE LEFT VERTEBRAL COLUMN. HE WAS HYPOTENSIVE AT 94/70 MMHG. AT 70 BPM HIS HEART RATE WAS NORMAL. MOTTLING WAS NOT PRESENT AND THERE WERE NO SIGNS OF DEPOSITS. THERE WAS NO INTER-ARM BLOOD PRESSURE DIFFERENT. HE WAS PANTING AT 30 BREATH
Patient problems: Clinical Endoleak; Ruptured Aneurysm
MAUDE report 19671216 ↗InjuryJan 26, 2023Fluid/Blood Leak; Improper or Incorrect Procedure or Method; Fluid/Blood Leak; Improper or Incorrect Procedure or Method
NO ADDITIONAL INFORMATION REGARDING PATIENT AND/OR EVENT DETAILS HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: Clinical Endoleak
MAUDE report 16245715 ↗InjuryJan 26, 2023Fluid/Blood Leak; Improper or Incorrect Procedure or Method; Fluid/Blood Leak; Improper or Incorrect Procedure or Method
NO ADDITIONAL INFORMATION REGARDING PATIENT AND/OR EVENT DETAILS HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: Clinical Endoleak
MAUDE report 16245690 ↗InjuryJul 11, 2019Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT A TYPE IIIB ENDOLEAK WAS OBSERVED IN ZENITH FLEX WITH SPIRAL-Z TECHNOLOGY AAA ENDOVASCULAR GRAFT ILIAC LEG, RPN ZSLE-11-74-ZT. THE (B)(6) YEAR OLD PATIENT HAD A PRE-EXISTING (B)(6). THE PATIENT ORIGINALLY HAD THE FOLLOWING DEVICES PLACED ON (B)(6) 2019: ZENITH FLEX AAA ENDOVASCULAR GRAFT BIFURCATED MAIN BODY (TFFB-26-82-ZT). ZENITH BRANCH ENDOVASCULAR GRAFT- ILIAC BIFURCATION (ZBIS-10-45-41). ZENITH FLEX WITH SPIRAL-Z TECHNOLOGY AAA ENDOVASCULAR GRAFT ILIAC LEG (ZSLE-16-39-ZT). ZENITH FLEX WITH SPIRAL-Z TECHNOLOGY AAA ENDOVASCULAR GRAFT ILIAC LEG (ZSLE-11-74-ZT). DURING THIS PROCEDURE, BALLOONING WAS COMPLETED USING A CODA BALLOON AT THE OVERLAP AND THE DISTAL END OF THE GRAFT. A 30ML SYRINGE WAS USED TO INFLATE THE BALLOON TO AN UNKNOWN VOLUME OF INFLATION. AN INSUFFLATOR WAS NOT USED FOR INFLATION. A SECONDARY INTERVENTION WAS COMPLETED ON (B)(6) 2019. ANOTHER ZENITH FLEX WITH SPIRAL-Z TECHNOLOGY AAA ENDOVASCULAR GRAFT ILIAC LEG (ZSLE-11-74-ZT) WAS IMPLANTED TO COVER THE GRAFT DEFECT.
Patient problems: No Code Available
MAUDE report 8781404 ↗InjurySep 13, 2017Occlusion Within Device; Occlusion Within Device
THE PATIENT HAD PREVIOUSLY BEEN TREATED FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) AND A RIGHT INTERNAL ILIAC OCCLUSION USING A SELF OR BALLOON EXPANDABLE UNCOVERED STENT THAT EXTENDED FROM THE PATIENT¿S AORTIC BIFURCATION ALL THE WAY INTO THE EXTERNAL ILIAC . IN (B)(6) 2015, A FENESTRATED ENDOVASCULAR ANEURYSM REPAIR (FEVAR) WAS PERFORMED USING A ZENITH FLEX WITH SPIRAL-Z TECHNOLOGY AAA ENDOVASCULAR GRAFT ILIAC LEG (ZSLE). THIS WAS PERFORMED BY A DIFFERENT PHYSICIAN AND AT ANOTHER FACILITY. THE PATIENTS RT INTERNAL ILIAC WAS ALL READY OCCLUDED PRIOR TO THIS ZFEN PROCEDURE. AS REPORTED TO COOK MEDICAL BY THE PATIENT'S NEW PHYSICIAN, THE PREVIOUSLY IMPLANTED ZENITH FLEX WITH SPIRAL-Z TECHNOLOGY AAA ENDOVASCULAR GRAFT ILIAC LEG (ZSLE) WAS THROMBOSED AND COMPLETELY OCCLUDED. THE PATIENT REQUIRED EMERGENT INTERVENTION TO TREAT CRITICAL LIMB ISCHEMIA WITH AN OCCLUDED RIGHT ILIAC LIMB. DURING THE INTERVENTION PROCEDURE, THE PREVIOUSLY IMPLANTED ZSLE WAS RE-ALIGNED WITH A LIMB FROM A DIFFERENT COMPANY. THE PHYSICIAN USED BALLOON EXPANDABLE COVERED STENTS PLACED JUST PROXIMAL TO THE FLOW DIVIDER IN THE AORTOILIAC SEGMENTS AND INFLATED TO 11 ATMOSPHERES. AS THERE WERE STILL RESIDUAL THROMBUS, A
Patient problems: No Code Available; Thrombosis
MAUDE report 6862615 ↗InjuryJun 8, 2017Leak/Splash; Leak/Splash
DESCRIPTION OF EVENT ACCORDING TO STUDY: ENDOLEAK OF UNKNOWN TYPE 5 YEARS POST-IMPLANT ON (B)(6) 2012, THE PATIENT RECEIVED ONE ZTLP-PT-38-167-CI PROXIMAL COMPONENT. THERE WAS NO DIFFICULTY DEPLOYING THE DEVICE. COIL EMBOLIZATION OF THE LEFT SUBCLAVIAN ARTERY WAS PERFORMED AFTER DEVICE DEPLOYMENT. ON THE COMPLETION ANGIOGRAM, THE CORE LAB NOTED THAT THE DEVICES WERE PATENT WITH NO KINK OR ENDOLEAK. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL ON (B)(6) 2012 (THREE DAYS POST-PROCEDURE). FOLLOW-UP CT SCANS ON (B)(6) 2012 (1-MONTH FOLLOW-UP, 23 DAYS POST-PROCEDURE) ANALYSIS: ULCER DEPTH 24 MM. DEVICES PATENT AND INTACT WITH NO KINKS OR ENDOLEAKS. ON (B)(6) 2012 (6-MONTH FOLLOW-UP, 186 DAYS POST-PROCEDURE) ANALYSIS: ULCER DEPTH 24 MM. DEVICES PATENT AND INTACT WITH NO MIGRATION, KINKS OR ENDOLEAKS. ON (B)(6) 2013 (12-MONTH FOLLOW-UP, 368 DAYS POST-PROCEDURE) ANALYSIS: ULCER DEPTH 25 MM. DEVICES PATENT AND INTACT WITH NO MIGRATION, KINKS OR ENDOLEAKS. ON (B)(6) 2014 (2-YEAR FOLLOW-UP, 732 DAYS POST-PROCEDURE) ANALYSIS: ULCER DEPTH 22 MM. DEVICES PATENT AND INTACT WITH NO MIGRATION, KINKS OR ENDOLEAKS. ON (B)(6) 2015 (3-YEAR FOLLOW-UP, 1082 DAYS POST-PROCEDURE) ANALYSIS: ULCER DEPTH 20 M
Patient problems: No Code Available
MAUDE report 6621268 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page